Skip to main content
GDCClass 1

Table, Operating-Room, Electrical

FDA adverse event reports and recall history for table, operating-room, electrical (product code GDC), from the public MAUDE corpus and the live FDA recall database.

Total reports
1,626
Deaths
1
Injuries
109
Malfunctions
1,376

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Recent recalls

  • 2026-07-01Open, ClassifiedDevice Design

    Adapter for Head Positioning, Model/Catalog Number: 1739994, when used in conjunction with the side rails of the Upper Back Section H V U component (product code 1909813) on Baxter operating tables

    The Adapter for Head Positioning accessory may become loose during patient positioning and unintentionally move when used in conjunction with the side rails of the Upper Back Section H V U component on Baxter operating tables.

  • 2025-03-21Open, ClassifiedSoftware design

    Baxter Operating table column TS7500 MOBIUS, Product Code 1704695

    There is a software issue which causes the upper back section to not be operable/adjustable when the "emergency mode" function is enabled.

  • 2025-03-21Open, ClassifiedSoftware design

    Baxter TruSystem 7500 Hybrid MR IMRIS, Product Code 2067886

    There is a software issue which causes the upper back section to not be operable/adjustable when the "emergency mode" function is enabled.

  • 2025-03-21Open, ClassifiedSoftware design

    Baxter TruSystem 7500, Product Code 4091000

    There is a software issue which causes the upper back section to not be operable/adjustable when the "emergency mode" function is enabled.

  • 2024-10-24Open, ClassifiedDevice Design

    Baxter Operating table TruSystem 7000 U (MB), Product codes: a) REF 1604786; b) REF 1604788; c) REF 1723633

    The firm received customer reports stating that the batteries and their connectors experienced electrical short-circuits and/or emitted smoke.

  • 2024-08-06Open, ClassifiedUnder Investigation by firm

    Baxter PST 500 U, Precision Surgical Table, Part number (product code) 4080300

    Baxter Healthcare Corporation is issuing an Urgent Medical Device Correction for the PST 500 U Precision Surgical Table due to a potential issue with loose spring pins in the spindle drives. This can cause the tabletop to unexpectedly tilt or move at any time, even without active use of the surgical table. This could potentially result in unintentional patient movement during surgical procedures and/or preparation for surgical procedures, including transport.

  • 2023-08-28Open, ClassifiedUnder Investigation by firm

    GS70 Salus Surgical Table, Product Code 18-070-70

    When the control of the operating table was used in Bluetooth mode, the operating table continued moving even after releasing the button on the control. Originally, the root cause was thought to be confined to Bluetooth/wireless communications between the control pendant and the surgical table. However, during investigation, Mizuho discovered that the failure could be replicated even in wired mode if communication between the pendant and table were somehow interrupted. The root cause appears to be: if communication between the controller and table is interrupted while a button is pressed, the table does not know if/when a button is released. This occurs in wireless or wired mode.

  • 2023-08-28Open, ClassifiedUnder Investigation by firm

    GS70H Salus Surgical Table, Product Code 18-071-70

    When the control of the operating table was used in Bluetooth mode, the operating table continued moving even after releasing the button on the control. Originally, the root cause was thought to be confined to Bluetooth/wireless communications between the control pendant and the surgical table. However, during investigation, Mizuho discovered that the failure could be replicated even in wired mode if communication between the pendant and table were somehow interrupted. The root cause appears to be: if communication between the controller and table is interrupted while a button is pressed, the table does not know if/when a button is released. This occurs in wireless or wired mode.

  • 2023-07-07Open, ClassifiedDevice Design

    Allen Advance Chest Support with Pad

    Potential for the device to crack where the Chest Base Prone support attaches to the carbon fiber operating room (OR) table rail.

  • 2022-11-03TerminatedOther

    Allen Advance Table, RSP-ADVANCE TABLE: a) REF A-71100-US, #107003, b) A-71101-EU, #107003. operating room table

    the adhesive, applied to screws and bolts as an additional holding force, may not cure correctly, and as a result, could result in a loss of holding and reduced mechanical integrity.

Monitor table, operating-room, electrical automatically

DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.