Microscope, Operating & Accessories, Ac-Powered, Ophthalmic
FDA adverse event reports and recall history for microscope, operating & accessories, ac-powered, ophthalmic (product code HRM), from the public MAUDE corpus and the live FDA recall database.
- Total reports
- 374
- Deaths
- 0
- Injuries
- 65
- Malfunctions
- 301
Reports per year
Source: FDA adverse event reports by year received, 2008–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.
Recent recalls
- 2022-12-09Open, ClassifiedProcess control
Leica MICROSYSTEMS Proveo 8, Part Number 10449058, surgical microscope system
There is a potential issue resulting in the Power Supply potentially becoming overheated due to internal components overstressing. Consequently, due to the overstressing of the internal components, one specific component can fail which will cause the device to unexpectedly shutdown.
- 2019-02-27TerminatedDevice Design
Alcon or Endure Ophthalmic Microscope Product Usage: Used for low magnification visualization during ophthalmic surgical procedures for cataract, retina and cornea.
Potential for the optical head to detach from the stand due to the lack of thread adhesive.
- 2015-08-21TerminatedDevice Design
LX3 Floor Stand for Endure Luxite Envision Optical System; Endure LuxOR Q-Vue. Product Usage: The LX3 Microscope Floor Stand is an accessory to the LuxOR Opthalmic Microscope and the LuxOR Opthalmic Microscope with Q-VUE 3-D Assistant Visualization and provides a portable suspension system for the microscope that integrates an AC power module and microscope controls (Primary Control Panel and connectivity of foot controllers)
Alcon is initiating a voluntary recall of all LX3 Microscope Floor Stands after receiving reports concerning loss of functionality due to issues with the cabling between the power, host and illumination modules, potentially causing the system to display a system message, and/or resulting in illumination failure.
- Open, ClassifiedSoftware design
Haag Streit Surgical Floor stands FS 2-11 (Ref no. 615H511) and FS 2-15 (Ref no. 615H515) in combination with the operating microscopes HS Hi-R NEO 900 (Ref no. 657 820) and HS Hi-R NEO 900A (Ref no. 657 821), with software REF 615 588 versions 2.0 to 3.3 - Product Usage: use in ophthalmology and in other disciplines with vertical direction of view
Software error -Software REF 615 588 versions 2.0 to 3.3, movement of the focusing is triggered by pressing the button of the footswitch it might occur that the movement does not stop when releasing the button, result in potential risk to the patient's eye
- TerminatedProcess control
Leica M220 F12 surgical microscope The Leica M220 F12 surgical microscope is an optical instrument for improving the visibility of objects through magnification and illumination. It can be applied for observation and documentation and for human and veterinary medical treatment.
The M220 optics may unintentionally drop into the surgical field, risking contact with the patient.
- TerminatedOther
PROVEO 8 Ophthalmic Microscope System optical instrument for improving the visibility of objects through magnification and illumination. It can be applied for observation and documentation and for human and veterinary medical treatment.
The observed issues can lead to unexpected interruptions of the optics carrier focus drive movement, intermittent stops of the X-Y coupling movement or sporadic failures in the proper on-screen display functionality of the HDR recording unit. Additionally, an anomaly in the device software may also lead to the unexpected behavior that when a user releases a handle or foot switch button this may not immediately end the actuated functionality. This behavior can only occur in unlikely workflows where a user changes a user profile or a mode change is initiated at the same time as a functional button is activated.
- TerminatedPackaging process control
LX3 Floor Stand for LuxOR Ophthalmic Microscope; Endure LuxOR Q-VUE Ophthalmic Microscope. LX3 Microscope Floor Stands are not stand-alone medical devices, but are used in combination with the LuxOR Ophthalmic Microscope and the LuxOR Ophthalmic Microscope with Q-VUE 3-D Assistant Visualization. These microscopes are used for low magnification visualization during cataract, retina, and cornea surgeries
Multiple shipments or forceful impact to a caster on a LuxOr/LX3 microscope may stress the microscope base, which could contribute to a caster breaking or detaching from the base.
Monitor microscope, operating & accessories, ac-powered, ophthalmic automatically