Skip to main content
HKOClass 1

Refractometer, Ophthalmic

FDA adverse event reports and recall history for refractometer, ophthalmic (product code HKO), from the public MAUDE/AEMS corpus and the live FDA recall database.

Total reports
169
Deaths
0
Injuries
52
Malfunctions
116

Reports per year

10
11
12
13
14
15
16
17
18
20
21
23
24
25
26

Source: FDA adverse event reports by year received, 20102026. Reporting paused during the 2026 MAUDE-to-AEMS transition and has since resumed.

Recent recalls

  • 2026-03-18Open, ClassifiedDevice Design

    Brand Name: Welch Allyn, Inc. Product Name: Welch Allyn Spot Vision Screener VS100 Model/Catalog Number: VS100 Software Version: Software version 3.2.0.1 Product Description: Welch Allyn Spot Vision Screener VS100, FDA product code name: Refractometer, ophthalmic One unit per package

    Display screen may unintentionally flicker due to a software issue. Exposure to flickering lights may induce photosensitive seizure activity in susceptible individuals.

  • 2015-08-27TerminatedSoftware design

    OPD-Scan III Refractive Power/Corneal Analyzer Opthalmic Software versions 1.00.08 1.05.07 1.10.01, 1.01.02 1.06.02 1.11.02, 1.02.01 1.07.01 1.12.03, 1.03.02 1.08.01 1.13.01, 1.04.03 1.09.01. Opthalmic: The OPD-Scan III is a diagnostic instrument that is indicated for use for Mapping of refractive error distribution of the eye by measurement and analysis of spherical power, cylindrical power, and cylinder axis.

    Software bug was found where there was no difference in Total and Corneal high-order aberrations, but differences were found in Internal high-order aberrations.

  • 2013-05-25TerminatedSoftware design

    Canon Full Auto Ref-Keratometer RK-F2, Catalog Number 6937B002AA. Product Usage: This device is intended to be used to measure the refractive power and the radius of corneal curvature of the human eye.

    Canon Inc. had received 17 claims from our foreign sales companies between October to November 2012 that measurement value of RK-F2s refractive power (SPH value) sometimes indicated abnormal value which occurs approximately 5% in probability.

  • TerminatedNonconforming Material/Component

    Carl Zeiss Meditec AG VISUREF150 REF 2227-967 - Product Usage: is a diagnostic instrument used to determine the objective refractive values and keratometric characteristics of the human eye. The results are used to assist the process of prescribing optical aids such as eyeglasses and contact lenses.

    Due to the incorrect DC/DC converter being assembled onto the communication board. As a result the effectiveness of the communication board cannot be guaranteed in that the safety function of the board would be insufficient resulting in an increased leakage of current and potentially cause electrical shocks to the user.

  • TerminatedEquipment maintenance

    AUTO REF/KERATOMETER ARK-1s

    There is possibly a difference between the measured dioptric value of the subjective measurement mode and the objective measurement mode on ARK-1s.

Monitor refractometer, ophthalmic automatically

DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.