Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph
FDA adverse event reports and recall history for electrode measurement, blood-gases (pco2, po2) and blood ph (product code CHL), from the public MAUDE corpus and the live FDA recall database.
- Total reports
- 2,419
- Deaths
- 49
- Injuries
- 69
- Malfunctions
- 2,131
Reports per year
Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.
Recent recalls
- 2026-08-06Open, ClassifiedNonconforming Material/Component
GEM Premier 7000 with iQM3 Part Numbers: 00077407508 00077407510 00077415004 00077415008 00077415010 00077430010 Product Description: The GEM Premier 7000 with iQM3 is a portable critical care system for use by health care professionals to rapidly analyze lithium heparinized whole blood samples at the point of health care delivery in a clinical setting and in a central laboratory.
The active ingredient concentration in the lysing bag solution of specific GEM Premier 7000 with iQM3 PAK lots did not meet established manufacturing specifications. The malfunction could result in delayed recognition and management of Co-Oximetry and total bilirubin abnormalities and their sequelae.
- 2026-08-06Open, ClassifiedNonconforming Material/Component
GEM Premier 5000 PAK Part Numbers: 00055360010 00055360011 00055407504 00055415010 00055430010 00055430011 00055445010 00055445011 00055360010.A
The active ingredient concentration in the lysing bag solution of specific GEM Premier 5000 lots did not meet established manufacturing specifications. The malfunction could result in delayed recognition and management of Co-Oximetry and total bilirubin abnormalities and their sequelae.
- 2026-07-29Open, ClassifiedSoftware Design Change
Brand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 Intelligent Quality Management 2 (iQM2) is used as the quality control and assessment system for the GEM Premier 5000 system.
BCR wands that received a configuration update (released under Service Information No. 2026-07 on or after April 8, 2026) may assign patient results to an incorrect patient ID. The root cause is a race condition introduced when the "price check" feature was disabled to reduce excess current draw. This causes buffered data from a previous scan to be transmitted instead of the most recent scan, populating a prior patient ID instead of the current one. This can result in erroneous results being attributed to the wrong patient, potentially leading to inappropriate clinical management.
- 2026-07-29Open, ClassifiedSoftware Design Change
Brand Name: GEM Premier 7000 with iQM3 Product Name: GEM Premier 7000 with iQM3 Model/Catalog Number: 00000015279 Intelligent Quality Management 3 (iQM3) is used as the quality control (QC) and assessment system for the GEM Premier 7000.
BCR wands that received a configuration update (released under Service Information No. 2026-07 on or after April 8, 2026) may assign patient results to an incorrect patient ID. The root cause is a race condition introduced when the "price check" feature was disabled to reduce excess current draw. This causes buffered data from a previous scan to be transmitted instead of the most recent scan, populating a prior patient ID instead of the current one. This can result in erroneous results being attributed to the wrong patient, potentially leading to inappropriate clinical management.
- 2026-07-29Open, ClassifiedSoftware Design Change
Brand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 Product Description: The GEM Premier 5000 system provides fast, accurate, quantitative measurements of heparinized whole blood pH, pCO2, pO2, Na+, K+, Cl-, Ca++, glucose, lactate, Hct, total bilirubin and CO-Oximetry (tHb, O2Hb, COHb, MetHb, HHb, sO2) from arterial, venous or capillary samples. Intelligent Quality Management 2 (iQM2) is used as the quality control and assessment system for the GEM Premier 5000 system. iQM2 is an active quality process control program designed to provide continuous monitoring of the analytical process before, during, and after sample measurement with real-time, automatic error detection, automatic correction of the system and automatic documentation of all corrective actions, replacing the use of traditional external quality controls (QC). Facilities should follow local, state and federal regulatory guidelines to ensure that a total quality management system is followed. The GEM Premier 5000 system makes use of potentiometric sensors to measure pCO2, pH, Na+, K+, Cl-, and Ca++. It uses amperometric sensors to measure pO2, glucose, and lactate concentrations. Blood conductivity is the method used to measure hematocrit. CO-Oximetry and total bilirubin measurements involve chemically lysing the whole blood sample and then utilizing a broad spectrum spectrophotometer to evaluate the sample at a variety of wavelengths.
BCR wands that received a configuration update (released under Service Information No. 2026-07 on or after April 8, 2026) may assign patient results to an incorrect patient ID. The root cause is a race condition introduced when the "price check" feature was disabled to reduce excess current draw. This causes buffered data from a previous scan to be transmitted instead of the most recent scan, populating a prior patient ID instead of the current one. This can result in erroneous results being attributed to the wrong patient, potentially leading to inappropriate clinical management.
- 2026-07-29Open, ClassifiedSoftware Design Change
Brand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 Product Description: The GEM Premier 5000 system provides fast, accurate, quantitative measurements of heparinized whole blood pH, pCO2, pO2, Na+, K+, Cl-, Ca++, glucose, lactate, Hct, total bilirubin and CO-Oximetry (tHb, O2Hb, COHb, MetHb, HHb, sO2) from arterial, venous or capillary samples. Intelligent Quality Management 2 (iQM2) is used as the quality control and assessment system for the GEM Premier 5000 system. iQM2 is an active quality process control program designed to provide continuous monitoring of the analytical process before, during, and after sample measurement with real-time, automatic error detection, automatic correction of the system and automatic documentation of all corrective actions, replacing the use of traditional external quality controls (QC). Facilities should follow local, state and federal regulatory guidelines to ensure that a total quality management system is followed. The GEM Premier 5000 system makes use of potentiometric sensors to measure pCO2, pH, Na+, K+, Cl-, and Ca++. It uses amperometric sensors to measure pO2, glucose, and lactate concentrations. Blood conductivity is the method used to measure hematocrit. CO-Oximetry and total bilirubin measurements involve chemically lysing the whole blood sample and then utilizing a broad spectrum spectrophotometer to evaluate the sample at a variety of wavelengths.
