Choledochoscope And Accessories, Flexible/Rigid
FDA adverse event reports and recall history for choledochoscope and accessories, flexible/rigid (product code FBN), from the public MAUDE corpus and the live FDA recall database.
- Total reports
- 2,380
- Deaths
- 10
- Injuries
- 132
- Malfunctions
- 2,236
Reports per year
Source: FDA adverse event reports by year received, 2011–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.
Recent recalls
- 2026-05-15Open, ClassifiedProcess control
Tangent Endoscopy, Tangent Single Use Digital Catheter. Model Numbers: (1) TNG4002-IND. (2) TNG-4007-6pk.
Affected devices exhibited fractures at the distal shaft tip without complete detachment.
- 2025-01-16Open, ClassifiedLabeling design
Olympus MAJ-891 Forceps/Irrigation Plug (Isolated Type). Model/Catalog Number: MAJ-891. This instrument is designed to be attached to the instrument channel port of Olympus endoscopes to allow both irrigation and the use of endo-therapy accessories.
Potential for inadvertent retention of biomaterial when the reprocessing procedure is not performed as instructed, so as not completely removing and disassembling the device.
- 2023-11-17Open, ClassifiedUnder Investigation by firm
Ultrasonic gastrovideoscope-Real time ultrasound imaging,ultrasound guided needle aspiration and other endoscopic procedures within the upper gastrointestinal tract and surrounding organs Model Number: TGF-UC180J.
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
- 2016-06-28TerminatedDevice Design
Choledocho-Nephroscope Intended to provide optical visualization of and therapeutic access to the Abdominal Cavity. This anatomy includes, but is not restricted to, the organs, tissues and subsystems: Biliary tree with access to the Common bile Duct, Hepatic Duct, Cystic Duct, Kidney, Renal Cavity and Renal Calyces. The instrument is introduced via percutaneous cut when indications consistent with the need for procedure are observed in adult and pediatric patient populations.
Pentax has become aware that operational/cleaning accessories and therapeutic devices (e.g., clips, stents, balloons, etc.) can become lodged in an endoscope's instrument channel.
- 2016-06-28TerminatedDevice Design
Choledochoscope Intended to provide optical visualization of and therapeutic access to the Biliary Tree. This anatomy includes, but is not restricted to, the organs, tissues and subsystems: Common Bile, Hepatic, Cystic and Pancreatic Ducts. The instrument is introduced via the working channel of a Duodenoscope positioned for access to the Common bile Duct when indications consistent with the need for procedure are observed in adult and pediatric patient populations.
Pentax has become aware that operational/cleaning accessories and therapeutic devices (e.g., clips, stents, balloons, etc.) can become lodged in an endoscope's instrument channel.
- TerminatedDevice Design
Olympus CHF-CB30S Choledoscope Intended Use: The device was designed to be used with an Olympus TV camera system, light source, TV monitor, endo-therapy accessories and other ancillary equipment for endoscopic diagnosis and treatment within the biliary tract (common bile duct, cystic duct, and hepatic duct).
The CHF-CB30S Device did not have a 510(k) clearance. Additionally, the device has been associated with breakage of the insertion section and protrusion of metal parts, as well as breakage and displacement of the rubber on the bending section during surgical procedures.
- TerminatedNonconforming Material/Component
Olympus DISPOSABLE GRASPING FORCEPS, FG-51D
Forceps do not comply with Olympus standards for the amount of force required to open and close the forceps
- TerminatedProcess design
SpyScope DS Access and Delivery diagnostic biliary catheter, M00546600
Certain lots of devices may have the working channel sleeve protruding from the camera cap which may result in injuries from minor tissue damage to significant bleeding.
- TerminatedUnder Investigation by firm
URF-P6 Uretero-reno fiberscope and URF-P6R Uretero-reno fiberscope
Olympus is recalling the URF-P6/P6R endoscopes due to a similar device (URF-V2/V2R endoscopes) but not identical, where adverse event complaints associated with tissue trauma, perforation and insertion tubes which were stuck inside of patients that had to be surgically removed. In an effort to mitigate a potential risk to patient health, Olympus is undertaking this action to notify users of these complaints and the need for careful inspection of the endoscope prior to use in accordance with the instructions provided int heir notification letter.
Monitor choledochoscope and accessories, flexible/rigid automatically
DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.