Pulse-Generator, Pacemaker, External
FDA adverse event reports and recall history for pulse-generator, pacemaker, external (product code DTE), from the public MAUDE/AEMS corpus and the live FDA recall database.
- Total reports
- 12,981
- Deaths
- 85
- Injuries
- 1,001
- Malfunctions
- 11,827
Reports per year
Source: FDA adverse event reports by year received, 2012–2026. Reporting paused during the 2026 MAUDE-to-AEMS transition and has since resumed.
Recent recalls
- 2018-06-29TerminatedDevice Design
Single Chamber Temporary External Pacemaker, Model 53401
A subset of Medtronic Model 53401 Single Chamber External Pulse Generators (EPGs) may revert from programmed settings to nominal settings during use with a patient due to an unexpected Power on Reset (POR).
- 2015-03-31TerminatedDevice Design
Medtronic Dual Chamber Temporary External Pacemaker, Model 5392. Product Usage: The temporary pacemaker is used with a cardiac pacing lead system for temporary single or dual chamber pacing in a clinical environment by trained personnel. The temporary pacemaker can be used where short-term demand (synchronous) or asynchronous pacing is indicated for therapeutic, prophylactic, or diagnostic purposes. The temporary pacemaker must be used in an environment where the patient is monitored continuously to ensure that it is operating properly and delivering appropriate therapy to the patient. Specific indications for temporary cardiac pacing include, but are not limited to, the following: - Complete heart block - Sinus bradycardia - Sick sinus syndrome - Bradycardia with congestive heart failure - Atrial and/or ventricular arrhythmias - Cardiac arrest - Support, management, and evaluation of a patient before permanent pacemaker implantation - Support during permanent pacemaker replacement - Cardiac complications during invasive or surgical procedures - Support following cardiac surgery - Acute myocardial infarction complicated by heart block - Atrial tachyarrhythmias that require high-rate burst pacing for treatment
Possible performance issue when used with specific AA-sized (LR6) batteries. The negative terminal of a commercially available AA (LR6) battery did not maintain a sufficient connection with the battery drawer electrical contact. This issue could prevent the EPG from powering on or cause the EPG to abruptly lose primary battery power, potentially stopping delivery of pacing therapy.
- 2010-09-21TerminatedDevice Design
Oscor PACE 203H, Dual-Chamber External Pulse Generator The PACE 203H / Model 3085 external pulse generator / temporary cardiac pacemaker is designed to be used with cardiac pacing lead systems for temporary atrial or ventricular or A-V sequential pacing. The PACE 203H / Model 3085 has applications where such pacing modes are indicated for therapeutic, prophylactic, or diagnostic purposes.
The recall was initiated after Osypka Medical received two (2) complaints from their Japanese distributor regarding an incompatibility of the PACE 203H with some new types of 9 V batteries. The investigation revealed that PACE 203H of a certain manufacturing period malfunction (i.e., not stimulate) when a type of 9 V batteries with higher than usual open circuit voltage (OCV) is used. Failure t
- 2010-09-21TerminatedDevice Design
St. Jude Medical, Model 3085, Dual-Chamber External Pulse Generator The PACE 203H / Model 3085 external pulse generator / temporary cardiac pacemaker is designed to be used with cardiac pacing lead systems for temporary atrial or ventricular or A-V sequential pacing. The PACE 203H / Model 3085 has applications where such pacing modes are indicated for therapeutic, prophylactic, or diagnostic purposes.
The recall was initiated after Osypka Medical received two (2) complaints from their Japanese distributor regarding an incompatibility of the PACE 203H with some new types of 9 V batteries. The investigation revealed that PACE 203H of a certain manufacturing period malfunction (i.e., not stimulate) when a type of 9 V batteries with higher than usual open circuit voltage (OCV) is used. Failure t
- TerminatedProcess control
ZOLL Pro-padz Liquid Gel Radiolucent, Part numbers: 8900-2105-01 (12/case) and 8900-2106-01 (single)
Some of the electrodes may have been assembled incorrectly, preventing delivery of defibrillation or pacing therapy.
Monitor pulse-generator, pacemaker, external automatically