CECClass 2

Radioimmunoassay, Free Thyroxine

FDA adverse event reports and recall history for radioimmunoassay, free thyroxine (product code CEC), from the public MAUDE corpus and the live FDA recall database.

Total reports
1,944
Deaths
0
Injuries
25
Malfunctions
1,913

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Recent recalls

  • 2020-06-02TerminatedOther

    Beckman Coulter Access Immunoassay Systems, Access Free T4 Thyroxine, REF 33880 (UDI:15099590225834) for use with Access Family of Immunoassay Systems including: Access 2, UniCel DxI 600, UniCel DxI 800, UniCel DxC 600i, UniCel DxC 660i, UniCel DxC 680i, UniCel DxC 860i, and UniCel DxC 880i systems.

    The recalling firm characterized the potential interference effects of biotin (up to 1,200 ng/mL) on all biotin-mediated Access assays. The following six assays: Free T3-REF A13422, Total T3-REF 33830, Free T4-REF 33880, GI Mon-387687, Tg-REF 33860, and TgAbII-REF A32898 displayed significant interference (defined as a result bias greater than +/-10%) at 1,200 ng/mL of biotin.

  • 2018-05-04TerminatedError in labeling

    Dimension Free Thyroxine (FT4L) Flex reagent cartridge

    The concentrations for biotin listed in the non-interfering substances section of the instructions for use for the Dimension TSHL, Dimension LOCI FT4L, Dimension Vista TSH, ADVIA Centaur Cyclosporine, and ADVIA Centaur DHEAS, incorrectly state the level at which biotin does not interfere.

  • 2013-12-19TerminatedProcess change control

    FastPack Free T4 Immunoassay Chemilunescence assay for the determination of Free T4 The FastPack free T4 Immunoassay is a paramagnetic particle immunoassay for the in vitro quantitative determination of FT4 in human serum. The FastPack free T4 Immunoassay is designed for use with the FastPack System.

    Qualigen initiated this recall because the FastPack kit may cause lower than expected results due to a manufacturing defect.

  • 2011-04-11TerminatedOther

    Access Free T4 Calibrator, Part Number: 33885, Lot Numbers: 014769 The Access Free T4 Calibrators are intended to calibrate the Access Free T4 assay for the quantitative determination of free thyroxine levels in human serum and plasma (heparin) using the Access Immunoassay Systems

    The recall was initiated because the stability claim for Access Free T4 calibrator lot 014769 is not met. The affected T4 calibrator lot can produce quality control and patient results that are erroneously low (by greater than 2 standard deviations). An erroneously low patient result may lead to misdiagnosis and/or delayed treatment.

  • 2010-11-24TerminatedOther

    The Dimension FT4L Flex reagent cartridge( Free Thyroxine) manufactured by Siemens Healthcare Diagnostics, Inc., Newark, DE.

    Some LOCI cartridges have sub-optimal seal of the lidstock that can cause rupture of the package at high elevations, causing the cartridge to leak. Use of leaking cartridges may result in inaccurate test results.

  • 2010-10-01TerminatedOther

    Access Immunoassay Systems Free T4, Part Number: 33880

    The recall was initiated after Beckman Coulter determined that kit boxes for Access Free T4 reagent (REF 33880), lot number 916293 may contain reagent packs labeled as Access Vitamin B12, (REF 33000), lot 916277. Using an Access Vitamin B12 reagent pack, lot 916277, filled with Access Free T4 reagents may generate incorrect patient results.

  • 2003-08-13TerminatedOther

    Free T4 Control Pack, list 9C03-10; the kit consists of 3 bottles of controls, L - 0.7, M - 1.2 and H - 3.0; Abbott Laboratories, Abbott Park, IL 60064

    The Free T4 Controls, may yield control values that are high and outside of the package insert ranges.

  • 2003-08-13TerminatedOther

    IMx Free T4 Reagent Kit, list 2222-20, 100 test kit; Abbott Laboratories, Abbott Park, IL 60064

    The Free T4 Controls, may yield control values that are high and outside of the package insert ranges.

  • 2003-08-13TerminatedOther

    Architect Free T4 Calibrator Kit, list 7G96-01; the kit consists of 2 bottles of calibrators, CAL 1 - 0.5 and CAL 2 - 6.0; Abbott Laboratories, Abbott Park, IL 60064

    The Free T4 Controls, may yield control values that are high and outside of the package insert ranges.

  • TerminatedNonconforming Material/Component

    Alinity i Free T4 Reagent Kit, List Number 07P7030

    Abbott has become aware of a manufacturing issue that can result in damage to a bottle neck in some of the reagent cartridges found in the lots listed in Appendix A. If a bottle neck is damaged, it may cause a failure during the reagent loading process on the Alinity i instrument.