Radioassay, Vitamin B12
FDA adverse event reports and recall history for radioassay, vitamin b12 (product code CDD), from the public MAUDE corpus and the live FDA recall database.
- Total reports
- 360
- Deaths
- 0
- Injuries
- 45
- Malfunctions
- 312
Reports per year
Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.
Recent recalls
- 2025-07-11CompletedProcess control
Access Vitamin B12 Reagent, Catalog Number 33000, Lot Number 439850
Beckman Coulter has received complaints of erroneous Vitamin B12 results when using Access Vitamin B12 Reagent lot 439850. The affected reagent lot may cause erroneously low or high Vitamin B12 results, leading to a delayed or improper diagnosis and/or repeat testing.
- 2024-09-05Open, ClassifiedNonconforming Material/Component
Power Express, REF B90918
The adhesive joint between the plastic socket and metal component of the Power Express Sample Holder can fail, causing the metal part to detach and enter the centrifuge. This detachment can result in damage to the centrifuge and potential destruction of the patient sample resulting in delayed results and possible exposure of lab technician to biohazardous material.
- 2024-09-05Open, ClassifiedNonconforming Material/Component
PE Centrifuge, REF: B36365, a component of the Power Express
The adhesive joint between the plastic socket and metal component of the Power Express Sample Holder can fail, causing the metal part to detach and enter the centrifuge. This detachment can result in damage to the centrifuge and potential destruction of the patient sample resulting in delayed results and possible exposure of lab technician to biohazardous material.
- 2024-09-05Open, ClassifiedNonconforming Material/Component
PE Centrifuge Temperature Controlled, REF: B36366, a component of the Power Express.
The adhesive joint between the plastic socket and metal component of the Power Express Sample Holder can fail, causing the metal part to detach and enter the centrifuge. This detachment can result in damage to the centrifuge and potential destruction of the patient sample resulting in delayed results and possible exposure of lab technician to biohazardous material.
- 2021-10-14Open, ClassifiedProcess control
Power Processor 3K, 3K High Speed Stockyards Module Nameplate: PP 3K Stockyard (Refrig) 9707 REF A88116
Missing splash guard which may expose users to hazardous materials. .
- 2018-05-04TerminatedError in labeling
IMMULITE /IMMULITE 1000 Vitamin B12
Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE assays, 3gAllergy Specific IgE, anti-HBc, BR-MA (CA15-3), CEA, Folic Acid, Gastrin, OM-MA (CA125), Thyroglobulin, and Vitamin B12, and these assays are susceptible to interference from biotin.
- 2012-06-27TerminatedDevice Design
Siemens Dimension Vista(R) System B12 Flex(R) Reagent Cartridge (K6442) B12 Flex Reagent Cartridge used with the Dimension Vista System for the in vitro diagnostic test for the quantitative measurement of Vitamin B12 in human serum and plasma.
Siemens Healthcare Diagnostics has confirmed that the presence of Intrinsic Factor Blocking Antibody in some patients can result in a false and significant elevation in the reported concentration of Vitamin B12 on the Dimension Vista System. This issue affects all lots of B12 Flex reagent cartridges on the Dimension Vista System.
- 2007-03-01TerminatedOther
Sample Management System software for in vitro diagnostic testing Product # 030102-03
Under limited circumstances, an incorrect patient result could be printed on the optional chartable patient report.
- 2005-02-11TerminatedOther
IMMULITE 2000 Vitamin B12 Kit, Catalog # L2KVB2, L2KVB6, kit lot 186.
A low CPS limit was not incorporated in the kit barcode for IMMULITE 2000 Vitamin B12 L2KVB 2, 6 kit lot 186 only.
- TerminatedUnder Investigation by firm
VITROS Immunodiagnostic Products Vitamin B12 Reagent Pack 1/2, Catalog Number 1453489 For in vitro diagnostic use only For the quantitative measurement of vitamin B12 in human serum and plasma (EDTA or heparin) using the VITROS ECi/ECiQ Immunodiagnostic Systems, the VITROS 3600 Immunodiagnostic System and the VITROS 5600 Integrated System, to aid in the differential diagnosis of anemia
The reagent pack exhibits the potential for an increased frequency of calibration failures.