Unit, Cryosurgical, Accessories

FDA adverse event reports and recall history for unit, cryosurgical, accessories (product code GEH), from the public MAUDE corpus and the live FDA recall database.

Total reports
3,228
Deaths
57
Injuries
844
Malfunctions
2,231

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 20122026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Recent recalls

  • 2026-08-12Open, ClassifiedDevice Design

    ERBECRYO 2 Cryosurgical Unit Model/Catalog Number: 10402-000

    A strong pull on the high-pressure transfer hose of cryosurgical unit may lead to pneumatic short circuit between he gas inlet and the connector resulting in CO2 leaking into the return path, leading to excessive pressure causing the cryoprobe to rupture.

  • 2026-03-20Open, ClassifiedProcess control

    Flexible Cryoprobe (OD 1.1mm, L1.15mm) w/ oversheath (OD2.6mm, L817mm) REF: 20402-401 STERILE EO. For surgical use

    Probes may rupture/burst during activation

  • 2026-03-20Open, ClassifiedProcess control

    Flexible Cryoprobe (OD 1.7mm, L1.15mm) REF: 20402-410. For surgical use

    Probes may rupture/burst during activation

  • 2026-03-20Open, ClassifiedProcess control

    Flexible Cryoprobe (OD 2.4mm, L1.15mm) REF: 20402-411. For surgical use

    Probes may rupture/burst during activation

  • 2025-09-11Open, ClassifiedProcess control

    ICEfx Cryoablation System, portable helium-free desktop cryoablation unit, Material Numbers (UPN) FPRCH8000, FPRCH8000-02, H749396118000

    Certain desiccant tube subassemblies were built with end caps that were not correctly tightened.

  • 2025-09-11Open, ClassifiedProcess control

    Preventive Maintenance (PM) Kit ICEFX, Material Number (UPN) SPRPM8000

    Certain desiccant tube subassemblies were built with end caps that were not correctly tightened.

  • 2025-04-29Open, ClassifiedProcess control

    IceSeed 1.5 CX S NEEDLE US, Cryoablation Needle, REF H7493967233100. The needles are intended to convert high-pressure gas to either a very cold Freezing application or to a warm Thawing application.

    Boston Scientific has identified that IceSeedTM CX needles manufactured between April 2024 and February 2025 have been programmed with DEMO settings instead of Commercial settings (see Affected Product Listing). Upon initial connection of an IceSeed CX needle to the Cryoablation System, the user will be prompted to perform the Needle Integrity and Functionality Test (NIT) per standard protocols. The affected needles perform as intended unless they are disconnected and reconnected to any channel after testing. In this case, the Cryoablation System will prompt re-execution of the NIT.

  • 2025-04-29Open, ClassifiedProcess control

    IceSeed 1.5 CX S NEEDLE OUS, Cryoablation Needle, REF H7493967334100. The needles are intended to convert high-pressure gas to either a very cold Freezing application or to a warm Thawing application.

    Boston Scientific has identified that IceSeedTM CX needles manufactured between April 2024 and February 2025 have been programmed with DEMO settings instead of Commercial settings (see Affected Product Listing). Upon initial connection of an IceSeed CX needle to the Cryoablation System, the user will be prompted to perform the Needle Integrity and Functionality Test (NIT) per standard protocols. The affected needles perform as intended unless they are disconnected and reconnected to any channel after testing. In this case, the Cryoablation System will prompt re-execution of the NIT.

  • 2025-04-29Open, ClassifiedProcess control

    IceSeed 1.5 CX NEEDLE US, Cryoablation Needle, REF H7493967433170. The needles are intended to convert high-pressure gas to either a very cold Freezing application or to a warm Thawing application.

    Boston Scientific has identified that IceSeedTM CX needles manufactured between April 2024 and February 2025 have been programmed with DEMO settings instead of Commercial settings (see Affected Product Listing). Upon initial connection of an IceSeed CX needle to the Cryoablation System, the user will be prompted to perform the Needle Integrity and Functionality Test (NIT) per standard protocols. The affected needles perform as intended unless they are disconnected and reconnected to any channel after testing. In this case, the Cryoablation System will prompt re-execution of the NIT.

  • 2025-04-29Open, ClassifiedProcess control

    IceSeed 1.5 CX NEEDLE OUS, Cryoablation Needle, REF H7493967534170. The needles are intended to convert high-pressure gas to either a very cold Freezing application or to a warm Thawing application.

    Boston Scientific has identified that IceSeedTM CX needles manufactured between April 2024 and February 2025 have been programmed with DEMO settings instead of Commercial settings (see Affected Product Listing). Upon initial connection of an IceSeed CX needle to the Cryoablation System, the user will be prompted to perform the Needle Integrity and Functionality Test (NIT) per standard protocols. The affected needles perform as intended unless they are disconnected and reconnected to any channel after testing. In this case, the Cryoablation System will prompt re-execution of the NIT.