Catheter, Cholangiography
FDA adverse event reports and recall history for catheter, cholangiography (product code GBZ), from the public MAUDE corpus and the live FDA recall database.
- Total reports
- 391
- Deaths
- 2
- Injuries
- 88
- Malfunctions
- 273
Reports per year
Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.
Recent recalls
- 2026-09-04Open, ClassifiedNonconforming Material/Component
Getinge PVC Thoracic Catheter, Soft PVC Straight, 10 per Case; Size(Model Number): 1. 16FR(8016), 2. 24FR(8024), 3. 28FR(8028), 4. 32FR(8032), 5. 36FR(8036);
An Out of Specification result (OOS) related to bacterial endotoxin testing in a single lot of catheters.
- 2026-09-04Open, ClassifiedNonconforming Material/Component
Getinge PVC Thoracic Catheter, Soft PVC Right Angle, 10 per Case; Size(Model Number): 1. 24FR(8124), 2. 28FR(8128), 3. 32FR(8132), 4. 36FR(8136);
An Out of Specification result (OOS) related to bacterial endotoxin testing in a single lot of catheters.
- 2026-09-04Open, ClassifiedNonconforming Material/Component
Getinge PVC Thoracic Catheter, Firm PVC Straight, 10 per Case; Size(Model Number): 1. 28FR(15028), 2. 36FR(15036);
An Out of Specification result (OOS) related to bacterial endotoxin testing in a single lot of catheters.
- 2023-12-02Open, ClassifiedNonconforming Material/Component
Atrium Firm PVC Thoracic Catheter 36FR Straight, part number 15036.
A sterilization nonconformance prematurely aged the product, resulting in an inaccurate labeled expiration date.
- 2015-08-05TerminatedEmployee error
Atrium 36Fr Right Angled Firm PVC Thoracic Catheter, (Pkg of 10) Part Number: 15236
Inner pouches incorrectly labeled as part 8136, 36Fr Right Angled Thermosensitive PVC catheter, but are part 15236-36Fr Right Angled Firm PVC Catheter
- 2015-07-22TerminatedProcess control
American Catheter / Cholangiogram Catheter 9100 Series.
Catheters were not sealed before sterilization. and was not detected through subsequent acceptance activities, product release and distribution.
- 2015-07-18TerminatedComponent change control
Ring Transjugular Intrahepatic Access Set Product Usage: The Ring Transjugular Intrahepatic Access Sets are intended for transjugular liver access in diagnostic and interventional procedures. The Ring Transjugular Intrahepatic Access Set consists of the following components: Flexor RTPS Guiding Sheath, needle, catheter, Dilator, Torcon NB Advantage Catheter, Van Andel Dilatation Catheter, Curved Newton Wire Guide, Amplatz Extra Stiff Wire Guide.
Cook Inc. is initiating a voluntary recall of the Ring Transjugular Intrahepatic Access Set and the Flexor RTPS Introduce Sheath due to complaints of the dilators being too long.
- 2014-06-12TerminatedEmployee error
Taut Operative Cholangiogram Catheter, 4.5FR (1.5mm) x 76cm To facilitate the introduction of contrast media into the biliary ducts to obtain a cholangiogram during laparoscopic or open cholecystectomy.
Several complaints were received for the metal support tube missing from the cholangiogram catheter.
- 2008-08-14TerminatedNonconforming Material/Component
HydraGlide Heparin Coated Silicone Thoracic Catheters - Straight, Atrium Code # 14120, Size (Fr) 20, # Eyelets 5, 10 per case
Heparin-coated Thoracic Catheters were manufactured with heparin allegedly contaminated with oversulfated chondroitin sulfate
- 2008-08-14TerminatedNonconforming Material/Component
HydraGlide Heparin Coated Silicone Thoracic Catheters - Straight, Atrium Code # 14124, Size (Fr) 24, # Eyelets 6, 10 per case
Heparin-coated Thoracic Catheters were manufactured with heparin allegedly contaminated with oversulfated chondroitin sulfate
Monitor catheter, cholangiography automatically
DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.