HSTClass 1

Apparatus, Traction, Non-Powered

FDA adverse event reports and recall history for apparatus, traction, non-powered (product code HST), from the public MAUDE corpus and the live FDA recall database.

Total reports
520
Deaths
3
Injuries
88
Malfunctions
422

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Recent recalls

  • 2024-03-20Open, ClassifiedUnder Investigation by firm

    HDS Traction Boot II, REF AR-6529-23, The Traction Boot is designed to position, support and/or distract the patient's foot for hip surgery

    Baxter Healthcare Corporation is issuing an Urgent Medical Device Recall for the Hip Distraction System, Traction Boot II, due to the boots being assembled incorrectly. The locking ridge on the front face of the boot key socket is facing upwards, which prevents the socket mechanism on the traction boot from attaching properly to the spar of the traction system. This will cause the boot to detach from the traction system.

  • 2017-02-14TerminatedDevice Design

    Arthrex Shoulder Suspension Tower; Catalog Number: AR-1650-01 Product Usage: The Arthrex Lateral Decubitus Shoulder Positioner is intended to allow for the controlled manipulation of the arm and shoulder joint during arthroscopic surgical procedures.

    During use, there is a potential scenario which could allow the fastening screws attaching covers to come loose and potentially land in the sterile field causing a delay in surgery.

  • 2015-01-20TerminatedLabeling False and Misleading

    Fitness Gear Inversion table (Style STE00059FG or Style STE00118FG), Distributed by ASLI, Wilmington, DE (on carton), Paradigm Health & Wellness, Inc. (in user manuals); The products are inversion tables, which allow the user to secure his or her ankles to the table and invert. The products were intended to be fitness equipment; i.e. consumer products.

    Reports of customers falling from the Inversion Table and sustaining injuries. The Inversion Table is a fitness product, not a medical device. DSG removed medical claims from product labels and advertising.

  • 2014-02-20TerminatedNonconforming Material/Component

    Gelzone Shoulder Sleeve, Catalog Number SLV634, Lot Number 852766. Product Usage: To be used where limb support and/or elevation is considered necessary.

    Implantech initiated the voluntary recall of Gelzone Shoulder Sleever, SLV-634, Lot Number 852766 because it may not satisfy customer requirements for the robustness of the Velcro closure.

  • 2013-07-22TerminatedComponent change control

    STERIS Fine Traction Device, one unit per box Product Usage: The Fine Traction Device is used to provide traction and enable fine adjustments to traction and orientation of the leg during orthopedic procedures.

    A bearing used in the assembly of the Fine Traction Device did not perform successfully in pre-release reliability testing; a replacement bushing passed reliability testing and was substituted per a change order. The change order authorized use of the substitute bushing in assembly of the Fine Traction Accessory; however, inadvertently the new part was not implemented. During complaint evaluati

  • 2013-06-14TerminatedDevice Design

    Zimmer Patient Helper Bed Frame Adapter Assembly The bed frame adapter assembly is intended to be used by qualified medical professionals to adapt a patient helper to a hospital bed. The patient helper is used to aid and assist the patient's mobility within a hospital bed and during bed ingress and egress.

    The bed frame adapter assemblies may have a weak weld where the patient helper adapter mounting bars attach to the tube.

  • 2013-03-18TerminatedProcess control

    Allen Traction Boot, Modular, OT 1000 Traction Device Boot Assembly, Fracture Table Orthopedic Leg Holders, non-invasive medical devices designed to attach to orthopedic fracture tables to aid clinical staff in applying traction to the leg of a patient during surgical procedures.

    Some units may have been assembled incorrectly which could result in the boot breaking free from the mounting base and allowing the patient's leg to fall.

  • 2010-05-14TerminatedNonconforming Material/Component

    Stryker Endoscopy PrepStrap Shoulder Sleeve, Model Number 3910900200, Manufactured by Stryker Endoscopy, San Jose, CA

    Latex-- Certain lots of Product contain undeclared latex.

  • 2009-10-21TerminatedProcess control

    Arthro-Knife, ConMed, Linvatec 11311 Concept Blvd, Largo FL 33773, Made in USA. RX only. U.S. PAT NO. 5.141.517. The Arthro-Knife is a one piece disposable assembly comprised of a plastic handle with a retractable blade, sheathed blade. These knives are supplied sterile, single-use. Each knife assembly contains a color coded sliding retractor that coincides with a specific blade type. The blade can be extended and locked to a half extended position or fully extended position.

    ConMed has determined there is a possibility that the device may not fully retract once it is deployed.

  • 2005-08-31TerminatedOther

    TM-300 Traction System; an Rx software driven device used for patient traction; Manufactured for Sammons Preston Roylan, An Ability One Company, Bolingbrook, IL 60440, Manufactured by: ITO Co., Ltd., 3-3-3 Toyotama-Minami, Nerima-ku, Tokyo, 176-8605, Japan

    The traction device may malfunction, producing oscillation and a jerking motion.