DSAClass 2

Cable, Transducer And Electrode, Patient, (Including Connector)

FDA adverse event reports and recall history for cable, transducer and electrode, patient, (including connector) (product code DSA), from the public MAUDE corpus and the live FDA recall database.

Total reports
632
Deaths
17
Injuries
95
Malfunctions
503

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Recent recalls

  • 2025-02-05Open, ClassifiedLabeling Change Control

    LIFEPAK 35 ECG cable REF 11111-000041

    Product shipped with incorrect IFU. IFU does not include proper cleaning/disinfection instructions

  • 2014-04-09TerminatedLabeling Change Control

    FAS-LOC 6 Foot Disposable Extension Cable with Safe Connect/Screw Down. Electrical extension cable intended to transmit signal from, or power or excitation signal to patient connected electrodes. The cable is a sterile, disposable device.

    Remington Medical discovered an error on the FL-601-97 IFU; part number IFU-RM-0154. Specifically, the polarity for the black chuck in the picture is wrong. It shows (+), but should show (-).

  • 2011-01-07TerminatedDevice Design

    CARTO 3 System Interface Cable, 34-pin to 10-pin, Red, 10' - Part Number: D-1286-18; Catalog Number CR3410CT This product provides a signal interface between a Biosense Webster electrophysiology catheter and the appropriate equipment. This cable may be re-used subject to established cleaning and sterilization restrictions

    When connecting this cable to the mapping system, loss of Body Surface ECG signals may occur. This may result in inaccurate locations appearing for catheters displayed 3D on the CARTO monitor.

  • 2010-08-13TerminatedProcess design

    Disposable Surgical Cable, Product Code: FL-601-97 (100), Single use, Rx Only, Remington Medical, Inc., 6830 Meadowridge Court, Alpharetta, GA 30005 Usage: Temporary pacing of open hearts: used with External Pacing Generators (EPGs).

    The Safe Connect component may separate from the cable, causing the cable to become unplugged from the EPG.

  • Open, ClassifiedDevice Design

    MM-TSM (Touch Screen Module/Secondary Controller/Bedside Controller), Part Numbers: 300002018101, 300002018102, used with IntraSight Integrated (Part Number: 797403) IVUS systems

    Multi-Modality Touch Screen Modules (MM-TSM), used with integrated IVUS systems, may crack, or become damaged, which may affect their usability. They could become non-functional. Therefore, bottom protective beams will be replaced with full protective metal brackets.

  • TerminatedProcess design

    ECG Trunk Cables: REF/Catalog Number Description 2106305-001 ECG Trunk Cable, 3/5-lead, AHA, 3.6 m/12 ft. 2106305-002 ECG Trunk Cable, 3/5-lead, AHA, 1.2 m/4 ft. 2106305-003 ECG Trunk Cable, 3/5-lead, IEC, 3.6 m/12 ft. 2106305-004 ECG Trunk Cable, 3/5-lead, IEC, 1.2 m/4 ft. 2106306-001 ECG Trunk Cable, Neonatal, DIN 3-lead, AHA, 3.6 m/12 ft. 2106306-002 ECG Trunk Cable, Neonatal, DIN 3-lead, AHA, 1.2 m/4 ft. 2106306-003 ECG Trunk Cable, Neonatal, DIN 3-lead, IEC, 3.6 m/12 ft. 2106306-004 ECG Trunk Cable, Neonatal, DIN 3-lead, IEC, 1.2 m/4 ft. 2106307-001 ECG Trunk Cable, 6-lead, AHA, 3.6 m/12 ft. 2106307-002 ECG Trunk Cable, 6-lead, AHA, 1.2 m/4 ft. 2106307-003 ECG Trunk Cable, 6-lead, IEC, 3.6 m/12 ft. 2106307-004 ECG Trunk Cable, 6-lead, IEC, 1.2 m/4 ft. 2106308-001 ECG Trunk Cable, 12-lead, AHA, 3.6 m/12 ft. 2106308-002 ECG Trunk Cable, 12-lead, AHA, 1.2 m/4 ft. 2106308-003 ECG Trunk Cable, 12-lead, IEC, 3.6 m/12 ft. 2106308-004 ECG Trunk Cable, 12-lead, IEC, 1.2 m/4 ft. 2106310-001 ECG Trunk Cable, 3/5-lead w/ESU filter, AHA, 3.6 m/12 ft. 2106310-002 ECG Trunk Cable, 3/5-lead w/ESU filter, IEC, 3.6 m/12 ft. 2106311-001 ECG Trunk Cable, 3/5-lead, AHA, 6 m/20 ft. 2106311-002 ECG Trunk Cable, 3/5-lead, IEC, 6 m/20 ft. The reusable ECG trunk cable enables ECG monitoring with GE monitoring equipment requiring rectangular 11-pin connector cable.

