Skip to main content
FNLClass 2

Bed, Ac-Powered Adjustable Hospital

FDA adverse event reports and recall history for bed, ac-powered adjustable hospital (product code FNL), from the public MAUDE corpus and the live FDA recall database.

Total reports
94,615
Deaths
650
Injuries
1,874
Malfunctions
90,319

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Recent recalls

  • 2026-06-04Open, ClassifiedDevice Design

    Pediatric care bed; Product Designation: KayserBett IDA;

    If the adjustment functions of the hand control, as instructed in the Manual, are not locked with the provided key when the patient (child) is unattended, there is an increased risk that the patient (child) or third parties (children) will operate the hand control and trap themselves or others under the bed frame or between the bed frame and the floor. This can lead to serious injuries or even death of the patient (child) or third parties (children).

  • 2024-10-25Open, ClassifiedSoftware design

    a. Hillrom Centrella Smart+ Medical-Surgical Hospital Bed, Product codes: P7900B000001, P7900B100010, P7900B100024, P7900B100208, P7900B100273, P7900B100304, P7900B000005, P7900B100011, P7900B100025, P7900B100212, P7900B100277, P7900B100307, P7900B000010, P7900B100017, P7900B100026, P7900B100222, P7900B100281, P7900B1SPL05, P7900B000011, P7900B100018, P7900B100028, P7900B100223, P7900B100287, P7900B1SPL06, P7900B000020, P7900B100019, P7900B100039, P7900B100224, P7900B100288, P7900B000022, P7900B100020, P7900B100090, P7900B100231, P7900B100292, P7900B100001, P7900B100021, P7900B100161, P7900B100237, P7900B100293, P7900B100005, P7900B100022, P7900B100203, P7900B100242, P7900B100294, P7900B100009, P7900B100023, P7900B100205, P7900B100264, and P7900B100296. b. Hillrom Centrella Smart+ Medical-Surgical Hospital Bed, Product codes: P7900A000001, P7900A000010, P7900A000019, P7900A000039, P7900A000005, P7900A000011, P7900A000022, and P7900A000063.

    The beds have a potential for an interface issue with the Rauland Responder 5 Nurse Call System where the bed exit system will alert at the bed, but may fail to send a remote alert through the nurse call system.

  • 2023-12-22Open, ClassifiedDevice Design

    Arjo medical beds, Models Enterprise 5000X, Enterprise 8000X, Enterprise 9000X and Citadel, assembled with IndiGo Drive Assistance module and retrofitted with IndiGo Drive Assistance modules

    Unintended movement of bed wheels

  • 2023-12-22Open, ClassifiedDevice Design

    Arjo medical beds, Models Enterprise 5000X, Enterprise 8000X, Enterprise 9000X and Citadel, assembled with IndiGo Drive Assistance module and retrofitted with IndiGo Drive Assistance modules

    Unintended movement of bed wheels

  • 2023-12-01Open, ClassifiedProcess control

    ProCuity bed series, model number 3009, item number: 300900000000,

    Patient beds are missing test values for electrical safety tests presenting risk of tissue burn or electrical shock.

  • 2023-12-01Open, ClassifiedProcess control

    ProCuity bed series, model number 3009, item number: 3009PX-L-100.

    Patient beds are missing test values for electrical safety tests presenting risk of tissue burn or electrical shock.

  • 2023-12-01Open, ClassifiedProcess control

    ProCuity bed series, model number 3009, item number: 3009PX-L-200

    Patient beds are missing test values for electrical safety tests presenting risk of tissue burn or electrical shock.

  • 2023-12-01Open, ClassifiedProcess control

    ProCuity bed series, model number 3009, item number: 3009PX-LE-400.

    Patient beds are missing test values for electrical safety tests presenting risk of tissue burn or electrical shock.

  • 2023-12-01Open, ClassifiedProcess control

    ProCuity bed series, model number 3009, item number: 3009PX-LE-450.

    Patient beds are missing test values for electrical safety tests presenting risk of tissue burn or electrical shock.

  • 2023-12-01Open, ClassifiedProcess control

    ProCuity bed series, model number 3009, item number: 3009PX-LEX-400.

    Patient beds are missing test values for electrical safety tests presenting risk of tissue burn or electrical shock.

Monitor bed, ac-powered adjustable hospital automatically