Skip to main content
DTMClass 2

Filter, Blood, Cardiopulmonary Bypass, Arterial Line

FDA adverse event reports and recall history for filter, blood, cardiopulmonary bypass, arterial line (product code DTM), from the public MAUDE corpus and the live FDA recall database.

Total reports
261
Deaths
7
Injuries
38
Malfunctions
197

Reports per year

10
11
12
13
14
15
16
17
18
19
20
21
22
24
26

Source: FDA adverse event reports by year received, 2010–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Recent recalls

  • 2013-09-20TerminatedNonconforming Material/Component

    Terumo Pall AL6X Arterial Blood Line Filter, labeled in part Terumo Cardiovascular Systems Corporation, packaged as part of a Terumo Cardiovascular Procedure Kit or as a single, sterile filter unit Single/Sterile Terumo Pall AL6X Arterial Filter: The Terumo Pall AL6X Arterial Filter is indicated for use in cardiopulmonary bypass procedures for the removal of micro-emboli greater than 40 microns in size, including gas emboli, fat emboli and aggregates composed of platelets, red blood cells and other debris from the arterial line and where the flow rate will not exceed 8 liters per minute. Cardiovascular Procedure (Convenience) Kit that includes the Terumo Pall AL6X Arterial Filter: The Cardiovascular Procedure Kits are indicated for use only in the extracorporeal circuit for the cardiopulmonary bypass procedures for which the user designed. The sterile cardiovascular procedure kit is intended to be used one time for periods up to 6 hours.

    Medical device used in cardiovascular procedures is defective.

  • 2013-08-14TerminatedEmployee error

    Medtronic Affinity Pediatric Arterial Filter, model number S4014, sterilized using ethylene oxide. Recall Update 9/6/13: Medtronic, Red cross Infant Pack 1/B, South-Africa, REF M272403C, Sterile EO. The Affinity Pediatric Arterial Filters are single-use, sterile, nonpyrogenic devices designed to filter microemboli greater than the specified micron size from the circuit for periods up to 6 hours during cardiopulmonary bypass surgery. These devices are available both in an uncoated and a Carmeda coated option. The Carmeda coating is a BioActive surface that is non-leaching and provides a thromboresistant blood contact surface.

    Medtronic is recalling 148 Affinity Pediatric Arterial Filters from 2 manufacturing lots because a limited quantity of filters may have a small breach in the filter media that may allow unwanted particulate or gaseous emboli to pass through the filter and the outlet of the device, which could result in serious injury to a pediatric patient including neurological damage, or potential patient death.

  • 2008-07-24TerminatedNonconforming Material/Component

    Medtronic Affinity Pediatric arterial Filter, Arterial Line Filter for Cardiopulmonary Bypass. Catalog Number: CB4014

    Medtronic is recalling selected Carmeda-coated cardiopulmonary bypass (CPB) products (e.g., oxygenators, reservoirs, pumps, cannulae and tubing packs) that were manufactured with contaminated heparin. Medtronic's heparin supplier recently reported that it had provided some batches to Medtronic that were contaminated with oversulfated chondroitin sulfate (OSCS).

  • 2005-12-02TerminatedOther

    Custom heart lung perfusion packs

    One way valve in the heart lung pack assembly may be missing.

  • TerminatedProcess control

    Medtronic Affinity, 38um Arterial Filter, Catalog number REF 351

    Medtronic has identified an out of-specification condition exhibiting excess plastic (flash) in the arterial filter directly above the outlet port.

Monitor filter, blood, cardiopulmonary bypass, arterial line automatically

DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.