Catheter, Steerable

FDA adverse event reports and recall history for catheter, steerable (product code DRA), from the public MAUDE corpus and the live FDA recall database.

Total reports
14,164
Deaths
255
Injuries
5,116
Malfunctions
8,787

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Recent recalls

  • 2026-07-20Open, ClassifiedUse error

    FARADRIVE Steerable Sheath Clear IDE, used to facilitate access to the vasculature and into the chambers of the heart, Material Number 21C602, IFU Reference Number 51636751-01

    Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body.

  • 2026-07-20Open, ClassifiedUse error

    FARADRIVE STEERABLE SHEATH CLEAR, used to facilitate access to the vasculature and into the chambers of the heart, Material Number 21M402, IFU Reference Number LBL1062

    Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body.

  • 2026-07-20Open, ClassifiedUse error

    FARADRIVE STEERABLE SHEATH (CLEAR) - US, used to facilitate access to the vasculature and into the chambers of the heart, Material Number M004PF21M402, IFU Reference Number 51654495-01

    Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body.

  • 2026-07-20Open, ClassifiedUse error

    FARADRIVE STEERABLE SHEATH CLEAR OUS, used to facilitate access to the vasculature and into the chambers of the heart, Material Number M004PFCE21M402, IFU Reference Number 51659727-01

    Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body.

  • 2024-03-26Open, ClassifiedProcess control

    Boston Scientific POLARSHEATH Steerable Sheath 12F, REF M004CRBS3050

    Boston Scientific has identified a tooling error in manufacturing which may have caused delamination of the inner lumen of the sheath shaft in a subset of POLARSHEATH devices. This could result in the embolization of a fragment during use (such as flushing of the sheath or introduction of the dilator or ablation catheter).

  • 2012-09-26TerminatedNonconforming Material/Component

    Boston Scientific, Z Flex 270, Steerable Sheath, Sterile EO, Rx only, Distributed by Boston Scientific Corporation, One Boston Scientific Place, Natick, MA 01760-1537, Manufactured by Greathbatch Medical, 2300 Berkshire Lane North, Minneapolis, MN 55441. To facilitate transvenous introduction of diagnostic / therapeutic catheters into the vasculature and into the chambers of the heart.

    During the investigation of four returned Z Flex 270 Steerable Sheaths that were used with cardiac cryoablation balloon catheters, the internal PTFE sheath liners were found to display varying degrees of damage that could pose an embolic risk to a patient.

  • 2011-09-12TerminatedPending

    Medtronic FlexCath, model 3FC12, Steerable Sheath. The FlexCath Steerable Sheath is a percutaneous catheter introducer fitted with a valve to allow for introduction, withdrawal, and swapping of catheters and wires while preventing air ingress and minimizing blood loss. A side-port with stopcock is integrated to allow continuous drip infusion, injection through the center lumen, flushing, aspiration, blood sampling and pressure monitoring. Rx only. Sterilized using ethylene oxide. Manufacturer: Medtronic CryoCath LP, Kirkland, Quebec H9H 5H3, Canada. Product Usage: The FlexCath Steerable Sheath is intended for percutaneous catheter introduction into the vasculature and into the chambers of the heart. The sheath deflection facilitates catheter positioning.

    Medtronic CryoCath would like to inform you of an observation related to a potential leak in the hemostatic valve of the Flex Cath 12 Steerable Sheath, Model 3FC12. There wave been reports of blood and saline solution leaking out of, and air ingress into the hemostatic valve. This includes reported instances of the introduction of air bubbles through the valve during aspiration of the side port

  • 2008-09-11TerminatedDevice Design

    Artisan Control Catheter, model number 04454, manufactured by Hansen Medical, Inc., Mountain View, CA. Steerable guide catheter for human use intended to facilitate manipulation, positioning and control for collecting electrophysiological data within the heart atria.

    The catheter's flexible bellows portion may develop a leak. This has the potential to cause loss of hemostasis, flush fluid leakage, and/or introduction of air into the catheter with a risk of subsequent embolism.

  • Open, ClassifiedProcess control

    DLP Left Heart Vent Catheters, 16 FR. Sterile, Rx Only

    Potential for a wire protrusion through the left heart vent catheter tip

  • Open, ClassifiedProcess control

    DLP Left Heart Vent Catheters, 18 FR. Sterile, Rx Only

    Potential for a wire protrusion through the left heart vent catheter tip

Monitor catheter, steerable automatically