FQOClass 1

Table, Operating-Room, Ac-Powered

FDA adverse event reports and recall history for table, operating-room, ac-powered (product code FQO), from the public MAUDE/AEMS corpus and the live FDA recall database.

Total reports
2,396
Deaths
16
Injuries
308
Malfunctions
1,491

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 20122026. Reporting paused during the 2026 MAUDE-to-AEMS transition and has since resumed.

Recent recalls

  • 2025-08-22Open, ClassifiedUnder Investigation by firm

    Corin Operating Tables Model Numbers: (1) 770001A0, (2) 770001A2, (3) 770001B0, (4) 770001B2, (5) 770001F0, (6) 770001F2; All Software Versions.

    Under certain conditions, such as the simultaneous use of the operating table with other medical devices, including high-frequency (HF) surgical equipment, interference may occur that could prevent the device from performing as intended. If the issue occurs, the operating table is unresponsive to command from both the universal remote control and override panel.

  • 2025-04-23Open, ClassifiedProcess change control

    Skytron Freedom 4FXS-60 Heavy Duty Monitor Mount

    The mounting hardware that is supplied with the 4FXS-60 monitor bracket does not provide adequate thread engagement to support the weight of the frame and monitor. This could lead to a failure of the mounting hardware and subsequent separation of the frame from the mounting column.

  • 2024-11-14Open, ClassifiedComponent design/selection

    MEERA CL mobile operating table - Designed for the placement and positioning of the patient immediately before, during and after surgical interventions, as well as for examination and treatment, Item Number: 720001B0

    Error code "50037" occurs on the IR-Hand Control when a MEERA-table is controlled, and the table stops the movement. This error may occur sporadically when the MEERA-table is controlled via an IR-Hand Control and may result in procedural delay

  • 2024-11-14Open, ClassifiedComponent design/selection

    MEERA CL mobile operating table - Designed for the placement and positioning of the patient immediately before, during and after surgical interventions, as well as for examination and treatment Item Number: 720001B2

    Error code "50037" occurs on the IR-Hand Control when a MEERA-table is controlled, and the table stops the movement. This error may occur sporadically when the MEERA-table is controlled via an IR-Hand Control and may result in procedural delay

  • 2024-11-14Open, ClassifiedComponent design/selection

    MEERA CL mobile operating table - Designed for the placement and positioning of the patient immediately before, during and after surgical interventions, as well as for examination and treatment Item Number: 720001F0

    Error code "50037" occurs on the IR-Hand Control when a MEERA-table is controlled, and the table stops the movement. This error may occur sporadically when the MEERA-table is controlled via an IR-Hand Control and may result in procedural delay

  • 2024-11-14Open, ClassifiedComponent design/selection

    MEERA CL mobile operating table - Designed for the placement and positioning of the patient immediately before, during and after surgical interventions, as well as for examination and treatment Item Number: 720001F2

    Error code "50037" occurs on the IR-Hand Control when a MEERA-table is controlled, and the table stops the movement. This error may occur sporadically when the MEERA-table is controlled via an IR-Hand Control and may result in procedural delay

  • 2024-11-14Open, ClassifiedComponent design/selection

    MEERA CL mobile operating table - Designed for the placement and positioning of the patient immediately before, during and after surgical interventions, as well as for examination and treatment Item Number: 710001B0

    Error code "50037" occurs on the IR-Hand Control when a MEERA-table is controlled, and the table stops the movement. This error may occur sporadically when the MEERA-table is controlled via an IR-Hand Control and may result in procedural delay

  • 2024-11-14Open, ClassifiedComponent design/selection

    MEERA CL mobile operating table - Designed for the placement and positioning of the patient immediately before, during and after surgical interventions, as well as for examination and treatment Item Number: 710001B2

    Error code "50037" occurs on the IR-Hand Control when a MEERA-table is controlled, and the table stops the movement. This error may occur sporadically when the MEERA-table is controlled via an IR-Hand Control and may result in procedural delay

  • 2024-11-14Open, ClassifiedComponent design/selection

    MEERA CL mobile operating table - Designed for the placement and positioning of the patient immediately before, during and after surgical interventions, as well as for examination and treatment Item Number: 700001B0

    Error code "50037" occurs on the IR-Hand Control when a MEERA-table is controlled, and the table stops the movement. This error may occur sporadically when the MEERA-table is controlled via an IR-Hand Control and may result in procedural delay

  • 2024-11-14Open, ClassifiedComponent design/selection

    MEERA CL mobile operating table - Designed for the placement and positioning of the patient immediately before, during and after surgical interventions, as well as for examination and treatment Item Number: 700001F0

    Error code "50037" occurs on the IR-Hand Control when a MEERA-table is controlled, and the table stops the movement. This error may occur sporadically when the MEERA-table is controlled via an IR-Hand Control and may result in procedural delay

Monitor table, operating-room, ac-powered automatically