Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase
FDA adverse event reports and recall history for analyzer, gas, carbon-dioxide, gaseous-phase (product code CCK), from the public MAUDE/AEMS corpus and the live FDA recall database.
- Total reports
- 12,579
- Deaths
- 34
- Injuries
- 128
- Malfunctions
- 12,392
Reports per year
Source: FDA adverse event reports by year received, 2012–2026. Reporting paused during the 2026 MAUDE-to-AEMS transition and has since resumed.
Recent recalls
- 2025-05-05Open, ClassifiedUnknown/Undetermined by firm
Microstream Advance Neonatal-Infant Intubated CO2 Filter Line marketed as Adt/Pedi Intub CO2 Line STerm , Adt/Pedi Intub CO2 Line STerm Lng , Adt/Pedi Intub CO2 Line High Humidity , Adt/Pedi Intub CO2 Line LTerm Lng , Adt/Pedi Intub CO2 Line LTerm , Neo/Inf Intub CO2 Line LTerm , Neo/Inf Intub CO2 Line High Humidity , Neo/Inf Intub CO2 Line LTerm Lng , VitaLine H Set Adult/Pediatric , VitaLine H Set Infant/Neonatal , FilterLine Set Long Adult/Pediatric , FilterLine H Set Long Adult/Pediatric , FilterLine H Set Long Infant/Neonatal , Trade Compliant: FilterLine H, Adult/Ped , Trade Compliant: FilterLine H, Infant/Neo , FilterLine Set Adult/Pedi , FilterLine H Set Adult/Pedi and FilterLine H Set Infant/Neonatal. Carbon dioxide gas analyzer.
Disconnection of the adapter from the endotracheal tube can potentially be difficult or not possible.
- 2025-03-20Open, ClassifiedProcess control
CAPNOCHECK BCI Airway Adapter; Model Number BSR; Product Code WW1100
Airway adapter does not meet specification and may break off when attaching a valve or tube. Device failure may result in inaccurate readings, a break in airway circuit, and loss of ventilation. A break in the airway circuit could result in inability to ventilate, hypoxia, cardiopulmonary collapse, or death.
- 2025-03-20Open, ClassifiedProcess control
CAPNOCHECK BCI Airway Adapter; Model Number BSR; Product Code WW8214
Airway adapter does not meet specification and may break off when attaching a valve or tube. Device failure may result in inaccurate readings, a break in airway circuit, and loss of ventilation. A break in the airway circuit could result in inability to ventilate, hypoxia, cardiopulmonary collapse, or death.
- 2025-03-17Open, ClassifiedLabeling design
Microstream Instructions for Use and Part Number used with - Product Description, REF: FilterLine H Set & VitaLine H Set Infant/Neonatal, 006147 - MICROSTREAM FilterLine H Set Infant/Neonatal CO2 Sampling Line and Airway Adapter for Humid Environments, 006324; MICROSTREAM FilterLine H Set Long Infant/Neonatal CO2 Sampling Line and Airway Adapter for Humid Environments, 007738; MICROSTREAM VitaLine H Set Infant/Neonatal CO2 Sampling Line and Airway Adapter for Humid Environments, 010807.
Neonatal-Infant and Adult-Pediatric CO2 Sampling Line/Airway Adapter/CO2 Oral-Nasal Filter Line/Intubated Filter (CO2) Line device users may experience difficulty/inability to disconnect adapter from endotracheal tube for procedures (e.g. suctioning), may cause treatment delay and/or unintended extubation, which could lead to respiratory failure, hypoxia, aspiration, and/or low oxygen saturation.
- 2025-03-17Open, ClassifiedLabeling design
Microstream Instructions for Use and Part Number used with - Product Description, REF: FilterLine Sets & VitaLine Sets Adult/Pediatric, MQ04616 - MICROSTREAM FilterLine H Set Long Adult-Pediatric CO2 Sampling Line and Airway Adapter for Humid Environments, 007737; MICROSTREAM FilterLine Set Long Adult-Pediatric CO2 Sampling Line and Airway Adapter, 007768; MICROSTREAM FilterLine Set Adult-Pediatric CO2 Sampling Line and Airway Adapter, 010579; MICROSTREAM FilterLine H Set Adult-Pediatric CO2 Sampling Line and Airway Adapter for Humid Environments, 010580; MICROSTREAM VitaLine H Set Adult-Pediatric CO2 Sampling Line and Airway Adapter for High Ambient Humidity X25, 010787; MICROSTREAM FilterLine Set Adult-Pediatric CO2 Sampling Line and Airway Adapter for Humid Environments, 015016; MICROSTREAM FilterLine Set Adult-Pediatric CO2 Sampling Line and Airway Adapter, 015021; MICROSTREAM FilterLine Set Adult-Pediatric CO2 Sampling Line and Airway Adapter, XS04620; MICROSTREAM FilterLine H Set Adult-Pediatric CO2 Sampling Line and Airway Adapter for Humid Environments, XS04624.