    ECG trunk cables and leadwires may reduce the amount of energy reaching the patient during defibrillation, potentially limiting success of defibrillating the patients rhythm.

  • TerminatedProcess design

    ECG leadwire sets: REF/Catalog Number Description 2106389-001 ECG Leadwire set, 5-lead, grouped, grabber, AHA, 74 cm/ 29 in 2106389-002 ECG Leadwire set, 5-lead, grouped, grabber, AHA, 130 cm/ 51 in 2106389-003 ECG Leadwire set, 5-lead, grouped, grabber, IEC, 74 cm/ 29 in 2106389-004 ECG Leadwire set, 5-lead, grouped, grabber, IEC, 130 cm/ 51 in 2106389-005 ECG Leadwire set, 5-lead, grouped, grabber, AHA, mix 74 cm/ 29 in, 130 cm/ 51 in 2106389-006 ECG Leadwire set, 5-lead, grouped, grabber, IEC, mix 74 cm/ 29 in, 130 cm/ 51 in 2106390-001 ECG Leadwire set, 3-lead, grabber, AHA, 74 cm/ 29 in 2106390-002 ECG Leadwire set, 3-lead, grabber, AHA, 130 cm/ 51 in 2106390-003 ECG Leadwire set, 3-lead, grabber, IEC, 74 cm/ 29 in 2106390-004 ECG Leadwire set, 3-lead, grabber, IEC, 130 cm/ 51 in 2106391-001 ECG Leadwire set, 5-lead, grabber, AHA, 74 cm/ 29 in 2106391-002 ECG Leadwire set, 5-lead, grabber, AHA, 130 cm/ 51 in 2106391-003 ECG Leadwire set, 5-lead, grabber, IEC, 74 cm/ 29 in 2106391-004 ECG Leadwire set, 5-lead, grabber, IEC, 130 cm/ 51 in 2106391-005 ECG Leadwire set, 5-lead, grabber, AHA, mix 74 cm/ 29 in, 130 cm/ 51 in 2106391-006 ECG Leadwire set, 5-lead, grabber, IEC, mix 74 cm/ 29 in, 130 cm/ 51 in 2106393-001 ECG Leadwire set, 5-lead V2-6, grabber, AHA, 130 cm/ 51 in 2106393-002 ECG Leadwire set, 5-lead C2-6, grabber, IEC, 130 cm/ 51 in 2106393-003 ECG Leadwire set, 5-lead V2-6, grabber, AHA, 74 cm/ 29 in 2106393-004 ECG Leadwire set, 5-lead C2-6, grabber, IEC, 74 cm/ 29 in 2106397-001 ECG Leadwire set, 6-lead, grouped, grabber, AHA, 74 cm/ 29 in 2106397-002 ECG Leadwire set, 6-lead, grouped, grabber, AHA, 130 cm/ 51 in 2106397-003 ECG Leadwire set, 6-lead, grouped, grabber, AHA, mix 74 cm/ 29 in, 130 cm/ 51 in 2106398-001 ECG Leadwire set, 6-lead, grouped, grabber, IEC, 74 cm/ 29 in 2106398-002 ECG Leadwire set, 6-lead, grouped, grabber, IEC, 130 cm/ 51 in The reusable ECG leadwire sets enable ECG monitoring with GE equipment requiring rectangular 11-pin ECG trunk cable. Use leadwire sets with ECG trunk cable or telemetry transmitter.

    ECG trunk cables and leadwires may reduce the amount of energy reaching the patient during defibrillation, potentially limiting success of defibrillating the patients rhythm.

