HBLClass 2

Holder, Head, Neurosurgical (Skull Clamp)

FDA adverse event reports and recall history for holder, head, neurosurgical (skull clamp) (product code HBL), from the public MAUDE/AEMS corpus and the live FDA recall database.

Total reports
5,812
Deaths
5
Injuries
1,660
Malfunctions
3,729

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 20122026. Reporting paused during the 2026 MAUDE-to-AEMS transition and has since resumed.

Recent recalls

  • 2026-01-08Open, ClassifiedDevice Design

    Brand Name: HFD100 Product Name: Head Fixation Device HFD100 Model/Catalog Number: 119695-000, 119629-000, 119630-000, 113803-000, 113803-600, 113802-000, 113802-600 Software Version: Not Applicable Product Description: The HFD100 Head Fixation Device provides rigid skeletal fixation within the optimal imaging envelope while introducing minimal artifacts in the acquired images. There are two different assemblies for the HFD100. The standard HFD100 is equipped with a table adapter for use with either the operating room table or the angiography table. For sites with a Hill-Rom TruSystem 7500 Hybrid Neuro MR Operating Table (ORT400), the HFD100 has a split block clamp assembly (Trumpf Mount Clamp Assembly) which connects directly to the table back section. The HFD100 system is also comprised of the linkage components to mount the skull clamp to the table adapter, and the 3-pin skull clamp. The linkage system is used to position the skull clamp in the necessary position, without colliding with the magnet bore. Component: Not a component.

    Torque screw may crack or separate. If used, an affected screw could lead to inadequate head stabilization, compromised surgical precision, procedural delays, requirement for alternate head fixation methods, and risk of patient injury.

  • 2025-05-28Open, ClassifiedDevice Design

    DORO Easy-Connect Navigation Adaptor; BRAINLAB (item No. 1204.002)

    Potential of compromised compatibility resulting in influence on the positioning of the reference array and an influence of the function of the navigation system cannot be excluded.

  • 2018-09-17TerminatedComponent change control

    DORO LUCENT Base Unit, REF 1101.021 Headrest System used in open and percutaneous craniotomies and spinal surgeries

    The combination of two different design revisions of DORO LUCENT Base Unit, Locking Transitional Member, and Transitional Member within one set up might lead to a malfunction and failure of the headrest system.

  • 2018-09-17TerminatedComponent change control

    DORO LUCENT Locking Transitional Member, REF 1101.031 Component of headrest system used in open and percutaneous craniotomies and spinal surgeries

    The combination of two different design revisions of DORO LUCENT Base Unit, Locking Transitional Member, and Transitional Member within one set up might lead to a malfunction and failure of the headrest system.

  • 2018-09-17TerminatedComponent change control

    DORO LUCENT Transitional Member, REF 1101.026 Component of headrest system used in open and percutaneous craniotomies and spinal surgeries

    The combination of two different design revisions of DORO LUCENT Base Unit, Locking Transitional Member, and Transitional Member within one set up might lead to a malfunction and failure of the headrest system.

  • 2015-07-24TerminatedOther

    Medical Device Exchange - Mayfield Disposable Skull Pin; Pouch, Tyvek/Mylar, Sterile, Single Use Only; OEM Catalog number A1072; EXP (repackager), Integra (OEM) Holder, head, neurosurgical (skull clamp)

    EXP did not register as a medical device establishment, list the devices being recalled, or establish its own quality system and instead relied on the fact that its vendors were registered and had their own quality systems.

  • 2013-11-13TerminatedDevice Design

    Composite Series Skull Clamp, one clamp per tote. The MAYFIELD Skull Clamp (A3059) is a cranial stabilization device, designed to provide rigid skeletal fixation. The Skull Clamp is placed on the patient's skull to hold their head and neck securely in a particular position when rigid fixation is desired. The clamp is indicated for use in open and percutaneous craniotomies as well as spinal surgery when rigid skeletal fixation is necessary

    An investigation of an adverse trend of complaints for the Skull Clamp index locking knob not easily unlocking or disengaging and malfunctions of the 80 lbs. torque knob was initiated. None of the complaints resulted in a report of patient injury, adverse health consequence, or a delay / prolongation of the surgical procedure.

  • 2012-12-19TerminatedUnder Investigation by firm

    MAYFIELD Composite Series Skull Clamps, Model A3059, 510(k) #K120633. The MAYFIELD Skull Clamps (A3059) are indicated for use in open and percutaneous craniotomies, as well as spinal surgery when rigid skeletal fixation is necessary.

    An investigation of an adverse trend of complaints conducted by Integra LifeSciences identified various product performance issues which could impact the functionality of the firm's Mayfield Composite Series Base Units and Mayfield Skull Clamps. In order to prevent additional complaint reports, the firm decided to recall and replace certain lot codes of the Base Units and Skull Clamps that were d

  • 2004-07-27TerminatedOther

    MAYFIELD Radiolucent Disposable Adult Skull Pins (Sterile), Model 4-0-A-2020

    Skull pins, which fractured below device specifications during QC testing, were distributed.

  • TerminatedDevice Design

    IMRIS MR/X-ray head fixation device, Model HFD200

    Breaking of the HFD200 Drive Knob force indicator due to over-torqueing of the drive knob beyond the maximum allowable pinning force specification may result in the clinical staff unknowingly pinning the patient's skull with insufficient force; the clinical staff unknowingly pinning the patient's skull with excessive force; or the clinical staff identifying the issue and not being able to continue with the procedure.

Monitor holder, head, neurosurgical (skull clamp) automatically