Calculator, Predicted Values, Pulmonary Function
FDA adverse event reports and recall history for calculator, predicted values, pulmonary function (product code BTY), from the public MAUDE/AEMS corpus and the live FDA recall database.
- Total reports
- 136
- Deaths
- 0
- Injuries
- 9
- Malfunctions
- 121
Reports per year
Source: FDA adverse event reports by year received, 2004–2024. Reporting paused during the 2026 MAUDE-to-AEMS transition and has since resumed.
Recent recalls
- 2015-02-05TerminatedUnder Investigation by firm
Shape-HF Cardiopulmonary Testing System Disposable Patient Interface (DPI). Part Number 0004-4001. The Shape-HF Cardiopulmonary Exercise Testing System is a stationary device that monitors parameters during laboratory or clinical conditions.
This recall has been initiated because use of the impacted DPIs could result in erroneous testing results potentially leading to incorrect diagnosis and incorrect treatment. The magnitude of error for VO2 or VCO2 parameters could be approximately 20% with use of the impacted product.
- 2014-07-14TerminatedUnder Investigation by firm
Shape HF Cardiopulmonary System. Made up of Shape HF System Analyzer (PN 0001-9001) and Disposable Patient Interface (DPI), (PN 0004-4001). A pulmonary gas exchange testing system used during cardiopulmonary exercise testing where collection and review of gas exchange variables are indicated. The system will provide physiological data to physicians to aid in patient assessment.
Shape Medical has initiated a correction due to a mandatory software upgrade for the Shape HF Cardiopulmonary Testing System prior to use of the impacted DPI lot numbers. Use of the incorrect software version with the impacted DPIs could result in erroneous testing results potentially leading to incorrect diagnosis and incorrect treatment.
- TerminatedComponent change control
EasyOne Pro Respiratory Analysis System Pulmonary Function Testing Device, Model 3000-1
The CO/CO2 sensor cable harness may have lower electromagnetic shielding, which might have resulted in a non-compliance with the IEC 60601-1-2 standard. As a consequence of a lower electromagnetic shielding provided by the cable harness, the CO/CO2 sensor could have been influenced by high frequencies emitted by very close devices.
- TerminatedComponent change control
EasyOne Pro LAB VOS Respiratory Analysis System Pulmonary Function Testing Device, Model 3100-1
The CO/CO2 sensor cable harness may have lower electromagnetic shielding, which might have resulted in a non-compliance with the IEC 60601-1-2 standard. As a consequence of a lower electromagnetic shielding provided by the cable harness, the CO/CO2 sensor could have been influenced by high frequencies emitted by very close devices.