FDA adverse event reports and recall history for hysteroscope (and accessories) (product code HIH), from the public MAUDE/AEMS corpus and the live FDA recall database.

Total reports
8,246
Deaths
41
Injuries
1,606
Malfunctions
6,546

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 20122026. Reporting paused during the 2026 MAUDE-to-AEMS transition and has since resumed.

Recent recalls

  • 2026-02-20Open, ClassifiedUnder Investigation by firm

    Olympus Inner Sheath. Model No. A4741. for endoscopic diagnosis and treatment in gynecological applications.

    Complaints of the ceramic tip of the resection sheath breaking have been received.

  • 2025-07-21Open, ClassifiedPackaging change control

    Blue Silicone, Reusable Silicone Seal 6 - 12 FR. Model Number: CS-B612. The reusable silicone seals fit over the biopsy port of endoscopes to seal off the working channel. Some contain a lumen which allows instruments to be passed through the seal and minimize or prevent fluid leakage during endoscopic procedures.

    Discrepancies between sterilization instructions in the Japanese language product package inserts and those published in the manufacturer product Instructions For Use (IFU).

  • 2025-07-21Open, ClassifiedPackaging change control

    Green Silicone Seal for Accessories, Reusable Silicone Seal 0 - 7 FR. Model Number: CS-G7. The reusable silicone seals fit over the biopsy port of endoscopes to seal off the working channel. Some contain a lumen which allows instruments to be passed through the seal and minimize or prevent fluid leakage during endoscopic procedures.

    Discrepancies between sterilization instructions in the Japanese language product package inserts and those published in the manufacturer product Instructions For Use (IFU).

  • 2024-06-06Open, ClassifiedProcess change control

    OES 4000 Hysteroscope, Model/Catalog Number: A4674A

    The required leakage test was not performed following of a re-work for a non-conformity. This could potentially result in liquid ingress into the eyepiece and cause a foggy image.

  • 2023-11-17Open, ClassifiedUnder Investigation by firm

    Hysteroscope-For endoscopic diagnosis within the uterus Model Number: HYF-1T.

    Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

  • 2023-11-17Open, ClassifiedUnder Investigation by firm

    Hysterovideoscope-For endoscopic diagnosis within the uterus Model Number: HYF-V.

    Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

  • 2023-08-29Open, ClassifiedReprocessing Controls

    Visera Hysterovideoscope Olympus HYF Type V

    IFU labeling is being updated to revise EtO gas mixture and elimination of reusable cleaning brushes.

  • 2023-08-09Open, ClassifiedProcess control

    UroLift Visual Obturator, 20Ch-The visual obturator is used with the UroLift system. Instrument for insertion of a cystoscope shaft into the bladder. Model: UL-VO

    There is an improperly performed weld between the cone of the sheath lock and tube of the obturator; mechanical load capacity is impaired, and fracture of the welded joint may occur. Contamination can occur in this capillary gap, which is not covered by the reprocessing validations performed.

  • 2019-06-13TerminatedUnder Investigation by firm

    E-Line Cutting Electrode 11.5Fr Mono 0 degrees, Sterile. One electrode is placed in a plastic tray and then inserted into a Tyvek pouch. Five packaged electrodes are placed in a white box. Product Usage: Resectoscopes are used for endoscopically controlled ablation of tissue. They are used, in combination with endoscopic accessories, for examination, diagnosis, and/or therapy in various medical disciplines such as urology and gynecology.

    E-Line Cutting Electrode package 8416.0305 contains E-Line Hook Electrode 8416.09 instead of E-Line Cutting Electrode 8416.03.

  • 2005-02-02TerminatedNonconforming Material/Component

    VERSAPOINT, Hysteroscopic system Connector Cable, Non-Sterile, Reusable (20 uses only). US Distribution/Manufactured for GYNECARE, A division of Ethicon, Inc., a Johnson & Johnson Company, Somerville, NJ 08876. The product subject of this recall is the connector cable which goes between the electrosurgical generator and the elctrode used for surgery. Specifically, the recall focuses on the hand grip on the electrode, formed by overmolding a polymer onto the cable, which deforms upon sterilization.

    Deformation of the Verspoint Hysteroscopic System Connector Cable hand grips when autoclaved causing possible non-sterile electrodes.

Monitor hysteroscope (and accessories) automatically

DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.