FDA adverse event reports and recall history for tube, drainage, suprapubic (product code FFA), from the public MAUDE corpus and the live FDA recall database.

Total reports
678
Deaths
5
Injuries
248
Malfunctions
417

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Recent recalls

  • 2024-09-13Open, ClassifiedPackaging process control

    Flexima APDL Drainage Catheter System and Kit, Material Numbers: a) REF M001271960, b) REF M001271970; all-purpose drainage catheter with locking pigtail

    There is the potential for holes in the clear sterile barrier of the device pouch which may impact device sterility.

  • 2024-09-13Open, ClassifiedPackaging process control

    Flexima Regular Kit Biliary Catheter System Kit, Material Numbers REF M001271650; bile drainage catheter with locking pigtail

    There is the potential for holes in the clear sterile barrier of the device pouch which may impact device sterility.

  • 2024-09-13Open, ClassifiedPackaging process control

    Flexima Regular Kit Nephrostomy Catheter System Kit, Material Numbers REF M001271860; To provide external drainage of the urinary tract.

    There is the potential for holes in the clear sterile barrier of the device pouch which may impact device sterility.

  • 2024-09-13Open, ClassifiedPackaging process control

    vanSonnenberg Sump Locking Pigtail Regular Kit Sump Catheter System Kit, Material Number REF M001273090; for use as a percutaneously placed drain for intra-abdominal fluid collections.

    There is the potential for holes in the clear sterile barrier of the device pouch which may impact device sterility.

  • 2024-09-13Open, ClassifiedPackaging process control

    VTC Regular Kit Nephrostomy Catheter System Kit, Material Number REF M001245300; to provide external drainage of the urinary tract.

    There is the potential for holes in the clear sterile barrier of the device pouch which may impact device sterility.

  • 2009-05-21TerminatedLabeling False and Misleading

    Boston Scientific, Flexima and Percuflex Drainage Catheter Family (Pigtail Style) UPN DESCRIPTION M001271770 FLEXIMA NEPH FIRM 8F/25CM M001271780 FLEXIMA NEPH FIRM 10F/25CM M001271790 FLEXIMA NEPH REG 8F/25CM M001271800 FLEXIMA NEPH REG 10F/25CM M001271810 FLEXIMA NEPH REG 12F/25CM M001271820 FLEXIMA NEPH REG 14F/25CM M001271830 FLEXIMA NEPH SOFT 10F/25CM M001271840 FLEXIMA NEPH SOFT 12F/25CM M001271850 FLEXIMA NEPH SOFT 14F/25CM M001281790 FLEXIMA NEPHROSTOMY 8/25 TT M001281800 FLEXIMA NEPHROSTOMY 10/25 TT M001281810 FLEXIMA NEPHROSTOMY 12/25 TT M001281820 FLEXIMA NEPHROSTOMY 14/25 TT M001271860 FLEXIMA NEPH 8F/25CM REG KIT M001271870 FLEXIMA NEPH 10F/25CM REG KIT M001271880 FLEXIMA NEPH 12F/25CM REG KIT M001271890 FLEXIMA NEPH 14F/25CM REG KIT M001271900 FLEXIMA NEPH 8F/25CM FIRM KIT M001271910 FLEXIMA NEPH 10F/25CM FIRM KIT M001281650 FLEXIMA NEPH 8/25 TEMP TIP KIT M001281660 FLEXIMA NEPH/10/25 TEMPTIP KIT Single Use Only, Sterilized with Ethylene Oxide Gas, Boston Scientific Corporation, One Boston Scientific Place, Natick, MA 01760-1537, Made in USA 780 Brookside Drive, Spencer, IN 47460 1080. Designed for percutaneous drainage of abscess fluid, the Biliary system and the urinary tract.

    Boston Scientific Corporation is initiating an urgent Field Correction related to its Flexima and Percuflex Drainage Catheter Family (Pigtail Style). During the attempted removal of the device, sutures were separated from the catheter and remained in the patient.

