Skip to main content
HTWClass 1

Bit, Drill

FDA adverse event reports and recall history for bit, drill (product code HTW), from the public MAUDE corpus and the live FDA recall database.

Total reports
18,028
Deaths
5
Injuries
2,418
Malfunctions
15,356

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Recent recalls

  • 2026-03-27Open, ClassifiedUnder Investigation by firm

    Embrace Drill Tower, Wedged (25mm); Item Number: 645-081/62;

    The surgical technique for the Embrace Shoulder Instruments - Drill Tower has mixed-up article numbers in the overview of the instruments. However, the individual surgical steps described in the surgical technique are correct.

  • 2026-03-27Open, ClassifiedUnder Investigation by firm

    Embrace Drill Tower, Standard/Lateral (25mm); Item Number: 645-081/63;

    The surgical technique for the Embrace Shoulder Instruments - Drill Tower has mixed-up article numbers in the overview of the instruments. However, the individual surgical steps described in the surgical technique are correct.

  • 2024-03-06Open, ClassifiedUnder Investigation by firm

    Embrace Drill Tower, Wedged (25mm)

    The surgical techniques for the Embrace Shoulder Instruments -Drill Tower (Article REF 645-081/62 & /63) have mixed-up type descriptions ("standard" vs "wedged") in the overview of the instruments.

  • 2024-03-06Open, ClassifiedUnder Investigation by firm

    Embrace Drill Tower, Standard/Lateral (25mm)

    The surgical techniques for the Embrace Shoulder Instruments -Drill Tower (Article REF 645-081/62 & /63) have mixed-up type descriptions ("standard" vs "wedged") in the overview of the instruments.

  • 2020-12-17TerminatedProcess control

    RingLoc+ Quick Connect Drill Bit, Item numbers 31-323215, 31-323220, 31-323230, 31-323240 - Product Usage: Bit, Drill.

    Specific lots of multiple sterile medical devices may have a breach in the sterile seal. A breach in the packaging seal could lead to loss of sterility of the device.

  • 2016-05-26TerminatedEquipment maintenance

    Various trauma and sports medicine instruments and implants. Bit, drill.

    Insufficient seal strength on sterile barrier packaging. The seal could be jeopardized causing the sterile barrier to be compromised and the product could become nonsterile. If the product becomes non-sterile and is used or implanted, it may lead to an infection.

  • 2016-05-23TerminatedDevice Design

    3.2 mm Three-fluted drill bit, sterile, Catalog ID # 03.010.060S Product Usage: Usage: The 3.2 mm Three-fluted Drill Bits are instruments that can be used with the following systems: Expert Lateral Femoral Nail (part number 03.010.060S only), Expert Tibial Nail, Expert Humeral Nailing, Multiloc Humeral Nailing System, Suprapatellar Instrumentation for Expert Tibial Nail, and Adolescent Lateral Femoral Nails (ALFN).

    The sterile packaging may be compromised. A protective cap is to be secured on the tip of the drill bit. If the protective cap is not secured, the drill bit may puncture the sterile barrier.

  • 2016-05-23TerminatedDevice Design

    3.2 mm three-fluted Radiolucent Drill Bit/Needle Point/145 mm Sterile Product Usage: The 3.2 mm Three-fluted Drill Bits are instruments that can be used with the following systems: Expert Lateral Femoral Nail (part number 03.010.060S only), Expert Tibial Nail, Expert Humeral Nailing, Multiloc Humeral Nailing System, Suprapatellar Instrumentation for Expert Tibial Nail, and Adolescent Lateral Femoral Nails (ALFN).

    The sterile packaging may be compromised. A protective cap is to be secured on the tip of the drill bit. If the protective cap is not secured, the drill bit may puncture the sterile barrier.

  • 2015-08-24TerminatedNonconforming Material/Component

    3.0/4.0 Cannulated Drill Bit; a component of the Tiger Cannulated Screw System The Tiger Cannulated Screw Fixation System implants are intended for fixation of fractures, non-unions, arthrodesis and osteotomies of the small bones in the hand and foot. The implants and guide wires are intended for single use only.

    The cannulation of the 3.0/4.0 Cannulated Drill Bit inner diamter is too small to allow for the guide wire to pass completely through.

  • 2015-06-04TerminatedComponent design/selection

    3.2 mm Cannulated Drill Bit/ QC 170 mm 2.7 Cannulated Drill Bit/ QC 160 mm 2.0 mm Cannulated Drill Bit / QC 150 mm there is no indication for Synthes Cannulated Drill Bits which is used with the Synthes Cannulated Screw. The Synthes Cannulated Screw, used with the threaded washer, is generally intended for intra-articular fixation of small bones, such as the hand, wrist, and forefoot.

    affected parts and lots of the Cannulated Drill Bits have the potential to break during use

Monitor bit, drill automatically