12 X-Ray Recall Records in One Day Trace to Two Recalls: Philips Drives, Canon Software

November 11, 20265 min read
DeviceWatch

DeviceWatch Team

Regulatory & Surveillance Experts


FDA posted 12 recall records for "Interventional Fluoroscopic X-Ray System" (product code OWB) on October 27, 2026. In the eight weeks before, it had posted one. The 12 records belong to two recalls from two firms. Nine come from Philips Medical Systems Nederland and cover its Azurion systems. Three come from Canon Medical Systems USA and cover its Alphenix angiography system. FDA classified all 12 as Class II.

What moved

Our weekly scan compares each device type's records against its trailing eight-week average. For this device type, the eight weeks from August 31 to October 25 held a single record: an Alphenix recall over loose bolts in the longitudinal movement mechanism, posted October 7. Against that baseline, 12 records in one day looks like a surge.

The longer view is calmer. In the 52 weeks before October 26, FDA posted 97 records for this device type, grouped into 31 recall events, or about two records a week. Two single Philips events were bigger than this week: 16 records posted January 13 and 22 records posted March 27. This device type tends to arrive in batches, because firms file one record per system configuration.

The 12 records

Philips, FDA recall event 99939, started by the firm October 6, 2026:

  • Z-0292-2027: Azurion 3 M12, models 722063 and 722221, 95 units
  • Z-0293-2027: Azurion 3 M15, models 722064, 722222 and 722280, 519 units
  • Z-0294-2027: Azurion 5 M12, model 722227, 60 units
  • Z-0295-2027: Azurion 5 M20, models 722228 and 722281, 465 units
  • Z-0296-2027: Azurion 7 B12, models 722067 and 722225, 209 units
  • Z-0297-2027: Azurion 7 B20, models 722068 and 722226, 457 units
  • Z-0298-2027: Azurion 7 M12, models 722078 and 722223, 602 units
  • Z-0299-2027: Azurion 7 M20, models 722079, 722224 and 722282
  • Z-0300-2027: Azurion S, model 722285, 1 unit

Canon, FDA recall event 99889, started by the firm October 12, 2026:

  • Z-0289-2027: Alphenix INFX-8000C, 15 systems
  • Z-0290-2027: Alphenix INFX-8000F, 3 units
  • Z-0291-2027: Alphenix INFX-8000V, 31 units

The Philips records list worldwide distribution, including the US. The Canon records list distribution in 18 US states.

Philips: drives that aren't in the device file

All nine Philips records give the same reason. Solid-state drives that are not specified in the Medical Device File may have been installed in various PCs in the system. The records say these drives may perform differently, including with reduced reliability over time. The effects they list are loss of imaging functionality or motorized movement, a blank or black display, and the inability to store newly acquired images or access previously acquired ones. FDA lists the cause as equipment maintenance.

Philips notified customers by certified letter, titled "URGENT Medical Device Correction." The records say systems may stay in use under their Instructions for Use and the letter's added guidance, which explains how to move the stand and table by hand if motorized movement fails. Philips plans to inspect affected systems and replace the drives, with that work expected to start in December 2026.

Canon: an error when two functions overlap

All three Canon records give the same reason. A software issue may cause an internal error if F-Store or image acquisition runs while the clinical analysis function is in use. FDA lists the cause as software design. Canon sent customers an urgent medical device correction on October 12, and a Canon representative will schedule installation of modified software that prevents the error.

The firms acted weeks earlier

Philips notified customers around October 6 and Canon on October 12. The classified records posted on October 27. A team that watches only the weekly recall feed learned of these actions two to three weeks after customers did.

Why the count matters for RA/QA teams

A recall count measures records, not problems. Twelve records here describe two problems, each filed once per system configuration. Trending by record count would show this week as a twelvefold jump. Trending by event shows two recalls, against 31 events over the prior year.

The two problems point to different controls. The Philips records describe parts that reached installed systems without being in the documented configuration, and FDA ties the cause to equipment maintenance. That makes it a question of how replacement parts are controlled through service as well as production. The Canon records describe an error that appears only when two functions run together, which is a question of how far concurrent-operation testing reaches.

What to do with this

A regulatory team should group recall records by FDA event number before trending volume, so one event filed per model counts once. Teams with imaging systems built on general-purpose PCs should check that parts installed in service match the documented configuration. Teams with image storage and analysis software should review whether their tests cover those functions running at the same time.