Electrode, Depth
FDA adverse event reports and recall history for electrode, depth (product code GZL), from the public MAUDE corpus and the live FDA recall database.
- Total reports
- 228
- Deaths
- 3
- Injuries
- 61
- Malfunctions
- 151
Reports per year
Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.
Recent recalls
- 2024-10-24CompletedEmployee error
AD-TECH Spencer Probe Depth Electrode, REF SD04R-AP58X-000 4-Contact 1.12mm (A style) Depth Electrode
Ad-Tech is recalling these A-Style SD Depth Electrodes because a label discrepancy was identified on A style contact SD electrodes where the MR Conditional symbol was incorrectly placed instead of the MR Unsafe symbol. The A style contacts were not cleared for MR Conditional use. Our records indicate you have received one or more of the affected lots listed above.
- 2024-10-24CompletedEmployee error
AD-TECH Spencer Probe Depth Electrode, REF SD06R-AP58X-000 6-Contact 1.12mm (A style) Depth Electrode
Ad-Tech is recalling these A-Style SD Depth Electrodes because a label discrepancy was identified on A style contact SD electrodes where the MR Conditional symbol was incorrectly placed instead of the MR Unsafe symbol. The A style contacts were not cleared for MR Conditional use. Our records indicate you have received one or more of the affected lots listed above.
- 2024-10-24CompletedEmployee error
AD-TECH Spencer Probe Depth Electrode, REF SD08R-AP58X-000 8-Contact 1.12mm (A style) Depth Electrode
Ad-Tech is recalling these A-Style SD Depth Electrodes because a label discrepancy was identified on A style contact SD electrodes where the MR Conditional symbol was incorrectly placed instead of the MR Unsafe symbol. The A style contacts were not cleared for MR Conditional use. Our records indicate you have received one or more of the affected lots listed above.
- 2019-05-06TerminatedDevice Design
Neuro Omega System, incorporating HaGuide software. For neurological and neurosurgical use. To monitor, record and display bioelectric signals, and stimulate nerves and brain tissue.
If cables are improperly connected, current may reach high charge density, causing tissue harm.
- 2014-10-21TerminatedProcess control
Box Label: 3 mm Cueva Electrode Kit, Sterile EO, Rx Only, REF. CNE1-3KDINX . Pouch labels: Each kit contains 3 pouches. 3 mm Cueva Electrode CNE-3X, Cueva Electrode Applicator Wand CNE-100X, Cueva Electrode Leadwire CNE-LW-2DINX. Box Label: 2 mm Cueva Electrode Kit, Sterile EO, Rx Only, REF. CNE1-2KDINX . Pouch labels: Each kit contains 3 pouches. 3 mm Cueva Electrode CNE-2X, Cueva Electrode Applicator Wand CNE-100X, Cueva Electrode Leadwire CNE-LW-2DINX.
The recall has been initiated due to concerns that the applicator wand may malfunction potentially causing the inability to monitor the VII cranial nerve. .
- 2010-03-16TerminatedSoftware change control
FHC microTargeting Guideline 4000 ver 1.4 (GL4K) Software, Catalog # MT-GL4K. Intended to assist in functional neurological procedures.
The raster tab for the GL4K version 1.4 software may assign incorrect depths for waveform recordings and if used to identify the implantation target for the DBS lead, the implant may be placed in the wrong location.
- 2009-07-07TerminatedPackaging process control
Cyberkinetics Inserter Wand & V2 Trigger Assembly, Expiration: July 14, 2009; Catalog Number: 4468.
Sterility of product cannot be assured
- 2008-12-18TerminatedPending
FHC MicroTargeting Electrode, MTDWBP(AR)(MP I ) Intended for use in intra-operative recording of single unit neuronal activity or intra-operative stimulation of neural elements in the brain.
Mislabeled: May contain an electrode longer than labeled size
- 2008-12-18TerminatedPending
FC4000 MicroTargeting Electrode Kit 5x For use with Nexframe and Nexdrive Distributed by Medtronic Inc., Minneapolis, MN REF: FCH 4000 (5 Prefilled single use kits -5 packages and 2 electrodes each) Intended for use in intra-operative recording of single unit neuronal activity or intra-operative stimulation of neural elements in the brain.
Mislabeled: May contain an electrode longer than labeled size
- 2008-12-18TerminatedPending
FC400SP microTargeting StimPilot Single Procedure Kit Distributed by Medtronic Inc., Minneapolis, MN Ref: FHC4000SP Intended for use in intra-operative recording of single unit neuronal activity or intra-operative stimulation of neural elements in the brain.
Mislabeled: May contain an electrode longer than labeled size
Monitor electrode, depth automatically