FBKClass 2

Endoscopic Injection Needle, Gastroenterology-Urology

FDA adverse event reports and recall history for endoscopic injection needle, gastroenterology-urology (product code FBK), from the public MAUDE corpus and the live FDA recall database.

Total reports
591
Deaths
0
Injuries
63
Malfunctions
520

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Recent recalls

  • 2024-11-15Open, ClassifiedProcess change control

    Posi-Stop Injection Needle, 23 Gauge, Catalog Number: 4720

    Injection solutions would not flow as intended through the tip of 25ga-sized needles. Investigations that some units contained debris from the manufacturing process within the inner sheath that blocked the ID of the needle.

  • 2024-11-15Open, ClassifiedProcess change control

    Posi-Stop Injection Needle, 23 Gauge, Catalog Number: 4702

    Injection solutions would not flow as intended through the tip of 25ga-sized needles. Investigations that some units contained debris from the manufacturing process within the inner sheath that blocked the ID of the needle.

  • 2024-11-15Open, ClassifiedProcess change control

    Posi-Stop Injection Needle, 23 Gauge, Catalog Number: 4706

    Injection solutions would not flow as intended through the tip of 25ga-sized needles. Investigations that some units contained debris from the manufacturing process within the inner sheath that blocked the ID of the needle.

  • 2024-11-15Open, ClassifiedProcess change control

    Posi-Stop Injection Needle, 23 Gauge, Catalog Number: 4716

    Injection solutions would not flow as intended through the tip of 25ga-sized needles. Investigations that some units contained debris from the manufacturing process within the inner sheath that blocked the ID of the needle.

  • 2024-11-15Open, ClassifiedProcess change control

    Posi-Stop Injection Needle, 23 Gauge, Catalog Number: 4721

    Injection solutions would not flow as intended through the tip of 25ga-sized needles. Investigations that some units contained debris from the manufacturing process within the inner sheath that blocked the ID of the needle.

  • 2024-11-15Open, ClassifiedProcess change control

    Posi-Stop Injection Needle, 25 Gauge, Catalog Number: 4704

    Injection solutions would not flow as intended through the tip of 25ga-sized needles. Investigations that some units contained debris from the manufacturing process within the inner sheath that blocked the ID of the needle.

  • 2024-11-15Open, ClassifiedProcess change control

    Posi-Stop Injection Needle, 25 Gauge, Catalog Number: 4718

    Injection solutions would not flow as intended through the tip of 25ga-sized needles. Investigations that some units contained debris from the manufacturing process within the inner sheath that blocked the ID of the needle.

  • 2024-11-15Open, ClassifiedProcess change control

    Posi-Stop Injection Needle, 25 Gauge, Catalog Number: 4722

    Injection solutions would not flow as intended through the tip of 25ga-sized needles. Investigations that some units contained debris from the manufacturing process within the inner sheath that blocked the ID of the needle.

  • 2024-11-15Open, ClassifiedProcess change control

    Posi-Stop Injection Needle, 25 Gauge, Catalog Number: 4723

    Injection solutions would not flow as intended through the tip of 25ga-sized needles. Investigations that some units contained debris from the manufacturing process within the inner sheath that blocked the ID of the needle.

  • 2023-12-18Open, ClassifiedDevice Design

    Injection Needle, 3 Fr., REF: 27030N; Injection Cannula, 6 Fr., REF: 27182A

    The efficacy of the manual cleaning process cannot be assured for the affected injection needles intended for general surgery and visceral surgery, which if used, may expose the patient to a higher risk of infection.