Broach
FDA adverse event reports and recall history for broach (product code HTQ), from the public MAUDE corpus and the live FDA recall database.
- Total reports
- 4,734
- Deaths
- 2
- Injuries
- 240
- Malfunctions
- 4,483
Reports per year
Source: FDA adverse event reports by year received, 2011–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.
Recent recalls
- 2011-07-05TerminatedProcess control
Foundation Knee System Instrumentation, Tibial Broach Assembly. Used in conjunction with the Tibial Sizing Template and Tibial Broach Guide to prepare the resected tibial plateau for acceptance of a keeled tibial baseplate stem.
Strike plate may dislodge from the broach stem handle during impaction of the device.
- 2009-07-07TerminatedDevice Design
Zimmer TM-400 Broach Handle; Part No. 96-210-10001. The product is a surgical instrument for use in conjunction with Zimmer TM-400 Trabecular Metal implants during spinal surgery. The Broach Handle threads onto various broaches and trial implants. The handle with these broaches or trials attached is used to prepare and measure a vertebral space to accept a TM-400 Trabecular Metal implant.
The broach handle may fracture at the threaded tip during use, which would result in surgical intervention to retrieve the broach or the trial as the threaded tip fragment prevents attachment of another handle to retrieve the device.
- 2009-05-12TerminatedFinished device change control
Zimmer Nexgen Complete Knee Solution Fluted Stemmed Tibial Broach, Size 7, 8, nonsterile; REF 5997-61-07. A reusable cutting instrument that is used in total knee arthroplasty surgery.
The surface finish inside of the broach is not to specification and the instrument may corrode if cleaned and sterilized according to instructions.
- 2008-09-19TerminatedDevice Design
Rejuvenate Broach Catalog No: 1601-1005, 1601-1006, 1601-1007, 1601-1008, 1601-1009, 1601-1010 Stryker Orthopaedics, Mahwah, NJ. 07430 Contours the proximal femur to the geometry of the stem to provide a complimentary fit of the stem to the bone.The broaches are designed with a series of teeth that are intended to seat against highly densified cancellous bone for the interference press-fit in the uncemented application.
Rejuvenate Broach system was designed in a manner that may result in the potential for a medical calcar fracture.
- TerminatedNonconforming Material/Component
Linear Straight Broach Handle (Rasp Handle) (Greatbatch Part # 510397; DJO Surgical Inc., # 803-03-159) and EZ Clean" Linear Broach Handle Greatbatch Part # T11824, DJO Surgical Inc. # 803-03-059).
Greatbatch Medical has voluntarily initiated a Recall for the Linear Straight Broach Handle (Rasp Handle) and EZ Clean" Linear Broach Handle. Greatbatch Medical has received reports that that the product is breaking at the welds and that the locating pin at the distal end of the instrument is breaking off during use.
Monitor broach automatically
DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.