Electrode, Ph, Stomach
FDA adverse event reports and recall history for electrode, ph, stomach (product code FFT), from the public MAUDE corpus and the live FDA recall database.
- Total reports
- 12,972
- Deaths
- 9
- Injuries
- 988
- Malfunctions
- 11,916
Reports per year
Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.
Recent recalls
- 2026-09-11Open, ClassifiedNonconforming Material/Component
Bravo CF capsule delivery device, 5-pack, Product Number FGS-0635, Bravo CF capsule delivery device, 1-pack, Product Number FGS-0636. Used to attach a pH monitoring capsule to the esophagus.
Capsule may fail to attach to the esophagus or detach from the delivery device. Risks include patient aspiration/inhalation, perforation of esophagus, obstruction of the airway, hemorrhage/blood loss/bleeding, laceration of the esophagus, a delay in diagnosis, and foreign bodies remaining in the patient.
- 2025-07-10Open, ClassifiedProcess control
Bravo CF capsule delivery device, designed to place and attach a pH monitoring capsule to the esophagus using endoscopic or manometric guidance, Product Number FGS-0635 (5 pack) and FGS-0636 (1 pack)
It has been determined that misapplied adhesive on the Bravo CF capsule delivery device may lead to its malfunction. Specifically, the misapplied adhesive may prevent the capsule from attaching to the patient's esophagus or detaching from the delivery device.
- 2016-05-06TerminatedDevice Design
Given Imaging Bravo pH capsule delivery device, 5-pack. In-Vitro Diagnostic for pH monitoring of gastric reflux Item Number: FGS-0312 The Bravo pH Monitoring System is intended to be used for gastroesophageal pH measurement and monitoring of gastric reflux in adults and children from 4 years of age. The Bravo pH Capsule can be attached following either endoscopy or manometry. The RAPID pH software is intended to record, store, view, and analyze gastroesophageal pH data.
Potential of an allergic reaction in a patient with nickel sensitivity.
- 2016-05-06TerminatedDevice Design
Given Imaging Bravo pH capsule delivery device, 1-pack. In- Vitor Diagnostic for pH monitoring of gastric reflux Item Number: FGS-0313 The Bravo pH Monitoring System is intended to be used for gastroesophageal pH measurement and monitoring of gastric reflux in adults and children from 4 years of age. The Bravo pH Capsule can be attached following either endoscopy or manometry. The RAPID pH software is intended to record, store, view, and analyze gastroesophageal pH data.
Potential of an allergic reaction in a patient with nickel sensitivity.
- 2008-05-07TerminatedLabeling Change Control
Medtronic Bravo pH Capsule with Delivery System, Stomach pH electrode, Product Numbers 9012B1011 and 9012B1001. Medtronic, Inc., Minneapolis, MN 55432
Unable to detach from source: The capsule may not detach from the delivery system following attachment to the esophageal wall.
- 2006-05-03TerminatedOther
Medtronic Zinetics 24ME Multi-Use External Reference pH Catheter. For Gastroesophageal pH Measurements. Catalog number 9012P2121. Distributed by Medtronic, Inc., 710 Medtronic Parkway NE, Minneapolis, MN 55432-5604, USA.
Lot 0001491 has a manufacturing defect in the sensor spacing. Sensors have been placed 5 cm apart instead of 15 cm apart. Medtronic has determined that patient safety is not compromised. Clinicians however should review and reassess the results of any patient studies for which catheters from lot 0001491 were used.
- 2006-01-13TerminatedOther
Medtronic Slimline Single-Use Internal Reference pH Catheter for Gastroesophageal pH measurements. Models 9012P3101, 9012P3111, 9012P3121, 9012P3131 and 9012P3141. Distributed by / Medtronic Inc. 710 Medtronic Parkway NE, Minneapolis, MN 55432-5604 USA
Certain lots have a manufacturing defect that allows the water pushed through the catheter to travel back into the connector of the Digitrapper pH recorder. Medtronic has determined that patient safety is not compromised, but catheter function may be affected and damage may result to the Digitrapper pH recorder.
- 2003-11-11TerminatedOther
ComforTEC Plus Impedance Probe, a part of the Insight S980000 Gastrointestinal motility system
Pouch containing pediatric probes was labeled as containing adult probes.
- 2003-01-23TerminatedOther
''ComforTEC II Single Use pH Probe *** Latex Free *** Do Not Re-use''. Packaged individually.
Failure of the dry silver chloride reference, resulting in malfunction and a 'Lead Off' indication.
- Open, ClassifiedProcess change control
CF Delivery Dev Caps Bravo x5, FGS-0635/FGS 0636 - Product Usage: intended to be used for gastroesophageal pH measurement and monitoring of gastric reflux in adults and children from 4 years of age.
The capsule may fail to attach to the esophageal mucosa, which in turn may lead to the aspiration of the capsule. In the event of capsule aspiration, immediate intervention to remove the capsule is required. Potential outcomes following capsule aspiration include low oxygen saturation, intervention to retrieve the capsule, potential need to intubate the patient, extended hospital stay and a delay of treatment. Thirteen reports of capsule aspiration have been received the last 2 years.
Monitor electrode, ph, stomach automatically
DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.