  • 2009-05-21TerminatedLabeling False and Misleading

    Boston Scientific, Flexima and Percuflex Drainage Catheter Family (Pigtail Style) UPN DESCRIPTION M001271240 FLEXIMA APDL FIRM 10F/25CM M001271310 FLEXIMA APDL FIRM 6.3F/25CM M001271320 FLEXIMA APDL FIRM 8F/25CM M001271330 FLEXIMA APDL REG 6.3F/25CM M001271340 FLEXIMA APDL REG 8F/25CM M001271350 FLEXIMA APDL REG 10F/25CM M001271380 FLEXIMA APDL REG 12F/25CM M001271390 FLEXIMA APDL REG 14F/25CM M001271460 FLEXIMA QUICKSTICK LL 8FR M001271470 FLEXIMA QUICKSTICK LL 10FR M001275060 APDTL/7.3FR FLEXIMA SINGLE TT M001275080 APDLTL/7.3FR FLEXIMA SINGLE TT M001281340 FLEXIMA APDL 8/25 TT M001281350 FLEXIMA APDL 10/25 TT M001281380 FLEXIMA APDL 12/25 TT M001281390 FLEXIMA APDL 14/25 TT M001271960 FLEXIMA APDL 8FR REG KIT M001271970 FLEXIMA APDL 10FR REG KIT M001271980 FLEXIMA APDL 8FR FIRM KIT M001271990 FLEXIMA APDL 10FR FIRM KIT M001275070 APDTL/7.3FR FLEXIMA KIT TT M001275010 APDLTL/7.3 FLEXIMA FIRM SINGLE M001275090 APDLTL/7.3FR FLEXIMA KIT TT M001275030 FLEXIMA APDLTL/7/30 STAND KIT M001275050 FLEXIMA APDLTL/7/30 DELUXE KIT Single Use Only, Sterilized with Ethylene Oxide Gas, Boston Scientific Corporation, One Boston Scientific Place, Natick, MA 01760-1537, Made in USA 780 Brookside Drive, Spencer, IN 47460 1080. Designed for percutaneous drainage of abscess fluid, the Biliary system and the urinary tract.

    Boston Scientific Corporation is initiating an urgent Field Correction related to its Flexima and Percuflex Drainage Catheter Family (Pigtail Style). During the attempted removal of the device, sutures were separated from the catheter and remained in the patient.

  • 2009-05-21TerminatedLabeling False and Misleading

    Boston Scientific, Flexima and Percuflex Drainage Catheter Family (Pigtail Style) UPN DESCRIPTION M001271540 FLEXIMA BILIARY FIRM 8F/35CM M001271550 FLEXIMA BILIARY FIRM 10F/35CM M001271560 FLEXIMA BILIARY REG 8F/35CM M001271570 FLEXIMA BILIARY REG 10F/35CM M001271580 FLEXIMA BILIARY REG 12F/35CM M001271590 FLEXIMA BILIARY REG 14F/35CM M001271600 FLEXIMA BILIARY SOFT 10F/35CM M001271610 FLEXIMA BILIARY SOFT 12F/35CM M001271620 FLEXIMA BILIARY SOFT 14F/35CM M001281560 FLEXIMA BILIARY 8/35 TT M001281570 FLEXIMA BILIARY 10/35 TT M001281580 FLEXIMA BILIARY 12/35 TT M001281590 FLEXIMA BILIARY 14/35 TT M001271630 FLEXIMA BIL 8F/35CM FIRM KIT M001271640 FLEXIMA BIL 10F/35CM FIRM KIT M001271650 FLEXIMA BIL 8F/35CM REG KIT M001271660 FLEXIMA BIL 10F/35CM REG KIT Single Use Only, Sterilized with Ethylene Oxide Gas, Boston Scintific Corporation, One Boscton Scintific Place, Natick, MA 01760-1537, Made in USA 780 Brookside Drive, Spencer, IN 47460 1080. Designed for percutaneous drainage of abscess fluid, the Biliary system and the urinary tract.

    Boston Scientific Corporation is initiating an urgent Field Correction related to its Flexima and Percuflex Drainage Catheter Family (Pigtail Style). During the attempted removal of the device, sutures were separated from the catheter and remained in the patient.

  • 2009-05-21TerminatedLabeling False and Misleading

    Boston Scientific, Flexima and Percuflex Drainage Catheter Family (Pigtail Style) UPN# DESCRIPTION 435428-01 M001273030 FLEXIMA VSCL REG 12FR/35CM M001273040 FLEXIMA VSCL REG 14FR/35CM M001273090 FLEXIMA VSDC REG LOOP 12/35KIT M001273100 FLEXIMA VSDC REG LOOP 14/35KIT M001273150 FLEXIMA VSCL FIRM 12FR/35CM M001273160 FLEXIMA VSCL FIRM 14FR/35CM M001273210 FLEXIMA VSDC FIRM LOOP 12/35 M001273220 FLEXIMA VSDC FIRM LOOP 14/35 Single Use Only, Sterilized with Ethylene Oxide Gas, Boston Scintific Corporation, One Boscton Scintific Place, Natick, MA 01760-1537, Made in USA 780 Brookside Drive, Spencer, IN 47460 1080. Designed for percutaneous drainage of abscess fluid, the Biliary system and the urinary tract.

    Boston Scientific Corporation is initiating an urgent Field Correction related to its Flexima and Percuflex Drainage Catheter Family (Pigtail Style). During the attempted removal of the device, sutures were separated from the catheter and remained in the patient.

  • 2006-10-28TerminatedOther

    Boston Scientific 10 Fr (3.3 mm) Flexima Regular APDL All Purpose Drainage catheter set; REF/catalog no. 27-135. Universal Product Number (UPN) M001271350.

    Sterility compromised/package integrity : the bottom of the Tyvek pouch may not have been sealed.

Monitor tube, drainage, suprapubic automatically