Enzyme Immunoassay, Diphenylhydantoin
FDA adverse event reports and recall history for enzyme immunoassay, diphenylhydantoin (product code DIP), from the public MAUDE corpus and the live FDA recall database.
- Total reports
- 924
- Deaths
- 0
- Injuries
- 2
- Malfunctions
- 921
Reports per year
Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.
Recent recalls
- 2014-10-29TerminatedUnder Investigation by firm
siemens Dimension Phenytoin Flex Reagent Cartridge The PTN method used on the Dimension clinical chemistry system is an in vitro diagnostic test intended to measure phenytoin (dilantin, diphenylhydantoin), an anti-epileptic drug, in human serum and plasma. PTN test results are used in the diagnosis and treatment of phenytoin overdose and in monitoring levels of phenytoin to ensure appropriate therapy.
Internal testing has confirmed a low absorbance range between the levels 1 and 5 calibrators with this lot resulting in the potential for a positive or negative bias and imprecision across the PTN assay range.
- 2014-10-02TerminatedSoftware Manufacturing/Software Deployment
VITROS Chemistry Products PHYT (Phenytoin) Slides, REF/Product Code 829 8671, IVD. For in vitro diagnostic use only. VITROS Chemistry Products PHYT Slides quantitatively measure phenytoin (PHYT) concentration in serum and plasma using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.
Ortho Clinical Diagnostics identified a potential for biased results to be generated when using VITROS PHYT Slides, Lot 2613-0150-4913. Some positively biased results were identified during testing of a different VITROS PHYT Slide lot; the affected slides were not released for distribution. Since Lot 2613-0150-4913 was manufactured in the same event, there is the potential for slides from this lot to also be affected.
- 2014-02-04TerminatedUnder Investigation by firm
ADVIA Centaur Systems Phenytoin Calibrator N ( PHTN) 2-pack SMN 10311391 For in vitro diagnostic use in the quantitative determination of phenytoin in serum or plasma using the ADVIA Centaur and ADVIA Centaur XP systems
ADVIA Centaur Systems Phenytoin Calibrator N, lots CN58 and higher, change in recovery observed with the ADVIA Centaur Systems Phenytoin assay compared to recoveries using previous Calibrator N lots.
- 2014-02-04TerminatedUnder Investigation by firm
ADVIA Centaur Systems Phenytoin Calibrator N ( PHTN) 6-pack SMN 10314977 For in vitro diagnostic use in the quantitative determination of phenytoin in serum or plasma using the ADVIA Centaur and ADVIA Centaur XP systems
ADVIA Centaur Systems Phenytoin Calibrator N, lots CN58 and higher, change in recovery observed with the ADVIA Centaur Systems Phenytoin assay compared to recoveries using previous Calibrator N lots.
- 2013-11-15TerminatedEquipment maintenance
VITROS Chemistry Products PHYT Slides packaged as 90 slides/pack catalog number 8298671 VITROS PHYT Slides quantitatively measure phenytoin (PHYT) concentration in serum and plasma using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.
Ortho Clinical Diagnostics is recalling Slide Dispense cartridges used with some VITROS Chemistry Product Slides due to slide dispense issues.
- 2008-09-16TerminatedOther
ABBOTT Clinical Chemistry Phenytoin, List Number: 1E07-20, is used for the quantitation of phenytoin in human serum or plasma.
An Abbott investigation has determined that the phenytoin reagent does not maintain an onboard stability of 28 days as stated in the phenytoin package insert. In addition, the 14 day calibration interval is also not being met. Further investigation is required to determine the root cause of this issue. Until the cause has been determined and a corrective action defined, specific instructions ar
- 2004-07-20TerminatedOther
Roche/Hitachi brand ONLINE TDM Phenytoin; Contents of kit: 1- 2 x 17 ml phenytoin conjugate and 2- 2 x 18 ml anti-phenytoin antibody (mouse monoclonal); latex microparticle. Product 03800458190.
May report falsely decreased patient results.
- 2004-07-20TerminatedOther
Roche/Hitachi brand ONLINE TDM Phenytoin; Contents of kit: 1- 1 x 37 ml phenytoin conjugate and 2- 1 x 38 ml anti-phenytoin antibody (mouse monoclonal); latex microparticle. Product 03516989190.
May report falsely decreased patient results.
- TerminatedDevice Design
VITROS Chemistry Products PHYT Slides, REF/CATALOG 829 8671, Rx ONLY, IVD; Ortho-Clinical Diagnostics, Inc. For in vitro diagnostic use only. VITROS Chemistry Products PHYT Slides quantitatively measure phenytoin (PHYT) concentration in serum and plasma using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.
The firm identified the potential for biased results to be generated when processing VITROS PHYT Slide product with citrate plasma specimen samples. The current Instructions for Use (IFU) and Assay Summary Chart state serum, heparin plasma, and citrate plasma as acceptable specimen types for use with the PHYT assay.
- TerminatedUnder Investigation by firm
VITROS Chemistry Products VersaTips - Product Usage: used to dispense samples to MicroSlide assays and sample and reagent fluid for MicroWell assays. In addition, VersaTips are used to dispense R1 Reagent Fluid into cuvettes in combination with sample fluid for Micro Tips Assays.
A deformity affecting a specific lot of VITROS Chemistry Products VersaTips. Deformed VersaTips results may be biased by up to +/-16% for both immunometric and competitive assays.
Monitor enzyme immunoassay, diphenylhydantoin automatically