Skip to main content
FCMClass 2

Tray, Catheterization, Sterile Urethral, With Or Without Catheter (Kit)

FDA adverse event reports and recall history for tray, catheterization, sterile urethral, with or without catheter (kit) (product code FCM), from the public MAUDE/AEMS corpus and the live FDA recall database.

Total reports
800
Deaths
2
Injuries
97
Malfunctions
692

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 20122026. Reporting paused during the 2026 MAUDE-to-AEMS transition and has since resumed.

Recent recalls

  • 2024-09-06Open, ClassifiedUnder Investigation by firm

    Medline Convenience kits labeled as: 1) URETEROSCOPY CDS SC-LF, Pack Number CDS983922F ; 2) URETEROSCOPY TRAY-RF, Pack Number DYNJ27442U ; 3) URETHROPLASTY PACK, Pack Number DYNJ39700B ; 4) RRP PACK, Pack Number DYNJ44109B ; 5) PROSTATECTOMY PACK-LF , Pack Number DYNJ66109A ; 6) PERINEAL PACK , Pack Number DYNJ69895; 7) LIVER 2018, Pack Number DYNJ905655A; 8) UROLOGY PACK, Pack Number DYNJT2255M

    Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.

  • 2024-02-16Open, ClassifiedNonconforming Material/Component

    Dover" 100% SILICONE 400 ML U/M FOLEY TRAY, NEEDLE PORT, 16 FR (5.3 MM), 5 CC, 2000 ML DRAIN BAG; Dover" HYDROGEL COATED LATEX 400 ML U/M FOLEY TRAY, NEEDLE, 16 FR (5.3 MM), 5 CC, 2000 ML BAG; Dover" UNIVERSAL TRAY, 1000 ML BASIN, PVP SOLUTION, PREPPING COMPONENTS, 10 ML PRE-FILLED SYRINGE; Dover" UNIVERSAL TRAY, 1000 ML BASIN, BZK SWAB STICKS, PREPPING COMPONENTS, 10 ML PRE-FILLED SYRINGE; Dover" UNIVERSAL TRAY, 1000 ML BASIN, PVP SWAB STICKS, PREPPING COMPONENTS, 10 ML PRE-FILLED SYRINGE; Dover" UNIVERSAL TRAY, 1200 ML BASIN, PVP SWAB STICKS, PREPPING COMPONENTS, 10 CC PRE-FILLED SYRINGE; Dover" 100% SILICONE FOLEY INSERTION TRAY, 18 FR (6.0 MM), 5 CC, PREP TRAY; Dover" HYDROGEL COATED LATEX FOLEY INSERTION TRAY, 18 FR (6.0 MM), 5 CC, PREP TRAY; Dover" HYDROGEL COATED LATEX FOLEY INSERTION TRAY, 16 FR (5.3 MM), 5 CC, PREP TRAY; Dover" UNIVERSAL TRAY, 1000 ML BASIN, PVP SOLUTION, PREPPING COMPONENTS, 10 CC PRE-FILLED SYRINGE; Dover" 100% SILICONE FOLEY INSERTION TRAY, 16 FR (5.3 MM), 5 CC, PREP TRAY; Dover" URINE DRAINAGE ADD-A-FOLEY TRAY, NEEDLE SAMPLING, DRAIN TUBE, PREP TRAY; Dover" HYDROGEL COATED LATEX 400 ML U/M FOLEY TRAY, NEEDLE, 18 FR (6.0 MM), 5 CC, 2000 ML BAG; Dover" UNIVERSAL TRAY, 1000 ML BASIN, BZK SWAB STICKS, PREPPING COMPONENTS, 30 CC PRE-FILLED SYRINGE; Dover" UNIVERSAL TRAY, 1000 ML BASIN, PVP SOLUTION, PREPPING COMPONENTS, 30 CC PRE-FILLED SYRINGE; Dover" URINE METER ADD-A-FOLEY TRAY, 400 ML, NEEDLE SAMPLING, DRAIN SPOUT, PREP TRAY; Dover" SILICONE ELASTOMER COATED LATEX FOLEY TRAY, 16 FR (5.3 MM), 5 CC, 2000 ML DRAINAGE BAG; Dover" 100% SILICONE FOLEY CATHETER KIT, 5 CC, 2-WAY, 16 FR (5.3 MM), LUBE, STRL WATER, 10CC SYRINGE; Dover" URINE DRAINAGE ADD-A-FOLEY TRAY, NEEDLE SAMPLING, DRAIN TUBE, PREP TRAY; Dover" URINE DRAINAGE ADD-A-FOLEY TRAY, NEEDLE SAMPLING, A/R CHAMBER, DRAIN SPOUT, PREP TRAY; Dover" UNIVERSAL TRAY, 1000 ML BASIN, PVP SWAB STICKS, PREPPING COMPONENTS, 30 CC PRE-FILLED SYRINGE

