Radioimmunoassay, Testosterones And Dihydrotestosterone
FDA adverse event reports and recall history for radioimmunoassay, testosterones and dihydrotestosterone (product code CDZ), from the public MAUDE/AEMS corpus and the live FDA recall database.
- Total reports
- 160
- Deaths
- 0
- Injuries
- 22
- Malfunctions
- 132
Reports per year
Source: FDA adverse event reports by year received, 2012–2026. Reporting paused during the 2026 MAUDE-to-AEMS transition and has since resumed.
Recent recalls
- 2026-07-16Open, ClassifiedComponent design/selection
ST AIA-Pack Testosterone, IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of testosterone in human serum on specific Tosoh AIA System analyzers, Model Number 0025204
The Test cup lot F level 1 low-quality control results values for the device were lower than expected on the Automated Immunoassay Analyzer (AIA) which could cause falsely elevated or falsely low total testosterone results. Such erroneous test results may lead to additional testing/diagnostic interventions and in limited cases, may lead to a delay in diagnosis and/or initiation of unnecessary or incorrect clinical treatment measures.
- 2025-12-01Open, ClassifiedNo Marketing Application
Men s Hormone Test
Distribution without premarket approval/clearance.
- 2025-06-02Open, ClassifiedEquipment maintenance
Access Testosterone assay, Catalog Number 33560
Inadequate preventative maintenance (PM) performed on the scales of the RPL5 fill line, led to splashing (cross-contamination) between the wells of testosterone reagent packs lot 538081, which could potentially result in up to a 34% increase or a 23% decrease in patient results.
- 2025-01-03Open, ClassifiedLabeling Change Control
The FastPack IP Sex Hormone Binding Globulin Immunoassay, Part #25000081, is for the quantitative measurement of SHBG in human serum and plasma. For in vitro diagnostic Use and contains 30 FastPacks.
Three lots of the FastPacks kitted into the FastPack SHBG Complete Kits were labeled with the incorrect expiration date. The following is a listing of affected lots: Lot Number: 2408018-1, 2409015-1, and 2409015-2. The likelihood of patient harm is expected to be low as routine Quality Control should detect kits that are impacted by signal deterioration due to age.
- 2024-08-15Open, ClassifiedNonconforming Material/Component
Product Name: Salivary Testosterone ELISA Model/Catalog Number: SLV-3013 Product Description: An enzyme immunoassay for the quantitative in vitro diagnostic measurement of free active testosterone in saliva.
Invalid Controls. The Low Control is out of range; too high which causes invalid runs that would result in repeating the runs and delays for reporting results to patients.
- 2022-12-02Open, ClassifiedProcess control
IMMULITE and IMMULITE 1000 Total Testosterone Assay- In vitro diagnostic use with the IMMULITE and IMMULITE 1000 Analyzers for the quantitative measurement of total testosterone in serum and heparinized plasma.
Confirmed an average positive bias of 40% for testosterone concentrations of approximately 660 ng/dL (22.9 nmol/L) and above with the IMMULITE/IMMULITE 1000 Total Testosterone Assay when compared to the IMMULITE 2000/IMMULITE 2000 XPi Total Testosterone Assay. Two (2) lots of Total Testosterone assays for the IMMULITE/IMMULITE 1000, kit lots #515 and #516 were actively distributed to customers when the issue was discovered by Siemens, and they are subject to the correction or removal in this recall (867 total quantity of reagents distributed for each lot). The violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. In both males and female, testosterone results are not used in isolation but are interpreted in conjunction with clinical history and symptomology as well as with other laboratory testing (e.g. gonadotropins and other sex steroids). A delay in the diagnosis may be experienced and medical intervention, such as repeat testing may be required.
- 2017-06-29TerminatedUnder Investigation by firm
VIDAS Testosterone, Ref 30418
Complaints were received from customers observing falsely overestimate results or external quality control higher results than expected when performing tests with VIDAS Testosterone.
- 2017-05-03TerminatedNonconforming Material/Component
ARCHITECT SHBG Reagent The ARCHITECT SHBG assay is a chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of sex hormone binding globulin (SHBG) in human serum and plasma on the ARCHITECT iSystem. The ARCHITECT SHBG assay is used as an aid in the diagnosis of androgen disorders.
Firm has identified that ARCHITECT SHBG results generated with the Potassium EDTA specimen tube type may demonstrate a negative shift relative to results generated with serum. There is a potential for falsely decreased ARCHITECT SHBG results to be generated when using the ARCHITECT SHBG assay with the Potassium EDTA specimen tube type.
- 2017-04-19TerminatedNonconforming Material/Component
SHBG IMMULITE 2000/IMMULITE 2000 XPi For the in vitro diagnostic use with the IMMULITE 2000 Systems Analyzers-For the quantitative measurement of SHBG in serum, as aid in the differential diagnosis of hirsutism.
Certain lots of Sex Hormone Binding Globulin (SHBG) do not meet the thirty (30) day open vial adjustor stability claim when reconstituted and stored at 2-8C, as published in the Instructions for Use (IFU). Adjustors stored at 2-8C after 14 days may show a bias in patient samples and quality control falling outside of published ranges. Lookback not recommended.
- 2013-12-19TerminatedProcess change control
FastPack Testosterone Immunoassay Chemilunescence assay for the determination of Testosterone The FastPack Testo Immunoassay is a paramagnetic particle immunoassay for the in vitro quantitative determination of total testosterone in human serum and plasma. It is intended strictly for in-vitro diagnostic use as an aid in the diagnosis and management of conditions involving excess or deficiency of this androgen. The FastPack Testo Immunoassay is designed for use with the FastPack System.
Qualigen initiated this recall because the FastPack kit may cause lower than expected results due to a manufacturing defect.