Sterilizer, Steam
FDA adverse event reports and recall history for sterilizer, steam (product code FLE), from the public MAUDE corpus and the live FDA recall database.
- Total reports
- 1,525
- Deaths
- 1
- Injuries
- 140
- Malfunctions
- 905
Reports per year
Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.
Recent recalls
- 2026-08-18Open, ClassifiedComponent design/selection
Fort Defiance Industries (FDI), Automated Field Steam Sterilizer REF: P2131
The Automated Field Steam Sterilizer's water recovery system 120V Amphenol connector may experience an intermittent loss of ground continuity, which could result in an unreliable fault current return path.
- 2023-08-03Open, ClassifiedEmployee error
STATCLAVE G4 Chamber Autoclave -Tabletop steam sterilizer intended for use by a health care provider to sterilize medical products by means of pressurized steam. Model: G4 621103
The STATCLAVE G4 is a dynamic air removal (prevacuum) tabletop steam sterilizer intended for use by a health care provider to sterilize medical products by means of pressurized steam. It is suitable for the sterilization of dental and medical instruments that are validated to be sterilized by steam. The STATCLAVE G4 has not been designed to sterilize liquid loads, bio medical waste or materials not compatible with steam sterilization. The processing of such loads may result in incomplete sterilization and / or damage to the autoclave.
- 2023-08-03Open, ClassifiedEmployee error
STATCLAVE G4 Chamber Autoclave -Tabletop steam sterilizer intended for use by a health care provider to sterilize medical products by means of pressurized steam. Model: G4 622303
The STATCLAVE G4 is a dynamic air removal (prevacuum) tabletop steam sterilizer intended for use by a health care provider to sterilize medical products by means of pressurized steam. It is suitable for the sterilization of dental and medical instruments that are validated to be sterilized by steam. The STATCLAVE G4 has not been designed to sterilize liquid loads, bio medical waste or materials not compatible with steam sterilization. The processing of such loads may result in incomplete sterilization and / or damage to the autoclave.
- 2015-11-04TerminatedSoftware design
AMSCO 400 and AMSCO C Small Steam Sterilizers
In AMSCO Small Steam Sterilizers equipped with the CS-iQ feature, the sterilizer software inadvertently prevents the capability to operate the foot pedal, which allows the Operator to open and close the sterilizer door.
- 2015-04-30TerminatedSoftware design (manufacturing process)
AMSCO C and AMSCO 400 Steam Sterilizers, STERIS Corporation. AMSCO C Small Sterilizer, AMSCO 400 Small Sterilizer, AMSCO 400 Medium Sterilizer. Designed for sterilization of heat and moisture-stable materials used in healthcare facilities.
STERIS has identified that the control board software in select AMSCO 400 and AMSCO C units will interrupt and cancel a processing cycle should the selected sterilizer shutdown time coincide with a processing cycle. This may result in a procedure delay.
- 2013-12-11TerminatedDevice Design
AMSCO Evolution steam sterilizer Transfer Carriage. Product Usage: Mobile, wheeled transportation unit utilized to carry, load, & unload the Loading Car into and out of steam sterilizer units.
The firm discovered that unlatching can occur when overloaded transfer carriages are pulled or maneuvered incorrectly or with excessive force.
- 2013-03-06TerminatedSoftware design
AMSCO C and AMSCO 400 Steam Sterilizers designed for sterilization of heat and moisture-stable materials used in healthcare facilities.
The control boards alarm set points and temperature tolerances allowed for erroneous activation of the AD (analog/digital) alarm as temperature fluctuated during operation. When this alarm occurs, the sterilizer locks up and cannot be used until serviced, causing Customer nuisance.
- 2012-05-11TerminatedDevice Design
The Evolution steam sterilizers, manufactured by Steris Corporation, utilize a loading system to assist the operator in transporting, loading and unloading items to be sterilized. It consists of two major components: -Loading Car: Shelving unit which holds the items to be sterilized; and the -Transfer Carriage: Wheeled unit which loads/unloads the loading car and transports it from one location to another. Product shipped in crate. Product Usage: The Evolution Transfer Carriage is a wheeled unit which loads/unloads the sterilizer loading car and transports it from one location to another.
Through Customer feedback and field service experience, STERIS learned that operators may experience difficulty when inserting or removing equipment from the Evolution sterilizer using the transfer carriage due to the following issues with the transfer carriage. 1) Deformation of the latch pin or cracked ball bearings preventing the roller from turning. 2) The rollers on the ends of the pivot a
- 2012-01-11TerminatedEmployee error
Midmark M11 Ultra Steam Sterilizer, Distributed by: Midmark Corporation, 60 Vista Drive, Versailles, Ohio 45380. Sterilize reusable equipment that is heat and moisture stable.
During and FDA inspection on 8/23/2011, it was discovered that the firm received numerous reports of the steam sterilizer malfunctioning.
- 2008-09-17TerminatedEmployee error
Steris Small Renaissance Eagle 3000 -- Remanufactured Gravity Models 3011 and 3021 and Prevacuum Models 3013 and 3023. Products are designed for efficient sterilization of non-porous and porous, heat and moisture-stable materials used in healthcare facilities.
The firm discovered that some Eagle 3000 units may have a door post pin missing which could cause incomplete or no engagement of the radial arms into the lock rails of the door. The sterilizer is designed with switches that prevent cycle initiation if the door is not locked. However, in the event the door switches are not properly adjusted they may falsely indicate that the door is locked even t
Monitor sterilizer, steam automatically