Lead, Atomic Absorption
FDA adverse event reports and recall history for lead, atomic absorption (product code DOF), from the public MAUDE corpus and the live FDA recall database.
- Total reports
- 723
- Deaths
- 0
- Injuries
- 2
- Malfunctions
- 720
Reports per year
Source: FDA adverse event reports by year received, 2004–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.
Recent recalls
- 2026-04-17Open, ClassifiedDevice Design
(1)(2) LeadCare II Blood Lead Test Kit, (3)LeadCare II 5X Blood Lead Test Kit, (4)LeadCare II 9X Blood Lead Test Kit ; Catalog Number(s): (1) 70-6762, (2) 70-6762RUO, (3) 70-8404 (Box containing 5 Test Kits), (4) 70-9000 (Box containing 9 test kits); The LeadCare II kit is provided with the following materials: " Sensors (2 Containers of 24 sensors each; 48 tests) " Treatment Reagent Tubes (2 packages of 24 tubes each; 48 tests) " Heparinized Capillary Tubes/Plungers (50 @) " Transfer Droppers (50 @) " Calibration Button (1 @) " Lead Control Level 1 (1 @ 2 mL) " Lead Control Level 2 (1 @ 2 mL)
Customers reported observing unexpectedly elevated results when using certain third-party micro-collection devices to collect capillary samples for testing with the LeadCare Test Systems. False positive lead results may cause delayed results and additional unnecessary testing.
- 2026-04-17Open, ClassifiedDevice Design
LeadCare Plus Blood Lead Test Kit; Catalog Number: 82-0004; The LeadCare Plus kit is provided with the following materials: " Sensors (4 containers of 24 each; 96 tests) " Treatment Reagent Tubes (4 packages of 24 each; 96 tests) " Calibration Button (1 @) " Lead Control Level 1 (1 @ 2 mL) " Lead Control Level 2 (1 @ 2 mL)
Customers reported observing unexpectedly elevated results when using certain third-party micro-collection devices to collect capillary samples for testing with the LeadCare Test Systems. False positive lead results may cause delayed results and additional unnecessary testing.
- 2026-04-17Open, ClassifiedDevice Design
LeadCare Ultra Blood Lead Test Kit; Catalog Number: 70-8098; The LeadCare Ultra kit is provided with the following materials: " Sensors (8 containers of 24 each; 192 tests) " Treatment Reagent Tubes (8 containers of 24 each; 192 tests) " Calibration Button (1 @) " Lead Control Level 1 (1 @ 2 mL) " Lead Control Level 2 (1 @ 2 mL)
Customers reported observing unexpectedly elevated results when using certain third-party micro-collection devices to collect capillary samples for testing with the LeadCare Test Systems. False positive lead results may cause delayed results and additional unnecessary testing.
- 2025-11-21Open, ClassifiedLabeling mix-ups
The LeadCare II Blood Lead Test System relies on electrochemistry and a unique Sensor to detect lead in whole blood. The LeadCare II Test Kit is provided with the following materials: " Sensors (2 Containers of 24 sensors each; 48 tests) " Treatment Reagent Tubes (2 packages of 24 tubes each; 48 tests) " Heparinized Capillary Tubes/Plungers (50) " Transfer Droppers (50) " Calibration Button (1) " Lead Control Level 1 (1 @ 2 mL) " Lead Control Level 2 (1 @ 2 mL)
Calibration Buttons being used in the kitting process were incorrectly labeled as Lot 2323M for lot 2333M. The use of mislabeled calibration Button with Lot #2333 can lead to delayed or erroneous high or erroneous low results.
- 2023-11-03Open, ClassifiedNonconforming Material/Component
LeadCare II Blood Lead Test Kit-For in vitro diagnostic use only. The test kit is for the quantitative measurement of lead in capillary whole blood.
During an internal audit a small percentage (<10%) of plastic caps used to close the treatment reagent tubes were found to not meet the minimum dimensional lip requirement and may have inadequate cap-to-tube seals. Manufacturing with these caps may result in cap-to-tube seals that allow leaking of HCl or may result in cap-to-tube seals that allow evaporation of treatment reagent. The consequence of this would be a change in the volume of treatment reagent and a change in the ratios of either added patient sample or control material to treatment reagent. The result of this might be a falsely elevated result (e.g., a patient false positive [PATFP] result or a Controls Out of Range-High [COOR-Hi] result).
- 2022-08-26Open, ClassifiedEnvironmental control
MAGELLAN DIAGNOSTICS LeadCare II Blood Lead Test Kit Rapid Test Kit LeadCare II General Chemistry Blood Lead Test Whole Blood Sample 48 Tests Model Number: 70-6762
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- 2021-07-01TerminatedNonconforming Material/Component
LeadCare II Blood Lead Test Kit Catalog Number 70-6762
Magellan has received reports that control tests of either the Low-Control and/or the High-Control generated a low result. The issue may cause the underestimation of blood lead levels in patient samples.
- 2021-07-01TerminatedNonconforming Material/Component
LeadCare Plus Blood Lead Test Kit Catalog Number 82-0004
Magellan has received reports that control tests of either the Low-Control and/or the High-Control generated a low result. The issue may cause the underestimation of blood lead levels in patient samples.
- 2021-07-01TerminatedNonconforming Material/Component
LeadCare Ultra Blood Lead Test Kit Catalog Number 70-8098
Magellan has received reports that control tests of either the Low-Control and/or the High-Control generated a low result. The issue may cause the underestimation of blood lead levels in patient samples.
- 2017-08-15TerminatedDevice Design
LeadCare II Blood Lead Test Kit, containing LeadCare II Lead Controls Product Usage: IVD test kit for the detection of lead in blood. The kit consists of sensors, treatment reagent, lead controls (two levels), heparinized capillary tubes and plungers, transfer droppers and a calibration button. The LeadCare II Lead Controls packaged in the kit lots listed in #2 below are the subject of this recall (Field Correction) action.
LeadCare II Test Kits controls out of range