Driver, Prosthesis
FDA adverse event reports and recall history for driver, prosthesis (product code HWR), from the public MAUDE corpus and the live FDA recall database.
- Total reports
- 9,892
- Deaths
- 1
- Injuries
- 245
- Malfunctions
- 9,644
Reports per year
Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.
Recent recalls
- 2016-01-29TerminatedEmployee error
Biosure Driver, Product No: 72201887 For delivery and placement of orthopedic screws. Orthopedic Manual Surgical Instrument
Driver does not meet specification, oversized. Driver may not be able to be fully inserted into the screw, or screw may become stuck on the driver.
- 2010-04-11TerminatedDevice Design
BIOMET MICROFIXATION, PATIENT DISTRACTOR DRIVER 10.87cm, (4.28")(L), RADEL, STAINLESS STEEL, NON-STERILE PRODUCT. Part Number SP-1918 BIOMET MICROFIXATION 1520 TRADEPORT DRIVE, JACKSONVILLE, FL 32218 USA.
The patient Distractor Driver, used in Osteotomy procedures, may not ratchet and advance the distraction device as intended. If the distraction device does not advance, bone consolidation may occur and a second surgery may be necessary.
- 2007-11-27TerminatedDevice Design
Odyssey Femoral Holder/Driver, Part Number K001-7410, Non-sterile, 1 each, Wright Medical Technology, Inc., 5677 Airline Road, Arlington, TN 38002
Breakage: breaking of the connecting foot that is on the distal surface of the instrument
- 2007-11-27TerminatedComponent change control
Pioneer Quantum Spinal Rod System, MIS Yoke Manipulator, Long Slot, non-sterile; Cat # 51-YOKEMAN-L, Pioneer Surgical Technology, Marquette, MI 49855
The pins may come out of the instrument and, if this happens, they could fall into/remain in the patient.
- TerminatedNonconforming Material/Component
Oxford Unicompartmental Knee Phase 3 Tibial Impactor Product Usage: Instruments for use with the Oxford Uni Partial Knee System
Zimmer Biomet is conducting a medical device field action for various lot-specific Oxford Knee instruments due to potential alumina inclusions in the raw material batch used to produce the affected products. Inclusions contained within the finished product could lead to the cracking and separation of the instrument.
- TerminatedNonconforming Material/Component
Oxford Partial Knee System Phase 3 Tibial Template Right Medial Size A Product Usage: Instruments for use with the Oxford Uni Partial Knee System
Zimmer Biomet is conducting a medical device field action for various lot-specific Oxford Knee instruments due to potential alumina inclusions in the raw material batch used to produce the affected products. Inclusions contained within the finished product could lead to the cracking and separation of the instrument.
Monitor driver, prosthesis automatically
DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.