Driver, Wire, And Bone Drill, Manual
FDA adverse event reports and recall history for driver, wire, and bone drill, manual (product code DZJ), from the public MAUDE corpus and the live FDA recall database.
- Total reports
- 3,406
- Deaths
- 1
- Injuries
- 113
- Malfunctions
- 3,287
Reports per year
Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.
Recent recalls
- 2013-10-28TerminatedNonconforming Material/Component
REF 5100-088, TPS Micro Driver, For Use with Stryker Total Performance System Caution Federal Law (USA) restricts the device to use by or on the order of a physician Driver, Wire and Bone Drill; The Stryker Total Performance System (TPS) is intended for use in the cutting, drilling, reaming, decorticating and smoothing of bone and other bone related tissues for a variety of applications such as ENT, dental, orthopedic, maxillofacial, spinal, and plastic surgery. The instruments are also used in the placement of screws, wires, pins, and other fixation devices.
It was discovered that the silicone potting failed to properly cure after undergoing the prescribed curing possess. The potting material is used to insulate, stabilize and provide a moisture barrier to the TPS MicroDriver main Printed Circuit Board Assembly (PCBA). Inadequate cure of the silicone potting could potentially allow it to leak to the outside of the handpiece or cause an electrical ma
- 2011-11-16TerminatedDevice Design
IBO Blade, Right, for use with Stryker TPS Reciprocating Saw, REF 5100-37-901, Stryker Instruments, Kalamazoo, MI. Product Usage: The intended use of these devices is to cut bone and hard tissue in maxillofacial procedures and have been specifically designed for osteotomies of the mandibular ramus.
There is a potential for these blades to break where the blade is welded on to the arbor. This hazard can result in the following harms: tissue damage due to the broken blade being aspirated or swallowed, additional surgery to remove a broken blade, infection due to an unretrieved device component, complications associated with increased time under anesthesia to retrieve a broken blade, osteomyeli
- 2011-11-16TerminatedDevice Design
IBO Blade, Left, for use with Stryker TPS Reciprocating Saw, REF 5100-37-902, Stryker Instruments, Kalamazoo, MI. Product Usage: The intended use of these devices is to cut bone and hard tissue in maxillofacial procedures and have been specifically designed for osteotomies of the mandibular ramus.
There is a potential for these blades to break where the blade is welded on to the arbor. This hazard can result in the following harms: tissue damage due to the broken blade being aspirated or swallowed, additional surgery to remove a broken blade, infection due to an unretrieved device component, complications associated with increased time under anesthesia to retrieve a broken blade, osteomyeli
- 2011-11-16TerminatedDevice Design
Right/Left IBO Blade Set, for use with Stryker TPS Reciprocating Saw, REF 5100-37-903, Stryker Instruments, Kalamazoo, MI. Product Usage: The intended use of these devices is to cut bone and hard tissue in maxillofacial procedures and have been specifically designed for osteotomies of the mandibular ramus.
There is a potential for these blades to break where the blade is welded on to the arbor. This hazard can result in the following harms: tissue damage due to the broken blade being aspirated or swallowed, additional surgery to remove a broken blade, infection due to an unretrieved device component, complications associated with increased time under anesthesia to retrieve a broken blade, osteomyeli
- 2011-11-16TerminatedDevice Design
Right IBO Blade , for use with Stryker TPS Reciprocating Saw, REF 5100-37-901S2, Stryker Instruments, Kalamazoo, MI. Product Usage: The intended use of these devices is to cut bone and hard tissue in maxillofacial procedures and have been specifically designed for osteotomies of the mandibular ramus.
There is a potential for these blades to break where the blade is welded on to the arbor. This hazard can result in the following harms: tissue damage due to the broken blade being aspirated or swallowed, additional surgery to remove a broken blade, infection due to an unretrieved device component, complications associated with increased time under anesthesia to retrieve a broken blade, osteomyeli
- 2011-11-16TerminatedDevice Design
Left IBO Blade , for use with Stryker TPS or Core Reciprocating Saw, REF 5100-37-902S2, Stryker Instruments, Kalamazoo, MI. Product Usage: The intended use of these devices is to cut bone and hard tissue in maxillofacial procedures and have been specifically designed for osteotomies of the mandibular ramus.
There is a potential for these blades to break where the blade is welded on to the arbor. This hazard can result in the following harms: tissue damage due to the broken blade being aspirated or swallowed, additional surgery to remove a broken blade, infection due to an unretrieved device component, complications associated with increased time under anesthesia to retrieve a broken blade, osteomyeli
- 2006-03-09TerminatedOther
Thermoplastic templates, single tooth Catalog Number: 040.526
Pilot drills may not fully extend through drill guide in template products
- 2006-03-09TerminatedOther
Thermoplastic templates, free-end situation Catalog Number: 040.527
Pilot drills may not fully extend through drill guide in template products
- 2005-08-23TerminatedOther
Stryker brand Consolidated Operating Room Equipment (CORE) System Powered Instrument Driver, REF 5400-50; Model 5400-050-000.
Potential electrical shock hazard, in that the potential exists for the patient to be exposed to an earth-referenced electrical current.
- 2005-08-23TerminatedOther
Stryker brand Consolidated Operating Room Equipment (CORE) SystemPowered Instrument Driver, REF 5400-50; Model 5400-050-000A.
Potential electrical shock hazard, in that the potential exists for the patient to be exposed to an earth-referenced electrical current.
Monitor driver, wire, and bone drill, manual automatically
DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.