BCR wands that received a configuration update (released under Service Information No. 2026-07 on or after April 8, 2026) may assign patient results to an incorrect patient ID. The root cause is a race condition introduced when the "price check" feature was disabled to reduce excess current draw. This causes buffered data from a previous scan to be transmitted instead of the most recent scan, populating a prior patient ID instead of the current one. This can result in erroneous results being attributed to the wrong patient, potentially leading to inappropriate clinical management.
- 2026-07-29Open, ClassifiedSoftware Design Change
Brand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 Product Description: The GEM Premier 5000 system provides fast, accurate, quantitative measurements of heparinized whole blood pH, pCO2, pO2, Na+, K+, Cl-, Ca++, glucose, lactate, Hct, total bilirubin and CO-Oximetry (tHb, O2Hb, COHb, MetHb, HHb, sO2) from arterial, venous or capillary samples. Intelligent Quality Management 2 (iQM2) is used as the quality control and assessment system for the GEM Premier 5000 system. iQM2 is an active quality process control program designed to provide continuous monitoring of the analytical process before, during, and after sample measurement with real-time, automatic error detection, automatic correction of the system and automatic documentation of all corrective actions, replacing the use of traditional external quality controls (QC). Facilities should follow local, state and federal regulatory guidelines to ensure that a total quality management system is followed. The GEM Premier 5000 system makes use of potentiometric sensors to measure pCO2, pH, Na+, K+, Cl-, and Ca++. It uses amperometric sensors to measure pO2, glucose, and lactate concentrations. Blood conductivity is the method used to measure hematocrit. CO-Oximetry and total bilirubin measurements involve chemically lysing the whole blood sample and then utilizing a broad spectrum spectrophotometer to evaluate the sample at a variety of wavelengths.
BCR wands that received a configuration update (released under Service Information No. 2026-07 on or after April 8, 2026) may assign patient results to an incorrect patient ID. The root cause is a race condition introduced when the "price check" feature was disabled to reduce excess current draw. This causes buffered data from a previous scan to be transmitted instead of the most recent scan, populating a prior patient ID instead of the current one. This can result in erroneous results being attributed to the wrong patient, potentially leading to inappropriate clinical management.
- 2026-07-29Open, ClassifiedSoftware Design Change
Brand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 Product Description: The GEM Premier 5000 system provides fast, accurate, quantitative measurements of heparinized whole blood pH, pCO2, pO2, Na+, K+, Cl-, Ca++, glucose, lactate, Hct, total bilirubin and CO-Oximetry (tHb, O2Hb, COHb, MetHb, HHb, sO2) from arterial, venous or capillary samples. Intelligent Quality Management 2 (iQM2) is used as the quality control and assessment system for the GEM Premier 5000 system. iQM2 is an active quality process control program designed to provide continuous monitoring of the analytical process before, during, and after sample measurement with real-time, automatic error detection, automatic correction of the system and automatic documentation of all corrective actions, replacing the use of traditional external quality controls (QC). Facilities should follow local, state and federal regulatory guidelines to ensure that a total quality management system is followed. The GEM Premier 5000 system makes use of potentiometric sensors to measure pCO2, pH, Na+, K+, Cl-, and Ca++. It uses amperometric sensors to measure pO2, glucose, and lactate concentrations. Blood conductivity is the method used to measure hematocrit. CO-Oximetry and total bilirubin measurements involve chemically lysing the whole blood sample and then utilizing a broad spectrum spectrophotometer to evaluate the sample at a variety of wavelengths.
BCR wands that received a configuration update (released under Service Information No. 2026-07 on or after April 8, 2026) may assign patient results to an incorrect patient ID. The root cause is a race condition introduced when the "price check" feature was disabled to reduce excess current draw. This causes buffered data from a previous scan to be transmitted instead of the most recent scan, populating a prior patient ID instead of the current one. This can result in erroneous results being attributed to the wrong patient, potentially leading to inappropriate clinical management.
- 2026-04-29Open, ClassifiedProcess change control
i-STAT EG7+ cartridge; List Number: 03P76-25;
Approximately 7.6% of cartridges from specific i-STAT EG7+ (list number 03P76-25), i-STAT EG6+ (list number 03P77-25) and i-STAT G3+ (list number 03P78-26) cartridge lots in the U.S. marketplace may report higher than expected PCO2 and lower than expected pH results due to a manufacturing issue. A falsely high pCO2 together with a falsely low pH makes the blood gas appear as if the patient is in severe respiratory acidosis. If the clinician believes the values are real, it may trigger unnecessary or even harmful interventions such as unnecessary ventilator changes, unnecessary medications, potential harm from overcorrection, or delays in identifying true problems.
- 2026-04-29Open, ClassifiedProcess change control
i-STAT EG6+ cartridge; List Number: 03P77-25;
Approximately 7.6% of cartridges from specific i-STAT EG7+ (list number 03P76-25), i-STAT EG6+ (list number 03P77-25) and i-STAT G3+ (list number 03P78-26) cartridge lots in the U.S. marketplace may report higher than expected PCO2 and lower than expected pH results due to a manufacturing issue. A falsely high pCO2 together with a falsely low pH makes the blood gas appear as if the patient is in severe respiratory acidosis. If the clinician believes the values are real, it may trigger unnecessary or even harmful interventions such as unnecessary ventilator changes, unnecessary medications, potential harm from overcorrection, or delays in identifying true problems.
Monitor electrode measurement, blood-gases (pco2, po2) and blood ph automatically
DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.