Skip to main content
GKLClass 2

Counter, Cell, Automated (Particle Counter)

FDA adverse event reports and recall history for counter, cell, automated (particle counter) (product code GKL), from the public MAUDE corpus and the live FDA recall database.

Total reports
215
Deaths
0
Injuries
13
Malfunctions
102

Reports per year

06
07
08
09
11
12
13
16
17
18
20
21
22
23
25

Source: FDA adverse event reports by year received, 2006–2025. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Recent recalls

  • 2015-01-20TerminatedProcess control

    CD8 (Leu-2a) PE (In vitro diagnostic), Catalog number 340046. Hematology: Single color direct immunofluorescence reagent for enumerating percentages of mature human suppressor/cytotoxic (CD8+) lymphocytes in erythrocyte-lysed whole blood or peripheral blood mononuclear cell suspensions.

    One lot of CD8 PE is contaminated with CD4 FITC and may cause an unexpected staining pattern and incorrect patient data.

  • 2013-08-28TerminatedEquipment maintenance

    ADVIA 120 Hematology Analyzer, Catalog Numbers SMN 10360958 and 10360959 . The ADVIA 120 is a hematology system that utilizes the principles of flow cytometry in order to provide complete blood counts.

    The ADVIA 120 optics cover or hood is located on the very top of the ADVIA 120 analyzer. When the optics cover is raised, it is supported by two gas struts (SMN 10309266) attached to the cover on the left and right sides. Over time, it is possible that these struts may lose their effectiveness and fail to support the optics cover in the open position. This could cause the cover to fall during maintenance procedures.

  • 2011-04-12TerminatedSoftware design

    Coulter GEN's The COULTER LH 500 Analyzer is a quantitative, automated hematology analyzer and leukocyte differential cell counter for In Vitro Diagnostic Use in clinical laboratories. The LH500 Analyzer also provides a semi-automated reticulocyte analysis.

    This recall was initiated because Beckman Coulter identified a problem with some date fields that are transmitted to the Laboratory Information System (LIS). Starting Jan 1, 2011, the date fields will transmit as a blank field to the LIS if the year is greater than 2010. The following dates are affected: Run date, Control expiration date, Date of birth (For individuals born in 2011 or later), a

  • 2007-12-22TerminatedSoftware design

    COULTER GEN*S System, Part Number: 6605381, Beckman Coulter, Brea, CA. (Hematology analyzer)

    Patient Mis-identification can occur: 1)- When a positive identifier (Sample ID or Cassette/Position) is manually edited to a positive identifier that is already in the To Do list. The workstation will accept the entry and no error message will be generated, creating 2 samples with the identical positive identifier. 2)- When manually entering a Patient ID, if a blank space is entered between th

  • 2006-08-04TerminatedOther

    ADVIA 2120 systems -Automated Complete Blood Cell and Differential Cell Counter

    The ADVIA 2120 has reported ; highly intermittent low results on all primary results: White Blood Cells (WBC), Red Blood Cells (RBC), Hemoglobin (HGB) and Platelets (PLT) and an abnormal baso cytogram.

  • 2004-07-20TerminatedOther

    CXP software, Part No. 623560, version 1.0

    Software anomaly. Sample ID and the Run Date may become fixed within a header on the FlowPAGE printout. All samples subsequently run, with results printed using the FlowPAGE print format, will have the fixed Sample ID and Run Date, rather than the correct Sample ID and Run Date.

  • 2004-07-20TerminatedOther

    Cytomics MXP Software, Version 1.0, Part No. 623688.

    Software anomaly. Sample ID and the Run Date may become fixed within a header on the FlowPAGE printout. All samples subsequently run, with results printed using the FlowPAGE print format, will have the fixed Sample ID and Run Date, rather than the correct Sample ID and Run Date.

  • 2004-07-20TerminatedOther

    Cytomics RXP Software, Version 1.0, Part Numbers: 175488 6418489 175260 175261 175262 175263 175264 175265

    Software anomaly. Sample ID and the Run Date may become fixed within a header on the FlowPAGE printout. All samples subsequently run, with results printed using the FlowPAGE print format, will have the fixed Sample ID and Run Date, rather than the correct Sample ID and Run Date.

  • TerminatedDevice Design

    GloCyte RBC Reagent - Product Usage: The GloCyte Automated Cell Counter System is intended for use by trained healthcare professionals in clinical laboratories to provide quantitative determination of fluorescence labeled total nucleated cells and erythrocytes in cerebrospinal fluid collected from adult and pediatric patients. The GloCyte Low and High Level Controls are assayed hematology controls designed to monitor the performance of the GloCyte Automated Cell Counter System. Assayed parameters include total nucleated cells and erythrocytes.

    RBC Reagent failure. High fluorescent counts during RBC Reagent + Diluent QC check that will result in failed QC check, preventing user from testing patient samples for RBC.

  • TerminatedUnder Investigation by firm

    BD-Retic-Count Reticulocyte Reagent System; Hematology: Used for in vitro diagnostics

    The carton label for one lot of BD Retic-Count Reticulocyte Reagent System had the incorrect storage conditions. The labels on the reagent as well as the Instructions for Use are correct.

Monitor counter, cell, automated (particle counter) automatically

DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.