Skip to main content
CIXClass 2

Bromcresol Green Dye-Binding, Albumin

FDA adverse event reports and recall history for bromcresol green dye-binding, albumin (product code CIX), from the public MAUDE corpus and the live FDA recall database.

Total reports
117
Deaths
0
Injuries
0
Malfunctions
117

Reports per year

07
10
11
12
16
17
18
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 2007–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Recent recalls

  • 2024-08-16Open, ClassifiedUnder Investigation by firm

    The VITROS XT Chemistry Products ALB-TP Slides is a single device that contains both an albumin test and a total protein test side by side separated by a plastic barrier sealed within a single slide frame. In this format, individual reactions occur, and test results are generated for each analyte independently of the other analyte. The ALB test is a multilayered, analytical element coated on a polyester support. For the albumin measurement, a drop of patient sample is deposited on the slide and is evenly distributed by the spreading layer to the underlying layers. When the fluid penetrates the reagent layer, the bromocresol green (BCG) dye diffuses to the spreading layer and binds to albumin in the sample. This binding results in a shift in wavelength of the reflectance maximum of the free dye. The color complex that forms is measured by reflectance spectrophotometry. The amount of albumin-bound dye is proportional to the concentration of albumin in the sample. Model/Catalog Number: 6844293

    The reason for this recall is QuidelOrtho has determined that during normal use the VITROS XT ALB-TP Slides may shed dust and debris as they travel through the MicroSlide Subsystem of the VITROS XT 3400 or VITROS XT 7600 analyzer. This dust and debris may then settle onto VITROS Na+ (Sodium) Slides as they are processed through the same MicroSlide Subsystem, potentially resulting in an increase in non-reproducible, positively, or negatively biased Na+ results. The level of bias will depend on where the dust and debris settle on the Na+ MicroSlide and may impact Na+ measurements to varying degrees inside and outside of the measuring range depending on the amount of contamination. All of which could vary from no/low bias with no patient harm, to extreme bias (falsely elevated or reduced results) potentially resulting in incorrect diagnosis; treatment; and/or a serious patient health hazard (for example, hypernatremia or hyponatremia). This issue may also potentially generate error codes preventing the analyzer from processing the MicroSlide Assays and cause delay of the patient results. All current and expired VITROS XT Chemistry Products ALB-TP slides are affected by this issue.

  • 2019-10-16TerminatedUnder Investigation by firm

    Roche Diagnostics cobas Integra Albumin Gen.2 Catalog Number: 05166861190 - Product Usage: In vitro test for the quantitative determination of the albumin in human serum and plasma on Roche/Hitachi cobas c systems.

    Due to low Quality Control recovery and invalid Calibration.

  • 2013-11-15TerminatedEquipment maintenance

    VITROS Chemistry Products ALB Slides, 90 slides/pack; catalog number 1988211 VITROS ALB Slides quantitatively measure (ALB) concentration in serum and plasma using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.

    Ortho Clinical Diagnostics is recalling Slide Dispense cartridges used with some VITROS Chemistry Product Slides due to slide dispense issues.

  • 2012-04-18TerminatedMaterial/Component Contamination

    Clinical Chemistry Albumin BCG (AlbG) List Number 7D53-21 Clinical Chemistry Albumin BCG 510(k) K981758, Product Code CIX The Albumin BCG assay is used for the quantitation of albumin in human serum or plasma

    Observed particulate matter in some reagent cartridges.

  • 2008-10-09TerminatedLabeling Change Control

    Roche Diagnostics ALBT2, Tina-quant Albumin Gen. 2, COBAS INTEGRA, cobas c systems, Roche Diagnostics, Indianapolis, IN.; 04469658190. An in-vitro diagnostic reagent system intended for use on COBAS Integra systems for the quantitative determination of albumin concentration in serum and plasma.

    Some of the labels are glossy, resulting in the bar code being difficult to read by the bar code reader.

  • TerminatedUnder Investigation by firm

    Albumin Gen.2, Catalog 05166861190

    Roche has confirmed customer complaints of low quality (QC) recovery and invalid calibrations for the identified products. A discoloration (yellow color) was observed and is the result of cross contamination between the R1 and R3 reagents.

Monitor bromcresol green dye-binding, albumin automatically

DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.