Indicator, Biological Sterilization Process

FDA adverse event reports and recall history for indicator, biological sterilization process (product code FRC), from the public MAUDE corpus and the live FDA recall database.

Total reports
1,469
Deaths
0
Injuries
24
Malfunctions
1,346

Reports per year

11
12
13
14
15
16
17
18
19
20
21
22
23
24
25

Source: FDA adverse event reports by year received, 2011–2025. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Recent recalls

  • 2026-10-02Open, ClassifiedUnder Investigation by firm

    VERIFY Biological/Integrator Challenge Pack Model/Catalog Number: LCB006

    Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction

  • 2026-10-02Open, ClassifiedUnder Investigation by firm

    VERIFY Biological/Integrator Challenge Pack Model/Catalog Number: LCB007

    Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction

  • 2026-10-02Open, ClassifiedUnder Investigation by firm

    VERIFY Dual Species Self-Contained Biological Indicator Model/Catalog Number: S3060

    Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction

  • 2026-10-02Open, ClassifiedUnder Investigation by firm

    VERIFY Dual Species Self-Contained Biological Indicator Model/Catalog Number: S3061

    Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction

  • 2026-10-02Open, ClassifiedUnder Investigation by firm

    VERIFY Biological Steam Test Pack Model/Catalog Number: S3065

    Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction

  • 2026-10-02Open, ClassifiedUnder Investigation by firm

    VERIFY Biological Steam Test Pack Model/Catalog Number: S3069

    Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction

  • 2026-06-05Open, ClassifiedUnder Investigation by firm

    Brand Name: SPORETRAQ MAIL-IN SPORE TESTING SERVICE, 52 TESTS Product Name: SPORETRAQ MAIL-IN SPORE TESTING SERVICE, 52 TESTS Model/Catalog Number: 1023780 Software Version: N/A Product Description: The SporeTraq Mail-In Spore Testing Service includes 52 bacterial Spore Strips provided by Mesa Labs. The Spore Strip (single species G. stearothermophilus ATCC#7953 or dual species G. stearothermophilus ATCC #7953 and B. atrophaeus ATCC #9372) is intended for use in testing the efficacy of sterilizer systems used in healthcare facilities. The Spore Strips are compatible with Steam sterilization, Chemiclave sterilization, Ethylene Oxide sterilization, or Dry Heat sterilization processes. The test strip is placed in the customer s sterilizer during their sterilization cycle and upon completion the sterilized test strip along with the control strip is returned to Mesa Labs for lab testing in the return envelope. Spore test results are digitally stored in HealthFirst s OnTraq smart automation tracking and reporting system for easy and secure access to a paperless archive from any device with internet access. Component: N/A

    Due to incorrect incubation process being performed on product.

  • 2026-06-05Open, ClassifiedUnder Investigation by firm

    Brand Name: SPORETRAQ MAIL-IN SPORE TESTING SERVICE, 12 TESTS Product Name: SPORETRAQ MAIL-IN SPORE TESTING SERVICE, 12 TESTS Model/Catalog Number: 1023770 Software Version: N/A Product Description: The SporeTraq Mail-In Spore Testing Service includes 12 bacterial Spore Strips provided by Mesa Labs. The Spore Strip (single species G. stearothermophilus ATCC#7953 or dual species G. stearothermophilus ATCC #7953 and B. atrophaeus ATCC #9372) is intended for use in testing the efficacy of sterilizer systems used in healthcare facilities. The Spore Strips are compatible with Steam sterilization, Chemiclave sterilization, Ethylene Oxide sterilization, or Dry Heat sterilization processes. The test strip is placed in the customer s sterilizer during their sterilization cycle and upon completion the sterilized test strip along with the control strip is returned to Mesa Labs for lab testing in the return envelope. Spore test results are digitally stored in HealthFirst s OnTraq smart automation tracking and reporting system for easy and secure access to a paperless archive from any device with internet access. Component: N/A

    Due to incorrect incubation process being performed on product.

  • 2025-02-27Open, ClassifiedLabeling False and Misleading

    ASP BIOTRACE Auto Read 60 Steam BI Process Challenge Device, REF 73135.

    Error in the IFU lists an incorrect incubation time on one occasion of 20 minutes rather than the correct incubation time of 60 minutes.

  • 2025-02-11Open, ClassifiedError in labeling

    Biological, Bio52 HealthFirst Mail In Monitoring System, 52 Tests, Item: 1003600. Biological, Bio12 HealthFirst Mail In Monitoring System, 12 Tests, Item: 1003580

    Spore test strip monitoring system, designed for routine monitoring of sterilization processes, may have a printing error on the return envelope that may cause spore test results to be reported to the incorrect customer's account.

Monitor indicator, biological sterilization process automatically