GIFClass 1

Diluent, Blood Cell

FDA adverse event reports and recall history for diluent, blood cell (product code GIF), from the public MAUDE corpus and the live FDA recall database.

Total reports
144
Deaths
0
Injuries
7
Malfunctions
136

Reports per year

94
97
06
10
11
13
14
15
21
22

Source: FDA adverse event reports by year received, 1994–2022. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Recent recalls

  • 2026-01-30Open, ClassifiedUnder Investigation by firm

    COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conjunction with a cyanide-free lytic agent for counting and sizing blood cells on all UniCel DxH Coulter Cellular Analysis Systems. For use as a rinsing agent on the UniCel DxH Slidemaker Stainers. For use as a diluent in conjunction with DxH ECO Cell Lyse for counting and sizing blood cells. For use as a rinsing agent.

    Due to specific lots of diluents contributing to elevated platelet Daily Checks background counts cycle with use with analyzers.

  • 2023-07-12Open, ClassifiedUnder Investigation by firm

    COULTER DxH Diluent, REF 628017; diluent used for counting and sizing blood cells

    Specific diluent lot numbers showed that conductivity, osmolality, and pH were out of manufacturing specifications.

  • 2015-08-28TerminatedNonconforming Material/Component

    COULTER LH Series Diluent, Catalog No. 8547194, Product Labeling A40192-AD. The diluent provides the ability to analyze portions of the diluted blood sample for different blood cell types, such as red blood cells and platelets.

    Beckman Coulter is recalling the LH 750 and LH780 Hematology Systems because the use of the diluent lots may eventually result in a compromised Hgb cuvette.

  • 2015-08-28TerminatedNonconforming Material/Component

    COULTER ISOTON 4 Diluent, Catalog No. 8547148. The diluent provides the ability to analyze portions of the diluted blood sample for different blood cell types, such as red blood cells and platelets.

    Beckman Coulter is recalling the LH 750 and LH780 Hematology Systems because the use of the diluent lots may eventually result in a compromised Hgb cuvette.

  • 2015-05-15TerminatedNonconforming Material/Component

    COULTER DxH Diluent, Catalog No. 628017, Part No. A59956. For use as an isotonic buffered diluent in conjunction with a cyanide-free lytic agent for counting and sizing blood cells.

    Beckman Coulter has received an increased number of complaints related to Hemoglobin (HGB) on UniCel DxH 800 and DxH 600 Coulter Cellular Analysis Systems. The use of the DxH Diluent lots listed in the recall notice may eventually result in a compromised HGB chamber.

  • 2014-09-16TerminatedUnder Investigation by firm

    Coulter DxH Diluent, Catalog No. 628017 Blood cell analysis comprises diluting a whole-blood sample with a solution that functions as a diluent. The diluent provides the ability to analyze portions of the diluted blood sample for different blood cell types and the measurement of hemoglobin.

    Beckman Coulter is recalling Coulter DxH Diluent because it may be associated with elevated platelet (PLT) background counts between 3-7 x 10(3) cells/uL. This will result in failed Daily Checks on the UniCel DxH 800 and DxH 600 Coulter Cellular Analysis Systems.

  • 2007-04-03TerminatedOther

    CELL-DYN 3200 Diluent/Sheath Reagent, List Number 03H79-01,Distributed and Manufactured by: Abbott Diagnostics Division, 5440 Patrick Henry Drive, Santa Clara, CA

    Elevated platelet background count results-some products may show a higher than expected platelet background count when used on the CELL-DYN 3200 System and report patient results that are unacceptable-out-of-range

  • 2007-04-03TerminatedOther

    CELL-DYN Sapphire and CELL-DYN 4000 Diluent/Sheath Systems- List No. 01H73-01, Distributed and Manufactured by: Abbott Diagnostics Division, 5440 Patrick Henry Drive, Santa Clara, CA

    High platelet background count when using the CELL-DYN 4000 System and/or CELL-DYN Sapphire System; with patient results unacceptable -out-of-range.

  • 2007-03-17TerminatedOther

    CELL-DYN 4000 Diluent/Sheath Reagent, List Number 01H73-01, Distributed and Manufactured by: Abbott Diagnostics Division, 5440 Patrick Henry Drive, Santa Clara, CA

    Incorrect Test Results-product may show a higher than expected platelet background count when used on the CELL-DYN 4000 System and report patient results that are unacceptable (out-of-range).

  • 2007-03-16TerminatedOther

    CELL-DYN 4000 Diluent/Sheath Reagent, List Number 01H73-01, Distributed and Manufactured by: Abbott Diagnostics Division, 5440 Patrick Henry Drive, Santa Clara, CA

    Incorrect Test Results-test results may show a higher than expected platelet background count when used on the CELL-DYN 4000 System and report patient results that are unacceptable (out-of-range).

Monitor diluent, blood cell automatically