Mattress, Air Flotation, Alternating Pressure
FDA adverse event reports and recall history for mattress, air flotation, alternating pressure (product code FNM), from the public MAUDE/AEMS corpus and the live FDA recall database.
- Total reports
- 2,491
- Deaths
- 154
- Injuries
- 763
- Malfunctions
- 1,288
Reports per year
Source: FDA adverse event reports by year received, 2012–2026. Reporting paused during the 2026 MAUDE-to-AEMS transition and has since resumed.
Recent recalls
- 2026-02-26Open, ClassifiedSoftware design
Adapt Pump, Model/Catalog Number: 61600200-Adapt Pump, component of Adapt support surfaces (Adapt Air, Adapt Air EX, Adapt Air Pro, Adapt Air Pro EX, Adapt Convertible, Adapt Convertible EX), sofware version 1.3.0; Mattress, Air Flotation, Alternating Pressure
Agiliti identified an issue where the Adapt Pump (61600200-Adapt Pump) supplied with all Adapt Line support surfaces does not re-engage the Microclimate Management function after using the Autofirm function for 10 minutes. This issue only occurs when the 10-minute timer is allowed to elapse after the Autofirm button is pressed after initial setup, and it only affects the Microclimate Management function.
- 2025-12-11Open, ClassifiedRelease of Material/Component prior to receiving test results
Stryker Arise 1000EX mattress, Part Number 2236000000
Stryker has issued a voluntary recall of certain MV3 beds and Arise 1000EX mattresses because some of these products have been sold in close proximity, suggesting a potential intent for them to be used together. These products have not been tested for compatibility and thus are not designed to be used together. The combination of both is not recommended by Stryker.
- 2025-06-12Open, ClassifiedSoftware design
Brand Name: Guardian 2 System Product Name: Guardian System 2 Controller Model/Catalog Number: GS-2.5-C Software Version: 2.5.4-59 (Fix)
A software issue where "low" and "terminal" battery alerts did not display on the screen to the user in specific circumstances when the device was not running patient therapy.
- 2025-06-12Open, ClassifiedSoftware design
Brand Name: Guardian 2 System Product Name: Guardian System 2 Controller Model/Catalog Number: GS-2.5-C Software Version: 2.5.4-69 (Fix)
A software update corrected an issue where "low" and "terminal" battery alerts did not display on the screen to the user in specific circumstances when the device was not running patient therapy.
- 2016-03-08TerminatedComponent design/selection
Stryker Position Pro Mattress. Mattress, Air Flotation, Alternating Pressure. PositionPRO is a powered pressure relief support surface with low air loss (LAL) intended for medical purposes. PositionPRO consists of multiple air cells filled and emptied by an integrated control unit to provide changes in the distribution of body weight for pressure relief. PositionPRO offers turn assist. PositionPRO is for use with GoBed II, Secure II/3, InTouch, and Epic II 84 in. x 35 in. flat deck frames. This surface is powered via an AC power cord that comes in either a 4 foot long option (PN QDF8087) or 15 foot long option (PN QDF8088). The 4 foot long option is used when the surface is plugged into the auxiliary outlet of the bed. The 15 foot long option is utilized when the surface needs to be plugged into a wall outlet. This product uses a power cord, manufactured by ElectriCord that plugs into the wall outlet or the auxiliary outlet of the bed that the surface is placed onto.
Potential for power cord to melt; hazards include electric shock and fire.
- 2014-05-19TerminatedNonconforming Material/Component
MaXair True Low Air Loss Mattress Replacement System, True Air Technologies, Inc.
The firm is recalling the products due to lack of design controls, lack of a Device Master Record, and failure to approve specific vendors.
- 2010-03-31TerminatedNonconforming Material/Component
AccuMax Quantum Complete (Model CU3/CU4), an alternating pressure with turn assist & on demand low air loss system; Encompass Therapuetic Support Systems, Glendale, CA. The device is intended to treat or prevent decubitus ulcers, extensive burns, urinary tract blockage and to aid circulation.
Potential out gasing of the battery and short term ignition of the gas within the control unit enclosure. This failure mode includes a risk for fire.
- 2009-01-07TerminatedDevice Design
Synergy Pulse, Powered Flotation Therapy bed, Model number 5609. Designed to provide enhanced static and pulsation pressure relieving therapies. Provides sixteen pressure zones for customizing the support surface to each patient, and is capable of providing pulsation therapy.
These devices have the potential to overheat.
- 2009-01-07TerminatedDevice Design
Synergy Dynamic, Powered Patient Rotation Bed, Model number 5607. Designed to provide enhanced static, pulsation and rotational pressure relieving therapies. Provides sixteen pressure zones for customizing the support surface to each patient, and is capable of providing alternating pressure therapy and lateral rotation up to 45 degrees side to side.
These devices have the potential to overheat.
- 2009-01-07TerminatedDevice Design
Synergy Air Squared, Alternating Pressure Air Flotation Mattress, Model number 6361. Designed to prevent and treat decubitus ulcers. This is accomplished by achieving interface pressures well below capillary closure during the alternation cycle.
These devices have the potential to overheat.
Monitor mattress, air flotation, alternating pressure automatically
DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.