FDA adverse event reports and recall history for system, thermal regulating (product code DWJ), from the public MAUDE corpus and the live FDA recall database.

Total reports
26,820
Deaths
54
Injuries
2,197
Malfunctions
24,379

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Recent recalls

  • 2026-08-14Open, ClassifiedDevice Design

    Blanketrol II Model 222S; Thermal regulating system.

    Overheating occurring on 115V CPU boards for the Blanketrol III Model 233, Innercool CoolBlue Model STx, and Blanketrol II Model 222S products may cause the device to shut down.

  • 2026-08-14Open, ClassifiedDevice Design

    Innercool Coolblue Console, Model STx; Thermal regulating system.

    Overheating occurring on 115V CPU boards for the Blanketrol III Model 233, Innercool CoolBlue Model STx, and Blanketrol II Model 222S products may cause the device to shut down.

  • 2026-08-14Open, ClassifiedDevice Design

    Blanketrol III, Model 233; Thermal regulating system.

    Overheating occurring on 115V CPU boards for the Blanketrol III Model 233, Innercool CoolBlue Model STx, and Blanketrol II Model 222S products may cause the device to shut down,

  • 2025-06-12Open, ClassifiedDevice Design

    Allon 2001. Part Number: 200-00263

    An investigation into a report of the patient wrap (ThermoWrap), used with the Allon device, exceeding the set temperature determined that if: 1. The Human Sensor Board is disconnected from the Controller Board in the Allon device AND the temperature alarms are disabled, or 2. The Water In and Water Out and Thermostat sensors are disconnected from the Human Sensor Board AND the temperature alarms are disabled, the device will proceed to heat the water above the hardware fault limit without notifying the user of the faulty condition.

  • 2025-02-20Open, ClassifiedDevice Design

    Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 82 cm (32"), REF US501, to prevent or treat hypothermia

    There is over-molding that stabilizes the connection between the cable and cable-entry on the mattress. It turns out that after about a year of use, the connector cable on the mattress can experience the same type of deflection that caused the intermittent connection with the cables connecting to the ESU.

  • 2025-02-20Open, ClassifiedDevice Design

    Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 127 cm (50"), REF US502, to prevent or treat hypothermia

    There is over-molding that stabilizes the connection between the cable and cable-entry on the mattress. It turns out that after about a year of use, the connector cable on the mattress can experience the same type of deflection that caused the intermittent connection with the cables connecting to the ESU.

  • 2025-02-20Open, ClassifiedDevice Design

    Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 89 cm (35"), REF US530, to prevent or treat hypothermia

    There is over-molding that stabilizes the connection between the cable and cable-entry on the mattress. It turns out that after about a year of use, the connector cable on the mattress can experience the same type of deflection that caused the intermittent connection with the cables connecting to the ESU.

  • 2025-02-14Open, ClassifiedDevice Design

    Augustine Surgical Inc. HotDog Return Electrode accessory connector cables, Model A136 used in conjunction with Warming Blankets and Mattresses.

    Users were experiencing the following error message (M-0B-125) stated Check Neutral Electrode Contact .

  • 2025-02-14Open, ClassifiedDevice Design

    Augustine Surgical Inc. HotDog Return Electrode accessory connector cables, Model A138 used in conjunction with Warming Blankets and Mattresses.

    Users were experiencing the following error message (M-0B-125) stated Check Neutral Electrode Contact .

  • 2025-01-10CompletedDevice Design

    Augustine Surgical Inc. HOT DOG PATIENT WARMING, Temperature Management Controller, Single Port, Models: 1) Model # WC71, 2) Model # WC71R (refurbished); used in conjunction with Warming Blankets and Mattresses.

    HotDog Temperature Management Controllers Models WC71, WC77, and Temperature Management Controller MP Model 2083516 may experience a Power-on self-test error code "EA POST".

Monitor system, thermal regulating automatically