    Convenience kits were manufactured and distributed with components which were subsequently recalled by Nurse Assist.

  • 2023-08-03Open, ClassifiedNonconforming Material/Component

    Custom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. LIVER 2018, Model Number: DYNJ905655A;

    Potential for the labeled sterile ultrasound gel component within packs to not meet sterility specifications.

  • 2022-12-02Open, ClassifiedEquipment maintenance

    Dover Closed Urethral Tray with Vinyl Catheter 14 Fr/Ch (4.7 ml) 5ml Catheter Product Code: 2480 Intended for urinary drainage from the bladder of a patient

    Occlusion (blockage) of the bubble inlet port of the urinary drainage bag which may prevent urine draining into the bag. If the bladder is not emptied within a clinically appropriate time, this may increase the risk of urinary retention

  • 2022-12-02Open, ClassifiedEquipment maintenance

    Dover Closed Urethral Tray with Hydrogel Coated Red Rubber Catheter 16 Fr/Ch (5.3 mm) Product Code: 3408 Intended for urinary drainage from the bladder of a patient

    Occlusion (blockage) of the bubble inlet port of the urinary drainage bag which may prevent urine draining into the bag. If the bladder is not emptied within a clinically appropriate time, this may increase the risk of urinary retention

  • 2022-12-02Open, ClassifiedEquipment maintenance

    Dover Closed Urethral Tray with Hydrogel Coated Red Rubber Catheter 14 Fr/Ch (4.7 mm) Product Code: 3410 Intended for urinary drainage from the bladder of a patient

    Occlusion (blockage) of the bubble inlet port of the urinary drainage bag which may prevent urine draining into the bag. If the bladder is not emptied within a clinically appropriate time, this may increase the risk of urinary retention

  • 2022-12-02Open, ClassifiedEquipment maintenance

    Dover Closed Urethral Tray with Vinyl Catheter Product Code: 3420 Intended for urinary drainage from the bladder of a patient

    Occlusion (blockage) of the bubble inlet port of the urinary drainage bag which may prevent urine draining into the bag. If the bladder is not emptied within a clinically appropriate time, this may increase the risk of urinary retention

  • 2022-12-02Open, ClassifiedEquipment maintenance

    Dover Closed Urethral Tray with Vinyl Catheter 14 Fr/Ch (4.7 mm), 5 mL Catheter Pre-connected to 1000 mL Drainage Bag Product Code: 3450 Intended for urinary drainage from the bladder of a patient

    Occlusion (blockage) of the bubble inlet port of the urinary drainage bag which may prevent urine draining into the bag. If the bladder is not emptied within a clinically appropriate time, this may increase the risk of urinary retention

  • 2022-12-02Open, ClassifiedEquipment maintenance

    Dover Closed Urethral Tray with Vinyl Catheter 14 Fr/Ch (4.7 mm), 5 mL Catheter Pre-connected to 1000 mL Drainage Bag Product Code: 3450LF Intended for urinary drainage from the bladder of a patient.

    Occlusion (blockage) of the bubble inlet port of the urinary drainage bag which may prevent urine draining into the bag. If the bladder is not emptied within a clinically appropriate time, this may increase the risk of urinary retention

  • 2022-08-26Open, ClassifiedEnvironmental control

    MEDLINE INDUSTRIES Medline Catheter Insertion Tray Medline Foley Without Catheter Without Balloon Model Number: DYND10160

    Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.

Monitor tray, catheterization, sterile urethral, with or without catheter (kit) automatically

DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.