Cannula, Ophthalmic
FDA adverse event reports and recall history for cannula, ophthalmic (product code HMX), from the public MAUDE corpus and the live FDA recall database.
- Total reports
- 819
- Deaths
- 0
- Injuries
- 272
- Malfunctions
- 531
Reports per year
Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.
Recent recalls
- 2025-10-10Open, ClassifiedProcess control
Cyclopen Microinterventional Cyclodialysis System. Model Number: CP-300. Product Description: The CycloPen Microinterventional Cyclodialysis device is a manual surgical instrument used for the ab-interno construction or modification of a cyclodialysis, followed by delivery of viscous materials, bio-tissue/allograft, or other materials to reinforce and maintain the cleft.
Potential for compromised integrity of the sterile packaging.
- 2024-03-14Open, ClassifiedUnder Investigation by firm
BVI Visitec Irrigation Handle (20/SP)- Intended to be fit with a cannula for irrigation of the eye Catalog Number: 581012
Pack and pouch label is missing the manufacturing and expiration date.
- 2022-08-26Open, ClassifiedEnvironmental control
BEAVER-VISITEC INTERNATIONAL, INC. Visitec CANNULA, LASIK IRR (10/BX)BVRVIS Catalog #585278
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- 2015-06-19TerminatedUse error
Steriseal-Cannula Ophthalmic; Bi-Manual Irrigating/Aspirating Handpiece.
The metal cannula attached to the handpiece can rotate and become dislodged from the plastic handpiece.
- 2014-09-04TerminatedUnder Investigation by firm
iTRACK250A Canaloplasty Microcatheter Kit; Catalogue number: iT-250A; Rx only, sterile. Indicated for fluid infusion and aspiration during surgery.
One lot may not have been properly sealed, resulting in a non-sterile device.
- 2014-05-27TerminatedPackaging process control
Visitec Anterior Chamber Cannula, .40x22mm (27Gx7/8in) REF 585006 Cannulae are intended to irrigate and or aspirate ophthalmic solutions and liquids in or out of the eye, and remove materials where appropriate. These products are used in all sections of the eye. This cannula is a general use cannula mainly designed for the precise maintenance of optimum anterior chamber depth
Contains a sharp tip needle when the product should contain a blunt tip cannula
- 2013-10-22TerminatedLabeling Change Control
Synergetics 25 gauge Soft Tip Cannula, packaged in 6 units per box. Catalog number: 30.08.25. Single-use 25 gauge cannula used in vitreoretinal surgery.
One lot of 25 gauge Soft-Tip Disposable Cannulas are co-mingled with 23 gauge cannulas; the product is labeled as 25 gauge cannulas.
- 2013-05-10TerminatedUse error
Insight Instruments MIS ( Multi-Port Illumination System) Illuminated pilot tubes work like a ceiling light while simultaneously protecting the entry sites. Illumination automatically follows the movement of the two surgical instruments inserted into the M.I.S. cannulas. The large distance of more than 16mm between cannula ends and posterior pole guarantees sufficient light in the working area and a large observation field especially when used in conjunction with modern wide field panoramic observation systems.
Insight Instruments, Inc., Stuart, FL is recalling M.I.S. (Multi-Port Illumination System), Ophthalmic Cannula due to a misclassification of the medical device.
- 2013-01-17TerminatedNonconforming Material/Component
Bausch and Lomb 27G sterile cannula packed in the Bausch and Lomb Amvisc 1.2% Sodium Hyaluronate ( Model 59051, 59081, 59051L, 59081L) and Amvisc Plus 1.6% Sodium Hyaluronate (Model 60081, 60051, 60051L, 60081L) ophthalmic viscosurgical device. The 27G disposable cannula is attached to the syringe standard luer fitting to deliver the Amvis or Amvis Plus during use as a surgical aid in opthalmic anterior or posterior segment procedures.
Some disposable cannulas (with 4-digit lot numbers) provided with the Amvisc or Amvisc Plus OVD are not properly engaging to the Luer-Lock on the Amvisc and Amvisc Plus sterile glass syringes. The cannulas may leak viscoelastic material or detach from the syringe during injection. In rare incidents, detachment has resulted in serious patient injury.
- 2011-05-10TerminatedComponent change control
BD Visitec High Viscosity Injector 4mm .90 (20G) [Hammer] REF 585173 High Viscosity Injectors are used during retinal surgery for fluid infusion. A High Viscosity Injector is placed into the sclera and temporarily sutured into place.
Restricted or no fluid flow for Visitec High Viscosity Injector.
Monitor cannula, ophthalmic automatically
DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.