Skip to main content
HTJClass 1

Gauge, Depth

FDA adverse event reports and recall history for gauge, depth (product code HTJ), from the public MAUDE corpus and the live FDA recall database.

Total reports
3,500
Deaths
0
Injuries
60
Malfunctions
3,438

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Recent recalls

  • 2011-03-31TerminatedDevice Design

    Xcelerate Specialty Patella Milling System Per File K2399; Catalog Number I-K2399PR00, I-K2399PR10; Stryker Howmedica Osteonics, Howmedica Osteonics Corp., 325 Corporate Drive, Mahwah, NJ 07430 The Xcelerate patella milling system (as detailed in surgical protocol LSPK35) was designed to prepare a patella to accept all Scorpio or Duracon patellar implants using either a resurfacing or recessed technique.

    Stryker Orthopaedics has become aware that the 10 mm on the bone remaining depth gage is incorrect thus allowing for the reamer to leave as little as 8.9mm (in a nominal condition) patella bone.

  • 2010-02-19TerminatedProcess control

    Depth Gauge; Catalog Number: 703707; Stryker Trauma AG, CH 2545 Selzach; Distributed in USA by Howmedica Osteonics Corp., 325 Corporate Drive, Mahwah, NJ 07430. This product is sold separately or as part of Kit # 990256. The depth gauge is an instrument in the VariAx Elbow System. Depth gauges are tools used intra-operatively, to aid in correct screw selection. They can be used with a plate, or directly on bone.

    The Depth Measuring Gauges have 5 scale marking instead of the required 4 scale markings.

  • TerminatedUnder Investigation by firm

    DVR Crosslock ePAK Depth Gauge, Model Number 212000003 Product Usage: These DVR Express Disposable Instruments are used to facilitate implantation of orthopedic medical devices in trauma.

    There is a potential for weak seals of the sterile packaging.

Monitor gauge, depth automatically

DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.