Skip to main content
GADClass 1

Retractor

FDA adverse event reports and recall history for retractor (product code GAD), from the public MAUDE corpus and the live FDA recall database.

Total reports
1,671
Deaths
4
Injuries
313
Malfunctions
1,275

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Recent recalls

  • 2026-07-09Open, ClassifiedLabeling design

    OmniTract Surgical Retractor System, which includes 633 of the firm s product codes.

    The current Omni-Tract Surgical Retractor System IFU states a dry time of 30 minutes. Integra LifeSciences has identified that the dry time needs to be extended to 90 minutes, 110 minutes, or 120 minutes depending on the sterilization cycle exposure time used.

  • 2025-05-15Open, ClassifiedUnder Investigation by firm

    Centurion Sterile Weitlaner Retractor, Reorder: 67315

    Three is the potential for the retractors to puncture through the sterile packaging.

  • 2025-01-16Open, ClassifiedUnder Investigation by firm

    ECONO STERILE medical procedure kits labeled as: 1) ECONO STERILE ARMY RETR (SET OF 2) STER CS25, Model Number 96-2243; 2) ECONO STERILE ARMY RETR (SET OF 2)STRL CS/25, Model Number 96-2243M; 3) ECONO STERILE SENN RETR DE BL 6.25"STRLCS/25, Model Number 96-2267; 4) ECONO STERILE SENN RETR DE BL 6.25"STRLCS/25, Model Number 96-2267A; 5) ECONO STERILE SENN RETR DE SH 6.25"STRLCS/25, Model Number 96-2269A; 6) ECONO STERILE MINI HOHMANN 6MM 6.5" STERCS25, Model Number 96-2959; retractors

    Reports of various packaging issues that may result in a breach of the sterile barrier.

  • 2024-06-28Open, ClassifiedUnder Investigation by firm

    Brand Name: Fathom Pedicle-Based Retractor Product Name: C/C Blade, 35-55mm Model/Catalog Number: MR2102035 Product Description: Cranial Caudal Blade Component: Yes, a component of a retractor blade

    Cranial Caudal Blades manufactured incorrectly could result in fracture and/or damage to the bone structure

  • 2024-06-28Open, ClassifiedUnder Investigation by firm

    Brand Name: Fathom Pedicle-Based Retractor Product Name: C/C Blade, 50-80mm Model/Catalog Number: MR2102050 Product Description: Cranial Caudal Blade Component: Yes, a component of a retractor blade

    Cranial Caudal Blades manufactured incorrectly could result in fracture and/or damage to the bone structure

  • 2024-03-29Open, ClassifiedProcess control

    Centurion retractor kits labeled as: a) STERILE 6-1/2" SENN RETRACTOR (ST7185), Product Code 66795; b) STERILE RAGNELL RETRACTOR (RR602), Product Code 67525; c) ST. SENN RETRACTOR 3 PRG BLUNT (ST7085), Product Code 37535

    Product may have a weak seal which may result in a breach of sterility of the contents if the seal fails. The weak seal may not be detectable by all users.

  • 2024-02-23Open, ClassifiedDevice Design

    Rebstock Holding rod for fixation -For holding and fixation of flexible Leyla Arms and spatula Art. No. 06-06-285 and 06-06-285MOD. surgical instrument.

    Holding arm could move (twist) uncontrollably in the front area. This could lead to a serious deterioration of patient's safety if the movement is transferred to the spatula, which is attached to the holding arm.

  • 2023-07-05Open, ClassifiedOther

    Pilling LOWSLEY PROSTATIC TRACTOR CVD, REF 243200; retractor

    There are incomplete pre-cleaning instructions within the Instructions for Use (IFU).

  • 2023-07-05Open, ClassifiedOther

    Pilling LOWSLEY PROSTATIC TRACTOR STR, REF 243201; retractor

    There are incomplete pre-cleaning instructions within the Instructions for Use (IFU).

  • 2023-01-12Open, ClassifiedEquipment maintenance

    RadiaLux Lighted Retractor (Pink), REF: 50-101-1

    Defect in sterile pouch seal, incomplete seal, of lighted Retractors, increases the risk of contamination, which can lead to post-operative infection.

Monitor retractor automatically

DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.