Electrode, Ion Based, Enzymatic, Creatinine
FDA adverse event reports and recall history for electrode, ion based, enzymatic, creatinine (product code CGL), from the public MAUDE corpus and the live FDA recall database.
- Total reports
- 198
- Deaths
- 3
- Injuries
- 10
- Malfunctions
- 182
Reports per year
Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.
Recent recalls
- 2025-05-27Open, ClassifiedStorage
ABBOTT POINT OF CARE TEST CREA I-STAT (25/BX) ABBTPT CARTRIDGE.
transit delays of certain cold chain products due to severe thunderstorms which may caused the products to be stored outside the recommended labeled storage conditions. This failure could increase the probability of an incorrect test result or a delayed test result.
- 2015-03-18TerminatedUnder Investigation by firm
Crea A and B membranes, packaged 4 per box Used in conjunction with the ABL800 FLEX analyzers which are intended for: In Vitro Testing of samples of whole blood for the parameters pCO2, cK+,cNa+, cCa2+,cCl-,cGlu, cLac, cCrea, ctBil, and co-oximetry parameters,(ctHb, sO2, and the hemoglobin fractions FO2Hb, FCOHb, FMetHb, FHHb and FHbF), in vitro testing of samples of expired air for the parameters pO2 and pCO2, in vitro testing of pleura samples for the pH parameter.
Negative drift on QC during in-use lifetime of CREA membranes can be observed.
- 2012-02-13TerminatedDevice Design
S1827 Crea Calibration solution 1, 944-135, Lot DV-02 Radiometer Medical, Bronshoj Denmark Product Usage: The S1827 Crea Calibration Solution 1 (944-135) Lot DV-02 is a solution used by the ABL 8x7 series creatine analyzers to establish calibration points for measured parameters.
When the user activates the dosicap (containing the crea additive) and mounts it onto the bottle, the threads will not seal correctly causing approximately 10% of the solution to leak out when the bottle is placed horizontally for dissolving the additive. This will cause the concentration of creatinin and creatin in the calibration solution to be wrong, and results in a bias on the Crea results.
- 2010-11-24TerminatedOther
i-STAT CREA Cartridge, Abbott List Number: 06F10-01, 02; IVD. Manufactured by Abbott Point of Care, Inc., Princeton, NJ.
The i-STAT CREA Cartridges may produce inaccurate (depressed) results at or near the end of the labeled expiry date when stored above 25 degrees for 2 weeks.
- 2010-03-05TerminatedOther
Reference Electrode Membranes for use with blood gas analyzer ABL 800 FLEX, when used on ABL 827 and ABL 837 with Creatinine modules (8x7). The device is packaged four (4) units per box.
Analysis of customer data, (for ABL 8X7 customers) has shown that the reference membranes are out of analytical specifications when replaced as specified in the device labeling. These errors in analytical specifications could lead to inadequate or clinical interventions (therapy) or misleading interpretation of patient status in clinical settings.
- 2009-05-07TerminatedProcess control
Stat Profile Critical Care Xpress (CCX and CCX+), CCX Creatinine Membrane Kit for CCX analyzer Catalog # 35238 Intended for in vitro diagnostic use by health care professionals and for point-of-care usage in the quantitative determination of pH, PCO2, PO2, SO2%, Hematocrit (Hct), Ca++, total hemoglobin (tHb), Oxyhemoglobin (O2Hb), Carboxyhemoglobin (COHb), Methemoglobin (MetHb), reduced hemoglobin (HHb), Oxygen content (O2Ct) and Oxygen capacity (O2Cap) in heparinized whole blood; Na+, K+, CL-; Ca++, Mg++, Glucose heparinized whole blood, serum, or plasma.
Erroneous, low creatinine results on CCX analyzer
- 2008-02-16TerminatedDevice Design
Nova - EZ CHEM Creatinine Meter: SRS Versions 5.1 and 5.2 P/N 9600, Nova Biomedical, Waltham, MA 02454
Incorrect Results - Creatinine results were lower than the laboratory reference method.
- 2008-02-16TerminatedDevice Design
Nova - EZ CHEM Control Solution P/N 9612, Nova Biomedical, Waltham, MA 02454
Incorrect Results - Creatinine results were lower than the laboratory reference method.
- 2008-02-16TerminatedDevice Design
Nova - EZ CHEM Linearity Solution P/N 9618, Nova Biomedical, Waltham, MA 02454
Incorrect Results - Creatinine results were lower than the laboratory reference method.
Monitor electrode, ion based, enzymatic, creatinine automatically
DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.