  • TerminatedProcess design

    Replacement leadwires with grabber ends: REF/Catalog Number Description 2106388-005 Replacement ECG Leadwire, grabber, BRN V, AHA, 74 cm/ 29 in 2106388-101 Replacement ECG Leadwire, grabber, LL , AHA, 130 cm/ 51 in 2106388-103 Replacement ECG Leadwire, grabber, GRN RL, AHA, 130 cm/ 2106388-107 Replacement ECG Leadwire, grabber, BLK N, IEC, 130 cm/ 51 in 2106392-004 Replacement ECG Leadwire, grabber, ORG V, AHA, 74 cm/ 29 in The reusable ECG leadwire sets enable ECG monitoring with GE equipment requiring rectangular 11-pin ECG trunk cable. Use leadwire sets with ECG trunk cable or telemetry transmitter.

    ECG trunk cables and leadwires may reduce the amount of energy reaching the patient during defibrillation, potentially limiting success of defibrillating the patients rhythm.

  • TerminatedProcess design

    Leadwire sets with snap ends: REF/Catalog Number Description 2106381-001 ECG Leadwire set, 5-lead, snap, AHA, 74 cm/ 29 in 2106381-002 ECG Leadwire set, 5-lead, snap, AHA, 130 cm/ 51 in 2106381-003 ECG Leadwire set, 5-lead, snap, IEC, 74 cm/ 29 in 2106381-004 ECG Leadwire set, 5-lead, snap, IEC, 130 cm/ 51 in 2106381-005 ECG Leadwire set, 5-lead, snap, AHA, mix 74 cm/ 29 in, 130 cm/ 51 in 2106383-001 ECG Leadwire set, 5-lead, grouped, snap, AHA, 74 cm/ 29 in 2106383-002 ECG Leadwire set, 5-lead, grouped, snap, AHA, 130 cm/ 51 in 2106383-003 ECG Leadwire set, 5-lead, grouped, snap, IEC, 74 cm/ 29 in 2106383-004 ECG Leadwire set, 5-lead, grouped, snap, IEC, 130 cm/ 51 in 2106383-005 ECG Leadwire set, 5-lead, grouped, snap, AHA, mix 74 cm/ 29 in, 130 cm/ 51 in 2106383-006 ECG Leadwire set, 5-lead, grouped, snap, IEC, mix 74 cm/ 29 in, 130 cm/ 51 in 2106385-001 ECG Leadwire set, 3-lead, snap, AHA, 74 cm/ 29 in 2106385-002 ECG LEADWIRE SET, 3-LD GROUPED, SNAP, AHA, 130 CM/ 51 IN 2106385-003 ECG LEADWIRE SET, 3-LD GROUPED, SNAP, IEC, 74 CM/ 29 IN 2106385-004 ECG LEADWIRE SET, 3-LD GROUPED, SNAP, IEC, 130 CM/ 51 IN 2106394-003 ECG Leadwire set, 5-lead V2-6, snap, AHA, 74 cm/ 29 in 2106395-001 ECG Leadwire set, 6-lead, grouped, snap, AHA, 74 cm/ 29 in 2106396-001 ECG Leadwire set, 6-lead, grouped, snap, IEC, 74 cm/ 29 in 2106396-002 ECG Leadwire set, 6-lead, grouped, snap, IEC, 130 cm/ 51 in The reusable ECG leadwire sets enable ECG monitoring with GE equipment requiring rectangular 11-pin ECG trunk cable. Use leadwire sets with ECG trunk cable or telemetry transmitter.

    ECG trunk cables and leadwires may reduce the amount of energy reaching the patient during defibrillation, potentially limiting success of defibrillating the patients rhythm.

  • TerminatedProcess design

    Replacement ECG leadwires with snap ends REF/Catalog Number Description 2106386-101 Replacement ECG Leadwire, snap, LL, AHA, 130 cm/ 51 in The reusable ECG leadwire sets enable ECG monitoring with GE equipment requiring rectangular 11-pin ECG trunk cable. Use leadwire sets with ECG trunk cable or telemetry transmitter.

    ECG trunk cables and leadwires may reduce the amount of energy reaching the patient during defibrillation, potentially limiting success of defibrillating the patients rhythm.

Monitor cable, transducer and electrode, patient, (including connector) automatically

DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.