Device, Impotence, Mechanical/Hydraulic

FDA adverse event reports and recall history for device, impotence, mechanical/hydraulic (product code FHW), from the public MAUDE corpus and the live FDA recall database.

Total reports
38,769
Deaths
9
Injuries
35,201
Malfunctions
3,404

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Recent recalls

  • 2023-04-12Open, ClassifiedProcess change control

    Titan Touch Pump, Catalog Numbers 5177502400; inflatable penile Prosthesis

    A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.

  • 2023-04-12Open, ClassifiedProcess change control

    Titan Pump Assembly, Catalog Number 5177701400; inflatable penile Prosthesis

    A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.

  • 2023-04-12Open, ClassifiedProcess change control

    Titan Touch NB Infr Zero 16cm, Catalog Number EN28162400; inflatable penile Prosthesis

    A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.

  • 2023-04-12Open, ClassifiedProcess change control

    Titan Touch NB Infr Zero 18cm, Catalog Number EN28182400; inflatable penile Prosthesis

    A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.

  • 2023-04-12Open, ClassifiedProcess change control

    Titan Touch NB Scrot Zero 16cm, Catalog Number EN29162400; inflatable penile Prosthesis

    A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.

  • 2023-04-12Open, ClassifiedProcess change control

    Titan Touch NB Scrot Zero 18cm, Catalog Number EN29182400; inflatable penile Prosthesis

    A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.

  • 2023-04-12Open, ClassifiedProcess change control

    Titan NB Infra Zero Ang 16cm, Catalog Number EN88162400; inflatable penile Prosthesis

    A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.

  • 2023-04-12Open, ClassifiedProcess change control

    Titan NB Scrotal Zero Ang 18cm, Catalog Number EN88182400; inflatable penile Prosthesis

    A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.

  • 2023-04-12Open, ClassifiedProcess change control

    Titan Touch Infr Zero Ang 18cm, Catalog Number ES28182400; inflatable penile Prosthesis

    A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.

  • 2023-04-12Open, ClassifiedProcess change control

    Titan Touch Infr Zero Ang 22cm, Catalog Number ES28222400; inflatable penile Prosthesis

    A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.