Humidifier, Respiratory Gas, (Direct Patient Interface)

FDA adverse event reports and recall history for humidifier, respiratory gas, (direct patient interface) (product code BTT), from the public MAUDE corpus and the live FDA recall database.

Total reports
19,184
Deaths
126
Injuries
410
Malfunctions
18,450

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Recent recalls

  • 2026-08-21Open, ClassifiedUnder Investigation by firm

    Medline Hudson RCI, DBD-CIRCUIT,VENT,INSPIRATORY,HEATED,NEO, Medline Product Number/SKU HUD78019 (Teleflex Product Number/SKU 780-19); breathing circuit

    Neonatal and infant heated wire breathing circuits may experience sporadic failure. The heated wire connector (pigtail) at the humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage. Device failure may predispose infant to bronchospasm, inspissated secretions, and airway obstruction, as well as inadvertent positive end-expiratory pressure (PEEP), circuit obstruction, or accidental lavage of pooled fluid into the airway, elevated oxygen consumption, and worsening respiratory distress.

  • 2025-10-15Open, ClassifiedNonconforming Material/Component

    Airvo 2, REF: PT101US, and myAirvo 2, REF: PT100US, used with Disinfection Kit User Manual, REF: UI-185043723

    When unintentionally disconnected from power source, humidifier device (used to deliver high flow warmed and humidified respiratory gases) may sound audible alert for less than 120 seconds, which may delay user awareness to reconnect power, which could lead to oxygen desaturation, hypoxia. Disinfection kit user manual updated to include checking the audible alert prior to use on a new patient.

  • 2024-05-15Open, ClassifiedNonconforming Material/Component

    Airvo 2 Humidifier, REF: PT101US, and myAirvo 2 Humidifier, REF: PT100US

    Humidifier devices, used to deliver high flow respiratory therapy to patients, manufactured before 14 August 2017, have a speaker configuration that may result in distorted, intermittent or inaudible alarm sound levels. In the absence of an audible alarm, if there is an interruption to therapy, a patient may experience oxygen desaturation.

  • 2023-03-17Open, ClassifiedNonconforming Material/Component

    Low Flow Disposable Patient Circuit for the Precision Flow System, Product Code PF-DPC-LOW

    There have been reports of excessive condensation associated with certain lots of the Low Flow Disposable Patient Circuits (DPCs) used with Precision Flow System which may lead to excessive rainout.

  • 2020-02-07TerminatedDevice Design

    HUDSON RCI COMFORT FLO Corrugated Humidification System, REF 2415, humidifier nebulizer kit

    There is the potential for water to flood the column and enter the circuit under circumstances where an abnormal pressure differential is created between the water bottle and the column during high flow oxygen therapy.

  • 2020-02-07TerminatedDevice Design

    HUDSON RCI COMFORT FLO Corrugated Humidification System with Remote Temperature Port, REF 2416, humidifier nebulizer kit

    There is the potential for water to flood the column and enter the circuit under circumstances where an abnormal pressure differential is created between the water bottle and the column during high flow oxygen therapy.

  • 2020-02-07TerminatedDevice Design

    HUDSON RCI COMFORT FLO Humidification System, REF 2410, humidifier nebulizer kit

    There is the potential for water to flood the column and enter the circuit under circumstances where an abnormal pressure differential is created between the water bottle and the column during high flow oxygen therapy.

  • 2020-02-07TerminatedDevice Design

    HUDSON RCI COMFORT FLO Humidification System With Remote Temperature Port, REF 2414, humidifier nebulizer kit

    There is the potential for water to flood the column and enter the circuit under circumstances where an abnormal pressure differential is created between the water bottle and the column during high flow oxygen therapy.

  • 2018-01-24TerminatedProcess control

    Infant AirLife single-limb, continuous-flow circuit (Model AH132) containing AirLife Humidification Chamber, Model Number AH290. The AH290 Humidification Chamber and Heated Breathing Circuit Kit(s) operate as part of a system for delivering health and humidified respirator gasses to patients requiring invasive and non-invasive ventilation.

    Manufacturing error in humidification chamber may cause an overfill/overflow condition of the chamber. The may case a water back-up into the patient circuit resulting in an excessive amount of water entering the airway/lungs of the ventilated patient.

  • 2018-01-24TerminatedProcess control

    Infant AirLife Dual-,limb, dual-heat, high-flow circuit (Model AH265) with AirLife Humidification Chamber, Model Number AH290. The AH290 Humidification Chamber and Heated Breathing Circuit Kit(s) operate as part of a system for delivering health and humidified respirator gasses to patients requiring invasive and non-invasive ventilation.

    Manufacturing error in humidification chamber may cause an overfill/overflow condition of the chamber. The may case a water back-up into the patient circuit resulting in an excessive amount of water entering the airway/lungs of the ventilated patient.