Neonatal-Infant and Adult-Pediatric CO2 Sampling Line/Airway Adapter/CO2 Oral-Nasal Filter Line/Intubated Filter (CO2) Line device users may experience difficulty/inability to disconnect adapter from endotracheal tube for procedures (e.g. suctioning), may cause treatment delay and/or unintended extubation, which could lead to respiratory failure, hypoxia, aspiration, and/or low oxygen saturation.
- 2025-03-17Open, ClassifiedLabeling design
Microstream Instructions for Use and Part Number used with - Product Description, REF: Filter Line Sets Adult/Pediatric & Infant/Neonatal, 10129497 - CapnoLine Neonatal-Infant Intubated Filter Line with Microstream Technology H, Length 7 (2.0 m), 006324COV; CapnoLine Adult-Pediatric Intubated Filter Line with Microstream Technology Length 7 (2.0 m), XS04620COV; CapnoLine Adult-Pediatric Intubated Filter Line with Microstream Technology H, Length 7 (2.0 m), XS04624COV
Neonatal-Infant and Adult-Pediatric CO2 Sampling Line/Airway Adapter/CO2 Oral-Nasal Filter Line/Intubated Filter (CO2) Line device users may experience difficulty/inability to disconnect adapter from endotracheal tube for procedures (e.g. suctioning), may cause treatment delay and/or unintended extubation, which could lead to respiratory failure, hypoxia, aspiration, and/or low oxygen saturation.
- 2025-03-17Open, ClassifiedLabeling design
Microstream Instructions for Use and Part Number used with - Product Description, REF: Airway Adapter and CO2 Sampling Line, 01158 - Microstream Luer Adult-Pediatric Airway Adapter, 010989; Omnistream CO2 Sampling Line X25, 010991
Neonatal-Infant and Adult-Pediatric CO2 Sampling Line/Airway Adapter/CO2 Oral-Nasal Filter Line/Intubated Filter (CO2) Line device users may experience difficulty/inability to disconnect adapter from endotracheal tube for procedures (e.g. suctioning), may cause treatment delay and/or unintended extubation, which could lead to respiratory failure, hypoxia, aspiration, and/or low oxygen saturation.
- 2025-03-17Open, ClassifiedLabeling design
Microstream Instructions for Use and Part Number used with - Product Description, REF: OmniVentLine Set, 012808 - OmniVentLine Set Adult-Pediatric CO2 Sampling Line and Airway Adapter, 012495;
Neonatal-Infant and Adult-Pediatric CO2 Sampling Line/Airway Adapter/CO2 Oral-Nasal Filter Line/Intubated Filter (CO2) Line device users may experience difficulty/inability to disconnect adapter from endotracheal tube for procedures (e.g. suctioning), may cause treatment delay and/or unintended extubation, which could lead to respiratory failure, hypoxia, aspiration, and/or low oxygen saturation.
- 2025-03-17Open, ClassifiedLabeling design
Microstream Instructions for Use and Part Number used with - Product Description, REF: Microstream Adv. Adult-Pediatric Intubated, PT00134333 - Microstream Advance Adult-Pediatric Intubated CO2 Filter Line 2m/6.5ft Short-term use: Procedural/Emergency, QMVAI; Microstream Advance Adult-Pediatric Intubated CO2 Filter Line 2m/6.5ft Extended Duration, QMVAIH
Neonatal-Infant and Adult-Pediatric CO2 Sampling Line/Airway Adapter/CO2 Oral-Nasal Filter Line/Intubated Filter (CO2) Line device users may experience difficulty/inability to disconnect adapter from endotracheal tube for procedures (e.g. suctioning), may cause treatment delay and/or unintended extubation, which could lead to respiratory failure, hypoxia, aspiration, and/or low oxygen saturation.
- 2025-03-17Open, ClassifiedLabeling design
Microstream Instructions for Use and Part Number used with - Product Description, REF: Microstream Adv. Infant Intubated, PT00134354 - Microstream Advance Neonatal-Infant Intubated CO2 Filter Line 2m/6.5ft Extended Duration, QMVIIH
Neonatal-Infant and Adult-Pediatric CO2 Sampling Line/Airway Adapter/CO2 Oral-Nasal Filter Line/Intubated Filter (CO2) Line device users may experience difficulty/inability to disconnect adapter from endotracheal tube for procedures (e.g. suctioning), may cause treatment delay and/or unintended extubation, which could lead to respiratory failure, hypoxia, aspiration, and/or low oxygen saturation.
Monitor analyzer, gas, carbon-dioxide, gaseous-phase automatically